{"title":"ITS-90中锡点的开发与评价","authors":"K. Sato, J. Tamba, I. Kishimoto, M. Arai","doi":"10.1109/SICE.2002.1195319","DOIUrl":null,"url":null,"abstract":"The freezing point of tin, one of the temperature fixed-points defined in the ITS-90, is newly developed and evaluated. The temperature stability of the furnace is within /spl plusmn/1.5 mK for 24 hours and temperature distribution is within 2 mK over 20 cm. To realize a stable plateau after a large amount of supercooling, the air blow cooling method is introduced instead of the conventional method. The temperature during the plateaus decreases only 0.3 mK for 8 hours and a good repeatability within /spl plusmn/0.026 mK is obtained in ten measurements.","PeriodicalId":301855,"journal":{"name":"Proceedings of the 41st SICE Annual Conference. SICE 2002.","volume":"18 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2002-08-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Development and evaluation of the tin point in the ITS-90\",\"authors\":\"K. Sato, J. Tamba, I. Kishimoto, M. Arai\",\"doi\":\"10.1109/SICE.2002.1195319\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The freezing point of tin, one of the temperature fixed-points defined in the ITS-90, is newly developed and evaluated. The temperature stability of the furnace is within /spl plusmn/1.5 mK for 24 hours and temperature distribution is within 2 mK over 20 cm. To realize a stable plateau after a large amount of supercooling, the air blow cooling method is introduced instead of the conventional method. The temperature during the plateaus decreases only 0.3 mK for 8 hours and a good repeatability within /spl plusmn/0.026 mK is obtained in ten measurements.\",\"PeriodicalId\":301855,\"journal\":{\"name\":\"Proceedings of the 41st SICE Annual Conference. SICE 2002.\",\"volume\":\"18 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2002-08-05\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Proceedings of the 41st SICE Annual Conference. SICE 2002.\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/SICE.2002.1195319\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Proceedings of the 41st SICE Annual Conference. SICE 2002.","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/SICE.2002.1195319","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Development and evaluation of the tin point in the ITS-90
The freezing point of tin, one of the temperature fixed-points defined in the ITS-90, is newly developed and evaluated. The temperature stability of the furnace is within /spl plusmn/1.5 mK for 24 hours and temperature distribution is within 2 mK over 20 cm. To realize a stable plateau after a large amount of supercooling, the air blow cooling method is introduced instead of the conventional method. The temperature during the plateaus decreases only 0.3 mK for 8 hours and a good repeatability within /spl plusmn/0.026 mK is obtained in ten measurements.