{"title":"高压电容电桥校准装置","authors":"Dai Dong-xue, Wang Wei, Yang Yan, Chen Yan","doi":"10.1109/CPEM49742.2020.9191896","DOIUrl":null,"url":null,"abstract":"This paper describes a calibration device of high voltage capacitance bridge developed by National Institute of Metrology, China (NIM). The capacitance ratios are from 1/1 to 1000/1 with a MPE of $\\pm 1\\times 10^{-5}$, and dissipation factor (D) values are from 0.00001 to 0.1 with a MPE of $\\pm(0.1\\%\\times\\mathrm{D}+0.00001)$. A control program was developed by Labview software, which can automatically control the connection of the internal components, switch the measuring point under high voltage (HV).","PeriodicalId":373216,"journal":{"name":"2020 Conference on Precision Electromagnetic Measurements (CPEM)","volume":"71 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2020-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":"{\"title\":\"Calibration Device of High Voltage Capacitance Bridge\",\"authors\":\"Dai Dong-xue, Wang Wei, Yang Yan, Chen Yan\",\"doi\":\"10.1109/CPEM49742.2020.9191896\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This paper describes a calibration device of high voltage capacitance bridge developed by National Institute of Metrology, China (NIM). The capacitance ratios are from 1/1 to 1000/1 with a MPE of $\\\\pm 1\\\\times 10^{-5}$, and dissipation factor (D) values are from 0.00001 to 0.1 with a MPE of $\\\\pm(0.1\\\\%\\\\times\\\\mathrm{D}+0.00001)$. A control program was developed by Labview software, which can automatically control the connection of the internal components, switch the measuring point under high voltage (HV).\",\"PeriodicalId\":373216,\"journal\":{\"name\":\"2020 Conference on Precision Electromagnetic Measurements (CPEM)\",\"volume\":\"71 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2020-08-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"2\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2020 Conference on Precision Electromagnetic Measurements (CPEM)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/CPEM49742.2020.9191896\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2020 Conference on Precision Electromagnetic Measurements (CPEM)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/CPEM49742.2020.9191896","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Calibration Device of High Voltage Capacitance Bridge
This paper describes a calibration device of high voltage capacitance bridge developed by National Institute of Metrology, China (NIM). The capacitance ratios are from 1/1 to 1000/1 with a MPE of $\pm 1\times 10^{-5}$, and dissipation factor (D) values are from 0.00001 to 0.1 with a MPE of $\pm(0.1\%\times\mathrm{D}+0.00001)$. A control program was developed by Labview software, which can automatically control the connection of the internal components, switch the measuring point under high voltage (HV).