乳房手术中轻牵开器热损伤的预防:光缆口径的比较研究

Danielle Nizzero, Roland Deek, N. Dean
{"title":"乳房手术中轻牵开器热损伤的预防:光缆口径的比较研究","authors":"Danielle Nizzero, Roland Deek, N. Dean","doi":"10.34239/ajops.v5n1.289","DOIUrl":null,"url":null,"abstract":"Objective: Iatrogenic thermal injury can result from the use of lighted retractors during breast surgery due to the generation of heat at the connection between the optical cable and the lighted retractor. This study aims to determine whether a smaller calibre optical cable generates less heat at its connection with a Teb-betts lighted retractor.\nMethods: We measured the heat generated at the connection between an optical cable and a Tebbetts lighted retractor for a 3.0 mm optical cable and a 5.5 mm optical cable.\nResults: Both optical cables showed temperature rises. The 3.0 mm optical cable reached a peak temper-ature of 33.29 °C after 44 minutes and 47 seconds. The 5.5 mm optical cable reached a peak temperature of 49.23 °C after 53 minutes and 19 seconds. Light intensity was similar between the two cable types.\nConclusion: A 3.0 mm optical cable generates less heat than a 5.5 mm cable at the connection between the cable and the retractor. We recommend using the smaller calibre optical cable to reduce the risk of thermal injury to patients from the use of Tebbetts lighted retractors during breast surgery.","PeriodicalId":264055,"journal":{"name":"Australasian Journal of Plastic Surgery","volume":"47 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2022-03-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Prevention of thermal injury from lighted retractors in breast surgery: a comparative study of optical cable calibre\",\"authors\":\"Danielle Nizzero, Roland Deek, N. Dean\",\"doi\":\"10.34239/ajops.v5n1.289\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective: Iatrogenic thermal injury can result from the use of lighted retractors during breast surgery due to the generation of heat at the connection between the optical cable and the lighted retractor. This study aims to determine whether a smaller calibre optical cable generates less heat at its connection with a Teb-betts lighted retractor.\\nMethods: We measured the heat generated at the connection between an optical cable and a Tebbetts lighted retractor for a 3.0 mm optical cable and a 5.5 mm optical cable.\\nResults: Both optical cables showed temperature rises. The 3.0 mm optical cable reached a peak temper-ature of 33.29 °C after 44 minutes and 47 seconds. The 5.5 mm optical cable reached a peak temperature of 49.23 °C after 53 minutes and 19 seconds. Light intensity was similar between the two cable types.\\nConclusion: A 3.0 mm optical cable generates less heat than a 5.5 mm cable at the connection between the cable and the retractor. We recommend using the smaller calibre optical cable to reduce the risk of thermal injury to patients from the use of Tebbetts lighted retractors during breast surgery.\",\"PeriodicalId\":264055,\"journal\":{\"name\":\"Australasian Journal of Plastic Surgery\",\"volume\":\"47 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-03-31\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Australasian Journal of Plastic Surgery\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.34239/ajops.v5n1.289\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Australasian Journal of Plastic Surgery","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.34239/ajops.v5n1.289","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:在乳房手术中使用轻型牵开器时,由于光纤与轻型牵开器的连接处产生热量,可引起医源性热损伤。本研究旨在确定较小口径的光缆在与Teb-betts光牵开器连接时是否产生较少的热量。方法:测量3.0 mm光缆和5.5 mm光缆在光缆与Tebbetts点燃牵开器连接处产生的热量。结果:两根光缆均显示温度升高。3.0 mm光缆在44分47秒后达到最高温度33.29℃。5.5 mm光缆在53分19秒后达到49.23℃的峰值温度。两种电缆的光强相似。结论:3.0 mm光缆与牵开器连接处产生的热量小于5.5 mm光缆。我们建议使用较小口径的光缆,以减少乳房手术中使用Tebbetts点燃牵开器对患者造成热损伤的风险。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Prevention of thermal injury from lighted retractors in breast surgery: a comparative study of optical cable calibre
Objective: Iatrogenic thermal injury can result from the use of lighted retractors during breast surgery due to the generation of heat at the connection between the optical cable and the lighted retractor. This study aims to determine whether a smaller calibre optical cable generates less heat at its connection with a Teb-betts lighted retractor. Methods: We measured the heat generated at the connection between an optical cable and a Tebbetts lighted retractor for a 3.0 mm optical cable and a 5.5 mm optical cable. Results: Both optical cables showed temperature rises. The 3.0 mm optical cable reached a peak temper-ature of 33.29 °C after 44 minutes and 47 seconds. The 5.5 mm optical cable reached a peak temperature of 49.23 °C after 53 minutes and 19 seconds. Light intensity was similar between the two cable types. Conclusion: A 3.0 mm optical cable generates less heat than a 5.5 mm cable at the connection between the cable and the retractor. We recommend using the smaller calibre optical cable to reduce the risk of thermal injury to patients from the use of Tebbetts lighted retractors during breast surgery.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
0.30
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信