{"title":"射频刺激器的安全要求和试验方法","authors":"Sang-Geon Park","doi":"10.6109/jicce.2016.14.3.191","DOIUrl":null,"url":null,"abstract":"In this study, we investigate the safety requirements and test methods of a radiofrequency stimulator. The main test items include controls of a minimum output, accommodation range, and output parameters that have been known as the safety requirements in conformity with international standards. As the test criteria for controlling the minimum output, an increase or decrease in a unit of 1 mA or 1 V or less was applied to the output amplitude regulator for both continuous and discontinuous control, and the output at the minimum setting was manipulated to not exceed 2% of the maximum setting. For controlling the output parameters, one of the representative test criteria states that the current limit of 250 mA should be equal to or less than 1,500 Hz. Consequently, when applying the radiofrequency stimulator on the human body, we need to ensure that the safety requirements conform to the international standards.","PeriodicalId":272551,"journal":{"name":"J. Inform. and Commun. Convergence Engineering","volume":"22 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2016-09-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Safety Requirements and Test Methods of a Radiofrequency Stimulator\",\"authors\":\"Sang-Geon Park\",\"doi\":\"10.6109/jicce.2016.14.3.191\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"In this study, we investigate the safety requirements and test methods of a radiofrequency stimulator. The main test items include controls of a minimum output, accommodation range, and output parameters that have been known as the safety requirements in conformity with international standards. As the test criteria for controlling the minimum output, an increase or decrease in a unit of 1 mA or 1 V or less was applied to the output amplitude regulator for both continuous and discontinuous control, and the output at the minimum setting was manipulated to not exceed 2% of the maximum setting. For controlling the output parameters, one of the representative test criteria states that the current limit of 250 mA should be equal to or less than 1,500 Hz. Consequently, when applying the radiofrequency stimulator on the human body, we need to ensure that the safety requirements conform to the international standards.\",\"PeriodicalId\":272551,\"journal\":{\"name\":\"J. Inform. and Commun. Convergence Engineering\",\"volume\":\"22 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2016-09-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"J. Inform. and Commun. Convergence Engineering\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.6109/jicce.2016.14.3.191\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"J. Inform. and Commun. Convergence Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6109/jicce.2016.14.3.191","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Safety Requirements and Test Methods of a Radiofrequency Stimulator
In this study, we investigate the safety requirements and test methods of a radiofrequency stimulator. The main test items include controls of a minimum output, accommodation range, and output parameters that have been known as the safety requirements in conformity with international standards. As the test criteria for controlling the minimum output, an increase or decrease in a unit of 1 mA or 1 V or less was applied to the output amplitude regulator for both continuous and discontinuous control, and the output at the minimum setting was manipulated to not exceed 2% of the maximum setting. For controlling the output parameters, one of the representative test criteria states that the current limit of 250 mA should be equal to or less than 1,500 Hz. Consequently, when applying the radiofrequency stimulator on the human body, we need to ensure that the safety requirements conform to the international standards.