蒙特卡罗环境下闪烁相机固有分辨率的仿真

Xianling Dong, W. H. M. Saad, W. Adnan, S. Hashim, Nor Pai'za Mohd. Hassan, A. Nordin, M. I. Saripan
{"title":"蒙特卡罗环境下闪烁相机固有分辨率的仿真","authors":"Xianling Dong, W. H. M. Saad, W. Adnan, S. Hashim, Nor Pai'za Mohd. Hassan, A. Nordin, M. I. Saripan","doi":"10.1109/ICSIPA.2013.6707969","DOIUrl":null,"url":null,"abstract":"In a typical scintillation camera system, intrinsic resolution is dependent upon the accuracy of the identification of an interaction position. This paper intends to set up a evaluation tool based on Monte Carlo simulation for the purpose of estimating the intrinsic resolution of scintillation cameras. Monte Carlo N-Particles (MCNP) Code was applied to simulate the components of the model platform referred to Toshiba GCA-7100A with a monolithic Sodium Iodide (NaI) scintillator (40 × 40 × 0.9525 cm3). The simulation result was 3.7 mm full width at half maximum (FWHM) for intrinsic resolution which came out to be in a good agreement with the experimental result. This suggests that our proposed evaluation tool may help to optimize the parameters of the detector without physical experiments.","PeriodicalId":440373,"journal":{"name":"2013 IEEE International Conference on Signal and Image Processing Applications","volume":"2 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1900-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Simulation of intrinsic resolution of scintillation camera in Monte Carlo environment\",\"authors\":\"Xianling Dong, W. H. M. Saad, W. Adnan, S. Hashim, Nor Pai'za Mohd. Hassan, A. Nordin, M. I. Saripan\",\"doi\":\"10.1109/ICSIPA.2013.6707969\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"In a typical scintillation camera system, intrinsic resolution is dependent upon the accuracy of the identification of an interaction position. This paper intends to set up a evaluation tool based on Monte Carlo simulation for the purpose of estimating the intrinsic resolution of scintillation cameras. Monte Carlo N-Particles (MCNP) Code was applied to simulate the components of the model platform referred to Toshiba GCA-7100A with a monolithic Sodium Iodide (NaI) scintillator (40 × 40 × 0.9525 cm3). The simulation result was 3.7 mm full width at half maximum (FWHM) for intrinsic resolution which came out to be in a good agreement with the experimental result. This suggests that our proposed evaluation tool may help to optimize the parameters of the detector without physical experiments.\",\"PeriodicalId\":440373,\"journal\":{\"name\":\"2013 IEEE International Conference on Signal and Image Processing Applications\",\"volume\":\"2 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1900-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2013 IEEE International Conference on Signal and Image Processing Applications\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/ICSIPA.2013.6707969\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2013 IEEE International Conference on Signal and Image Processing Applications","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ICSIPA.2013.6707969","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

摘要

在典型的闪烁相机系统中,固有分辨率依赖于相互作用位置的识别精度。本文拟建立一种基于蒙特卡罗模拟的闪烁相机固有分辨率评估工具。采用蒙特卡罗n粒子(MCNP)代码,采用单片碘化钠(NaI)闪烁体(40 × 40 × 0.9525 cm3)模拟了东芝GCA-7100A模型平台的组件。本征分辨率为3.7 mm,与实验结果吻合较好。这表明我们提出的评估工具可以帮助优化探测器的参数,而无需物理实验。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Simulation of intrinsic resolution of scintillation camera in Monte Carlo environment
In a typical scintillation camera system, intrinsic resolution is dependent upon the accuracy of the identification of an interaction position. This paper intends to set up a evaluation tool based on Monte Carlo simulation for the purpose of estimating the intrinsic resolution of scintillation cameras. Monte Carlo N-Particles (MCNP) Code was applied to simulate the components of the model platform referred to Toshiba GCA-7100A with a monolithic Sodium Iodide (NaI) scintillator (40 × 40 × 0.9525 cm3). The simulation result was 3.7 mm full width at half maximum (FWHM) for intrinsic resolution which came out to be in a good agreement with the experimental result. This suggests that our proposed evaluation tool may help to optimize the parameters of the detector without physical experiments.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信