输液器分析仪中红外光电二极管传感器对输液器的性能分析

Anisa Rahma Astuti, Syaifudin, Triana Rahmawati, K. Phasinam
{"title":"输液器分析仪中红外光电二极管传感器对输液器的性能分析","authors":"Anisa Rahma Astuti, Syaifudin, Triana Rahmawati, K. Phasinam","doi":"10.35882/jeeemi.v5i1.274","DOIUrl":null,"url":null,"abstract":"Infusion pumps and syringe pumps are devices used to administer liquid medicines to patients. The frequency of using the infusion pump and syringe pump in the long term will affect the accuracy of the tool. Accuracy is very important in dosing to patients in critical conditions who require intensive care to prevent fluid balance in the body. Therefore, periodic calibration of medical devices is required at least once a year. Calibration according to Permenkes No. 54 of 2015 is a calibration activity to determine the correctness of a tool. The purpose of this research is to make an Infusion Device Analyzer (IDA) with a TFT LCD displaying a graph of flowrate parameters. The method used is to analyze the flowrate value using an infrared photodiode sensor and can see the stability of the flowrate graph on a 7-inch TFT LCD from the use of 2 brands of syringes and an infusion set. The results obtained can be stored on the SD Card. The measurement results show that the error in the performance of the syringe and infusion pump read by the module on Channel 1 with the Terumo syringe is 0.15 (10 ml/h), 0.1 (50ml/h) and 0.03 (100ml/h). . On Channel 2 it is 0.02(10ml/hour), 0.03 (50ml/hour) and 0.04(100ml/hour). When using the B-Braun Channel 1 syringe, 0.25 (10ml/h), 009(50ml/h) and 0.08(100ml/h) are used. And on Channel 2 it is 0.62 (10ml/h), 0.15 (50ml/h), and 7.3 (100ml/h). When using the Terumo Channel 1 brand infusion set at 0.05 (10ml/h), 0.3(50ml/h), and 0.04(100ml/h). On Channel 2 it is 0.14(10ml/hour), 0.02 (50ml/hour) and 0.18 (100 ml/hour). When using the B-Braun Channel 1 Infusion Set, it is 0.07(10ml/h), 0.02(50 ml/h), and 0.03 (100ml/h). Then on Channel 2 0.07 (10ml/hour), 0.02(50 ml/hour), and 0.1(100ml/hour). The conclusion of this study is that the use of 2 different infusion sets greatly affects the readings, other than that other factors can also affect the readings including the position of the hose and the placement of sensors on each channel. From the manufacture of this tool, it is expected that users can be more efficient in using a 2-channel Infusion Device analyzer which can be run at the same time.","PeriodicalId":369032,"journal":{"name":"Journal of Electronics, Electromedical Engineering, and Medical Informatics","volume":"1 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2023-01-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"The Performance Analysis of the Infrared Photodiode Sensor to Infusion Set on Infusion Device Analyzer Machine\",\"authors\":\"Anisa Rahma Astuti, Syaifudin, Triana Rahmawati, K. Phasinam\",\"doi\":\"10.35882/jeeemi.v5i1.274\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Infusion pumps and syringe pumps are devices used to administer liquid medicines to patients. The frequency of using the infusion pump and syringe pump in the long term will affect the accuracy of the tool. Accuracy is very important in dosing to patients in critical conditions who require intensive care to prevent fluid balance in the body. Therefore, periodic calibration of medical devices is required at least once a year. Calibration according to Permenkes No. 54 of 2015 is a calibration activity to determine the correctness of a tool. The purpose of this research is to make an Infusion Device Analyzer (IDA) with a TFT LCD displaying a graph of flowrate parameters. The method used is to analyze the flowrate value using an infrared photodiode sensor and can see the stability of the flowrate graph on a 7-inch TFT LCD from the use of 2 brands of syringes and an infusion set. The results obtained can be stored on the SD Card. The measurement results show that the error in the performance of the syringe and infusion pump read by the module on Channel 1 with the Terumo syringe is 0.15 (10 ml/h), 0.1 (50ml/h) and 0.03 (100ml/h). . On Channel 2 it is 0.02(10ml/hour), 0.03 (50ml/hour) and 0.04(100ml/hour). When using the B-Braun Channel 1 syringe, 0.25 (10ml/h), 009(50ml/h) and 0.08(100ml/h) are used. And on Channel 2 it is 0.62 (10ml/h), 0.15 (50ml/h), and 7.3 (100ml/h). When using the Terumo Channel 1 brand infusion set at 0.05 (10ml/h), 0.3(50ml/h), and 0.04(100ml/h). On Channel 2 it is 0.14(10ml/hour), 0.02 (50ml/hour) and 0.18 (100 ml/hour). When using the B-Braun Channel 1 Infusion Set, it is 0.07(10ml/h), 0.02(50 ml/h), and 0.03 (100ml/h). Then on Channel 2 0.07 (10ml/hour), 0.02(50 ml/hour), and 0.1(100ml/hour). The conclusion of this study is that the use of 2 different infusion sets greatly affects the readings, other than that other factors can also affect the readings including the position of the hose and the placement of sensors on each channel. From the manufacture of this tool, it is expected that users can be more efficient in using a 2-channel Infusion Device analyzer which can be run at the same time.\",\"PeriodicalId\":369032,\"journal\":{\"name\":\"Journal of Electronics, Electromedical Engineering, and Medical Informatics\",\"volume\":\"1 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-01-28\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Electronics, Electromedical Engineering, and Medical Informatics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.35882/jeeemi.v5i1.274\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Electronics, Electromedical Engineering, and Medical Informatics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.35882/jeeemi.v5i1.274","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

