{"title":"一个定制的高精度电位计装置的性能","authors":"E. Arri, N. Locci, M. Tosi","doi":"10.1109/IMTC.1989.36872","DOIUrl":null,"url":null,"abstract":"The authors present the results of laboratory tests performed on an automatic high-precision device for voltage balancing. The system tested is based on an 18-bit D/A (digital/analog) converter and a dedicated null detector, both controlled by a personal computer. The high relative resolution allowed by the D/A converter is exploited by the system in different full-scale ranges (0.1-1-10 V). The relative resolution of balancing is equal to 3.8*10/sup -6/, corresponding to the least significant bit of the 18-bit D/A converter used. The system has been implemented for the purpose of achieving high accuracy with low-cost components and short operating times. The experimental results are in accordance with design specifications, with no significant deterioration of metrological characteristics, especially the relative resolution of the D/A converter.<<ETX>>","PeriodicalId":298343,"journal":{"name":"6th IEEE Conference Record., Instrumentation and Measurement Technology Conference","volume":"70 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1989-04-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Performance of a custom high precision potentiometric device\",\"authors\":\"E. Arri, N. Locci, M. Tosi\",\"doi\":\"10.1109/IMTC.1989.36872\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The authors present the results of laboratory tests performed on an automatic high-precision device for voltage balancing. The system tested is based on an 18-bit D/A (digital/analog) converter and a dedicated null detector, both controlled by a personal computer. The high relative resolution allowed by the D/A converter is exploited by the system in different full-scale ranges (0.1-1-10 V). The relative resolution of balancing is equal to 3.8*10/sup -6/, corresponding to the least significant bit of the 18-bit D/A converter used. The system has been implemented for the purpose of achieving high accuracy with low-cost components and short operating times. The experimental results are in accordance with design specifications, with no significant deterioration of metrological characteristics, especially the relative resolution of the D/A converter.<<ETX>>\",\"PeriodicalId\":298343,\"journal\":{\"name\":\"6th IEEE Conference Record., Instrumentation and Measurement Technology Conference\",\"volume\":\"70 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1989-04-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"6th IEEE Conference Record., Instrumentation and Measurement Technology Conference\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/IMTC.1989.36872\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"6th IEEE Conference Record., Instrumentation and Measurement Technology Conference","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/IMTC.1989.36872","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Performance of a custom high precision potentiometric device
The authors present the results of laboratory tests performed on an automatic high-precision device for voltage balancing. The system tested is based on an 18-bit D/A (digital/analog) converter and a dedicated null detector, both controlled by a personal computer. The high relative resolution allowed by the D/A converter is exploited by the system in different full-scale ranges (0.1-1-10 V). The relative resolution of balancing is equal to 3.8*10/sup -6/, corresponding to the least significant bit of the 18-bit D/A converter used. The system has been implemented for the purpose of achieving high accuracy with low-cost components and short operating times. The experimental results are in accordance with design specifications, with no significant deterioration of metrological characteristics, especially the relative resolution of the D/A converter.<>