F.L.I.P. Drijber, J.B. Finlay
{"title":"万向关节滑移是霍夫曼半框架外固定架失效的原因之一","authors":"F.L.I.P. Drijber,&nbsp;J.B. Finlay","doi":"10.1016/0141-5425(92)90105-T","DOIUrl":null,"url":null,"abstract":"<div><p>Slippage of the universal joints of external fixation devices is known to occur but its significance or incidence is often overlooked. In this study, controlled experiments were used to determine the relationship between joint slippage and the maximum loads a single half-frame could bear for the Hofmann device. The experiments showed that: (a) the joints slipped at minimal loads and (b) frame failure, i.e. loss of initial alignment of the frame components, was determined by joint slippage. The importance of the control of slippage cannot be overstated; the orthopaedic community must educate itself and its patients and guard against the problem in order to avoid complications secondary to slippage.</p></div>","PeriodicalId":75992,"journal":{"name":"Journal of biomedical engineering","volume":"14 6","pages":"Pages 509-515"},"PeriodicalIF":0.0000,"publicationDate":"1992-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/0141-5425(92)90105-T","citationCount":"8","resultStr":"{\"title\":\"Universal joint slippage as a cause of Hofmann half-frame external fixator failure\",\"authors\":\"F.L.I.P. Drijber,&nbsp;J.B. Finlay\",\"doi\":\"10.1016/0141-5425(92)90105-T\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>Slippage of the universal joints of external fixation devices is known to occur but its significance or incidence is often overlooked. In this study, controlled experiments were used to determine the relationship between joint slippage and the maximum loads a single half-frame could bear for the Hofmann device. The experiments showed that: (a) the joints slipped at minimal loads and (b) frame failure, i.e. loss of initial alignment of the frame components, was determined by joint slippage. The importance of the control of slippage cannot be overstated; the orthopaedic community must educate itself and its patients and guard against the problem in order to avoid complications secondary to slippage.</p></div>\",\"PeriodicalId\":75992,\"journal\":{\"name\":\"Journal of biomedical engineering\",\"volume\":\"14 6\",\"pages\":\"Pages 509-515\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1992-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1016/0141-5425(92)90105-T\",\"citationCount\":\"8\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of biomedical engineering\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/014154259290105T\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of biomedical engineering","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/014154259290105T","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 8

摘要

众所周知,外固定装置万向关节会发生滑移,但其重要性或发生率往往被忽视。在本研究中,采用对照实验来确定关节滑移与Hofmann装置单个半框架所能承受的最大载荷之间的关系。试验表明:(a)节点在最小荷载下发生滑移;(b)节点滑移决定了框架的破坏,即框架构件的初始对中丢失。控制滑移的重要性怎么强调都不过分;骨科界必须对自己和患者进行教育,并对这一问题加以防范,以避免滑移引起的并发症。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Universal joint slippage as a cause of Hofmann half-frame external fixator failure

Slippage of the universal joints of external fixation devices is known to occur but its significance or incidence is often overlooked. In this study, controlled experiments were used to determine the relationship between joint slippage and the maximum loads a single half-frame could bear for the Hofmann device. The experiments showed that: (a) the joints slipped at minimal loads and (b) frame failure, i.e. loss of initial alignment of the frame components, was determined by joint slippage. The importance of the control of slippage cannot be overstated; the orthopaedic community must educate itself and its patients and guard against the problem in order to avoid complications secondary to slippage.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信