摘要

输液泵和注射泵是用于给病人注射液体药物的装置。长期使用注射泵和注射泵的频率会影响工具的精度。对于需要重症监护以防止体内液体平衡的危重病人,给药的准确性非常重要。因此,每年至少需要对医疗器械进行一次定期校准。根据2015年Permenkes No. 54的校准是一项确定工具正确性的校准活动。本研究的目的是制造一种具有TFT液晶显示流量参数图的输液器分析仪(IDA)。所采用的方法是使用红外光电二极管传感器分析流量值,可以在7英寸TFT LCD上看到使用2个品牌的注射器和输液器的流量图的稳定性。得到的结果可以存储在SD卡上。测量结果表明,该模块在通道1上读取的Terumo注射器和注射泵的性能误差分别为0.15 (10ml /h)、0.1 (50ml/h)和0.03 (100ml/h)。在频道2,它是0.02(10ml/小时),0.03 (50ml/小时)和0.04(100ml/小时)。当使用B-Braun Channel 1注射器时,使用0.25 (10ml/h), 009(50ml/h)和0.08(100ml/h)。通道2是0.62 (10ml/h), 0.15 (50ml/h)和7.3 (100ml/h)。当使用Terumo Channel 1品牌输液器时,0.05 (10ml/h), 0.3(50ml/h)和0.04(100ml/h)。在通道2,它是0.14(10毫升/小时),0.02(50毫升/小时)和0.18(100毫升/小时)。当使用B-Braun 1通道输液器时,它是0.07(10ml/h), 0.02(50ml /h)和0.03 (100ml/h)。然后在通道2上0.07 (10ml/小时),0.02(50ml /小时)和0.1(100ml/小时)。本研究的结论是,使用两种不同的输液器对读数有很大的影响,除此之外,其他因素也会影响读数,包括软管的位置和每个通道上传感器的放置。从该工具的制造来看,期望用户可以更有效地使用可以同时运行的2通道输注设备分析仪。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
The Performance Analysis of the Infrared Photodiode Sensor to Infusion Set on Infusion Device Analyzer Machine
Infusion pumps and syringe pumps are devices used to administer liquid medicines to patients. The frequency of using the infusion pump and syringe pump in the long term will affect the accuracy of the tool. Accuracy is very important in dosing to patients in critical conditions who require intensive care to prevent fluid balance in the body. Therefore, periodic calibration of medical devices is required at least once a year. Calibration according to Permenkes No. 54 of 2015 is a calibration activity to determine the correctness of a tool. The purpose of this research is to make an Infusion Device Analyzer (IDA) with a TFT LCD displaying a graph of flowrate parameters. The method used is to analyze the flowrate value using an infrared photodiode sensor and can see the stability of the flowrate graph on a 7-inch TFT LCD from the use of 2 brands of syringes and an infusion set. The results obtained can be stored on the SD Card. The measurement results show that the error in the performance of the syringe and infusion pump read by the module on Channel 1 with the Terumo syringe is 0.15 (10 ml/h), 0.1 (50ml/h) and 0.03 (100ml/h). . On Channel 2 it is 0.02(10ml/hour), 0.03 (50ml/hour) and 0.04(100ml/hour). When using the B-Braun Channel 1 syringe, 0.25 (10ml/h), 009(50ml/h) and 0.08(100ml/h) are used. And on Channel 2 it is 0.62 (10ml/h), 0.15 (50ml/h), and 7.3 (100ml/h). When using the Terumo Channel 1 brand infusion set at 0.05 (10ml/h), 0.3(50ml/h), and 0.04(100ml/h). On Channel 2 it is 0.14(10ml/hour), 0.02 (50ml/hour) and 0.18 (100 ml/hour). When using the B-Braun Channel 1 Infusion Set, it is 0.07(10ml/h), 0.02(50 ml/h), and 0.03 (100ml/h). Then on Channel 2 0.07 (10ml/hour), 0.02(50 ml/hour), and 0.1(100ml/hour). The conclusion of this study is that the use of 2 different infusion sets greatly affects the readings, other than that other factors can also affect the readings including the position of the hose and the placement of sensors on each channel. From the manufacture of this tool, it is expected that users can be more efficient in using a 2-channel Infusion Device analyzer which can be run at the same time.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信