手术室家用电子监控设备的安全性

Y. Sarpong, A. Dagal, S. Sharar, A. Avellino
{"title":"手术室家用电子监控设备的安全性","authors":"Y. Sarpong, A. Dagal, S. Sharar, A. Avellino","doi":"10.5580/24c8","DOIUrl":null,"url":null,"abstract":"Due to increasing financial burden on the penal systems to keep offenders in the prisons, there have been increasing uses of electronic home monitoring devices to place “less risky” offenders under house arrest. These devices on several occasions have malfunctioned and caused electrical harm to wearers in cases of power surges and lightning strikes. Due to its relative novelty, there is no consensus or guidelines in the literature on how to manage patients with electronic home monitoring devices in surgical cases using monopolar electrocautery which can deliver charges that could potentially cause harm or malfunction of the monitoring system. Recently, we encountered a patient who needed bilateral cranioplasty procedures but had been placed on house arrest unbeknownst to us. During the surgery, we packed several gauzes in between the patient and the device and placed the electrocautery pad on the contralateral leg with power set at 20 MHz. The surgery was performed without any adverse outcome and the patient did well after surgery.","PeriodicalId":396781,"journal":{"name":"The Internet Journal of Anesthesiology","volume":"4 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2010-12-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"3","resultStr":"{\"title\":\"Safety of Electronic Home Monitoring Devices in the Operating Room\",\"authors\":\"Y. Sarpong, A. Dagal, S. Sharar, A. Avellino\",\"doi\":\"10.5580/24c8\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Due to increasing financial burden on the penal systems to keep offenders in the prisons, there have been increasing uses of electronic home monitoring devices to place “less risky” offenders under house arrest. These devices on several occasions have malfunctioned and caused electrical harm to wearers in cases of power surges and lightning strikes. Due to its relative novelty, there is no consensus or guidelines in the literature on how to manage patients with electronic home monitoring devices in surgical cases using monopolar electrocautery which can deliver charges that could potentially cause harm or malfunction of the monitoring system. Recently, we encountered a patient who needed bilateral cranioplasty procedures but had been placed on house arrest unbeknownst to us. During the surgery, we packed several gauzes in between the patient and the device and placed the electrocautery pad on the contralateral leg with power set at 20 MHz. The surgery was performed without any adverse outcome and the patient did well after surgery.\",\"PeriodicalId\":396781,\"journal\":{\"name\":\"The Internet Journal of Anesthesiology\",\"volume\":\"4 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2010-12-31\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"3\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"The Internet Journal of Anesthesiology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.5580/24c8\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"The Internet Journal of Anesthesiology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5580/24c8","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 3

摘要

由于将罪犯关押在监狱的刑事制度的财政负担日益增加,越来越多的人使用电子家庭监控设备将“危险性较低”的罪犯软禁起来。这些设备曾多次发生故障,并在电涌和雷击的情况下对佩戴者造成电气伤害。由于其相对新颖,在使用单极电灼术的手术病例中,如何管理使用电子家庭监测设备的患者,文献中没有共识或指南,因为单极电灼术可能产生可能造成伤害或监测系统故障的电荷。最近,我们遇到了一个需要双侧颅骨成形术的病人,但在我们不知情的情况下被软禁了。在手术中,我们在患者和装置之间塞了几块纱布,并将电灼垫放置在对侧腿上,功率设置为20 MHz。手术无不良反应,患者术后恢复良好。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Safety of Electronic Home Monitoring Devices in the Operating Room
Due to increasing financial burden on the penal systems to keep offenders in the prisons, there have been increasing uses of electronic home monitoring devices to place “less risky” offenders under house arrest. These devices on several occasions have malfunctioned and caused electrical harm to wearers in cases of power surges and lightning strikes. Due to its relative novelty, there is no consensus or guidelines in the literature on how to manage patients with electronic home monitoring devices in surgical cases using monopolar electrocautery which can deliver charges that could potentially cause harm or malfunction of the monitoring system. Recently, we encountered a patient who needed bilateral cranioplasty procedures but had been placed on house arrest unbeknownst to us. During the surgery, we packed several gauzes in between the patient and the device and placed the electrocautery pad on the contralateral leg with power set at 20 MHz. The surgery was performed without any adverse outcome and the patient did well after surgery.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信