{"title":"电场磨机标定系统的不确定度分析","authors":"E. Auden","doi":"10.51843/wsproceedings.2017.22","DOIUrl":null,"url":null,"abstract":"Abstract: We present the uncertainty analysis for an electric field mill calibration system used at the Sandia Primary Standards Laboratory. The uncertainty analysis incorporates contributions from DC voltage supplied by a high voltage amplifier; the distance between an electrically isolated charge plate and the ground plate of a stainless steel test structure; type A measurement uncertainty; and uncertainty associated with operator variation, including cable arrangement, electrical connections, and the angle at which the field mill is loaded into the test structure.","PeriodicalId":432978,"journal":{"name":"NCSL International Workshop & Symposium Conference Proceedings 2017","volume":"19 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1900-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Uncertainty Analysis of an Electric Field Mill Calibration System\",\"authors\":\"E. Auden\",\"doi\":\"10.51843/wsproceedings.2017.22\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Abstract: We present the uncertainty analysis for an electric field mill calibration system used at the Sandia Primary Standards Laboratory. The uncertainty analysis incorporates contributions from DC voltage supplied by a high voltage amplifier; the distance between an electrically isolated charge plate and the ground plate of a stainless steel test structure; type A measurement uncertainty; and uncertainty associated with operator variation, including cable arrangement, electrical connections, and the angle at which the field mill is loaded into the test structure.\",\"PeriodicalId\":432978,\"journal\":{\"name\":\"NCSL International Workshop & Symposium Conference Proceedings 2017\",\"volume\":\"19 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1900-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"NCSL International Workshop & Symposium Conference Proceedings 2017\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.51843/wsproceedings.2017.22\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"NCSL International Workshop & Symposium Conference Proceedings 2017","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.51843/wsproceedings.2017.22","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Uncertainty Analysis of an Electric Field Mill Calibration System
Abstract: We present the uncertainty analysis for an electric field mill calibration system used at the Sandia Primary Standards Laboratory. The uncertainty analysis incorporates contributions from DC voltage supplied by a high voltage amplifier; the distance between an electrically isolated charge plate and the ground plate of a stainless steel test structure; type A measurement uncertainty; and uncertainty associated with operator variation, including cable arrangement, electrical connections, and the angle at which the field mill is loaded into the test structure.