降低影像医学科常规摄影检查重照率

李宗憶 李宗憶, 丁詩偉 Zong-Yi Li, 陳善慈 Shih-Wei Ding, 林佳蓁 Shan-Tzu Chen, 李尉琦 李尉琦, 陳建成 Wei-Ci Li, 張瓊文 張瓊文, 朱惠珍 Chiung-Wen Chang
{"title":"降低影像医学科常规摄影检查重照率","authors":"李宗憶 李宗憶, 丁詩偉 Zong-Yi Li, 陳善慈 Shih-Wei Ding, 林佳蓁 Shan-Tzu Chen, 李尉琦 李尉琦, 陳建成 Wei-Ci Li, 張瓊文 張瓊文, 朱惠珍 Chiung-Wen Chang","doi":"10.53106/199457952023071704007","DOIUrl":null,"url":null,"abstract":"\n <p>統計2020年1~3月本科常規攝影平均重照率為6.41%。透過品管圈手法將主要重照原因分為「擺位不佳」、「體外異物」、「病人躁動」、「閉氣不佳」、「吸氣不足」及「項目條件錯誤」。經圈員們參考文獻與腦力激盪,根據三現原則、真因驗證,擬定6個對策群組共18項改善對策。包含對策群組1.神龍「擺位」、對策群組2.「異」不容辭、對策群組3.「靜」善「靜」美、對策群組4.不怕英雄氣短、對策群組5.告別雞同鴨講、對策群組6.望「成」莫「急」。對策實施後,X光影像重照率從改善前6.41%降低至3.43%。改善成果達到業界第一,並成為同儕標竿水準,未來將持續落實及精進對策的實施。</p>\n<p> </p><p>According to statistics, the average re-take rate of routine radiography from January to March 2020 is 6.41%. Using the method of quality control circle, the main reasons for re-take were divided into \"\"poor positioning\"\", \"\"foreign body\"\", \"\"patient agitation \"\", \"\"poor breath retention\"\", \"\"insufficient breathing\"\" and \"\"wrong item conditions\"\". After referring to the literature and brainstorming by the circle members, according to the principle of three occurrences and True Cause Verification, a total of 18 improvement countermeasures were drawn up in 6 countermeasure groups.The countermeasure groups content 1.\"\"well positioning\"\",2.\"\"without foreign body \"\",3.\"\"stand still\"\",4.\"\" well breath retention \"\",5.\"\" sufficient breathing\"\",6.\"\"right item conditions\"\". After the implementation of the countermeasures, the re-take rate of X-ray radiography decreased from 6.41% to 3.43%. The result is to reach No.1 and become the benchmark in the industry. In the future, we will continue to implement and improve the implementation of countermeasures.</p>\n<p> </p>\n","PeriodicalId":260200,"journal":{"name":"醫療品質雜誌","volume":"42 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2023-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"降低影像醫學科常規攝影檢查重照率\",\"authors\":\"李宗憶 李宗憶, 丁詩偉 Zong-Yi Li, 陳善慈 Shih-Wei Ding, 林佳蓁 Shan-Tzu Chen, 李尉琦 李尉琦, 陳建成 Wei-Ci Li, 張瓊文 張瓊文, 朱惠珍 Chiung-Wen Chang\",\"doi\":\"10.53106/199457952023071704007\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"\\n <p>統計2020年1~3月本科常規攝影平均重照率為6.41%。透過品管圈手法將主要重照原因分為「擺位不佳」、「體外異物」、「病人躁動」、「閉氣不佳」、「吸氣不足」及「項目條件錯誤」。經圈員們參考文獻與腦力激盪,根據三現原則、真因驗證,擬定6個對策群組共18項改善對策。包含對策群組1.神龍「擺位」、對策群組2.「異」不容辭、對策群組3.「靜」善「靜」美、對策群組4.不怕英雄氣短、對策群組5.告別雞同鴨講、對策群組6.望「成」莫「急」。對策實施後,X光影像重照率從改善前6.41%降低至3.43%。改善成果達到業界第一,並成為同儕標竿水準,未來將持續落實及精進對策的實施。</p>\\n<p> </p><p>According to statistics, the average re-take rate of routine radiography from January to March 2020 is 6.41%. Using the method of quality control circle, the main reasons for re-take were divided into \\\"\\\"poor positioning\\\"\\\", \\\"\\\"foreign body\\\"\\\", \\\"\\\"patient agitation \\\"\\\", \\\"\\\"poor breath retention\\\"\\\", \\\"\\\"insufficient breathing\\\"\\\" and \\\"\\\"wrong item conditions\\\"\\\". After referring to the literature and brainstorming by the circle members, according to the principle of three occurrences and True Cause Verification, a total of 18 improvement countermeasures were drawn up in 6 countermeasure groups.The countermeasure groups content 1.\\\"\\\"well positioning\\\"\\\",2.\\\"\\\"without foreign body \\\"\\\",3.\\\"\\\"stand still\\\"\\\",4.\\\"\\\" well breath retention \\\"\\\",5.\\\"\\\" sufficient breathing\\\"\\\",6.\\\"\\\"right item conditions\\\"\\\". After the implementation of the countermeasures, the re-take rate of X-ray radiography decreased from 6.41% to 3.43%. The result is to reach No.1 and become the benchmark in the industry. In the future, we will continue to implement and improve the implementation of countermeasures.</p>\\n<p> </p>\\n\",\"PeriodicalId\":260200,\"journal\":{\"name\":\"醫療品質雜誌\",\"volume\":\"42 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"醫療品質雜誌\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.53106/199457952023071704007\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"醫療品質雜誌","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.53106/199457952023071704007","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

