{"title":"使用电感分压器进行交流电压比校准","authors":"K. Draxler, R. Styblíková","doi":"10.1109/I2MTC.2013.6555398","DOIUrl":null,"url":null,"abstract":"This paper presents the calibration of an electronic voltage divider and precise instrument voltage transformers in the range up to 1 kV at 50 Hz frequency using inductive voltage dividers. We also describe a way to evaluate the difference between a standard inductive divider and a calibrated instrument voltage transformer or electronic voltage divider expressed as the ratio error and phase displacement. The paper also presents the construction of a standard decadic voltage divider with a range of input voltage (10×100) V and (10×10) V.","PeriodicalId":432388,"journal":{"name":"2013 IEEE International Instrumentation and Measurement Technology Conference (I2MTC)","volume":"1 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2013-05-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Use of an inductive divider for AC voltage ratio calibration\",\"authors\":\"K. Draxler, R. Styblíková\",\"doi\":\"10.1109/I2MTC.2013.6555398\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This paper presents the calibration of an electronic voltage divider and precise instrument voltage transformers in the range up to 1 kV at 50 Hz frequency using inductive voltage dividers. We also describe a way to evaluate the difference between a standard inductive divider and a calibrated instrument voltage transformer or electronic voltage divider expressed as the ratio error and phase displacement. The paper also presents the construction of a standard decadic voltage divider with a range of input voltage (10×100) V and (10×10) V.\",\"PeriodicalId\":432388,\"journal\":{\"name\":\"2013 IEEE International Instrumentation and Measurement Technology Conference (I2MTC)\",\"volume\":\"1 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2013-05-06\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2013 IEEE International Instrumentation and Measurement Technology Conference (I2MTC)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/I2MTC.2013.6555398\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2013 IEEE International Instrumentation and Measurement Technology Conference (I2MTC)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/I2MTC.2013.6555398","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Use of an inductive divider for AC voltage ratio calibration
This paper presents the calibration of an electronic voltage divider and precise instrument voltage transformers in the range up to 1 kV at 50 Hz frequency using inductive voltage dividers. We also describe a way to evaluate the difference between a standard inductive divider and a calibrated instrument voltage transformer or electronic voltage divider expressed as the ratio error and phase displacement. The paper also presents the construction of a standard decadic voltage divider with a range of input voltage (10×100) V and (10×10) V.