统计2020年1~3月本科常规摄影平均重照率为6.41%。透过品管圈手法将主要重照原因分为「摆位不佳」、「体外异物」、「病人躁动」、「闭气不佳」、「吸气不足」及「项目条件错误」。经圈员们参考文献与脑力激荡,根据三现原则、真因验证,拟定6个对策群组共18项改善对策。包含对策群组1.神龙「摆位」、对策群组2.「异」不容辞、对策群组3.「静」善「静」美、对策群组4.不怕英雄气短、对策群组5.告别鸡同鸭讲、对策群组6.望「成」莫「急」。对策实施后,X光影像重照率从改善前6.41%降低至3.43%。改善成果达到业界第一,并成为同侪标竿水准,未来将持续落实及精进对策的实施。 According to statistics, the average re-take rate of routine radiography from January to March 2020 is 6.41%. Using the method of quality control circle, the main reasons for re-take were divided into ""poor positioning"", ""foreign body"", ""patient agitation "", ""poor breath retention"", ""insufficient breathing"" and ""wrong item conditions"". After referring to the literature and brainstorming by the circle members, according to the principle of three occurrences and True Cause Verification, a total of 18 improvement countermeasures were drawn up in 6 countermeasure groups.The countermeasure groups content 1.""well positioning"",2.""without foreign body "",3.""stand still"",4."" well breath retention "",5."" sufficient breathing"",6.""right item conditions"". After the implementation of the countermeasures, the re-take rate of X-ray radiography decreased from 6.41% to 3.43%. The result is to reach No.1 and become the benchmark in the industry. In the future, we will continue to implement and improve the implementation of countermeasures.
本文章由计算机程序翻译,如有差异,请以英文原文为准。
降低影像醫學科常規攝影檢查重照率

統計2020年1~3月本科常規攝影平均重照率為6.41%。透過品管圈手法將主要重照原因分為「擺位不佳」、「體外異物」、「病人躁動」、「閉氣不佳」、「吸氣不足」及「項目條件錯誤」。經圈員們參考文獻與腦力激盪,根據三現原則、真因驗證,擬定6個對策群組共18項改善對策。包含對策群組1.神龍「擺位」、對策群組2.「異」不容辭、對策群組3.「靜」善「靜」美、對策群組4.不怕英雄氣短、對策群組5.告別雞同鴨講、對策群組6.望「成」莫「急」。對策實施後,X光影像重照率從改善前6.41%降低至3.43%。改善成果達到業界第一,並成為同儕標竿水準,未來將持續落實及精進對策的實施。

 

According to statistics, the average re-take rate of routine radiography from January to March 2020 is 6.41%. Using the method of quality control circle, the main reasons for re-take were divided into ""poor positioning"", ""foreign body"", ""patient agitation "", ""poor breath retention"", ""insufficient breathing"" and ""wrong item conditions"". After referring to the literature and brainstorming by the circle members, according to the principle of three occurrences and True Cause Verification, a total of 18 improvement countermeasures were drawn up in 6 countermeasure groups.The countermeasure groups content 1.""well positioning"",2.""without foreign body "",3.""stand still"",4."" well breath retention "",5."" sufficient breathing"",6.""right item conditions"". After the implementation of the countermeasures, the re-take rate of X-ray radiography decreased from 6.41% to 3.43%. The result is to reach No.1 and become the benchmark in the industry. In the future, we will continue to implement and improve the implementation of countermeasures.

 

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信