{"title":"显微镜无菌系统。","authors":"S G Kramer,&nbsp;D Char","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The problem of microscope sterility during ophthalmic microsurgery has generally been managed by using sterilizable handle protectors, or by using some form of plastic sterilizable drape for the entire microscope. Both of these systems have disadvantages. A simple and effective alternative which has been designed for Zeiss operating microscopes and can be modified for other types of instruments is presented. No modification of the microscope is required. The instrument consists of two pieces: a metal bracket which mounts in the accessory shoe of the operating microscope, and a rigid plastic sterilizable portion which is readily attached on the microscope from below with three sterilizable screws. The entire assembly can be attached within 15 seconds. It affords complete sterility protection of the operative field; it provides a sterile surface for handling and moving the microscope; it does not interfere with accessory lighting systerms; and it allows open access to external accessories such as cameras and observer tubes.</p>","PeriodicalId":23219,"journal":{"name":"Transactions. Section on Ophthalmology. American Academy of Ophthalmology and Otolaryngology","volume":"83 5","pages":"869-71"},"PeriodicalIF":0.0000,"publicationDate":"1977-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Microscope sterility system.\",\"authors\":\"S G Kramer,&nbsp;D Char\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The problem of microscope sterility during ophthalmic microsurgery has generally been managed by using sterilizable handle protectors, or by using some form of plastic sterilizable drape for the entire microscope. Both of these systems have disadvantages. A simple and effective alternative which has been designed for Zeiss operating microscopes and can be modified for other types of instruments is presented. No modification of the microscope is required. The instrument consists of two pieces: a metal bracket which mounts in the accessory shoe of the operating microscope, and a rigid plastic sterilizable portion which is readily attached on the microscope from below with three sterilizable screws. The entire assembly can be attached within 15 seconds. It affords complete sterility protection of the operative field; it provides a sterile surface for handling and moving the microscope; it does not interfere with accessory lighting systerms; and it allows open access to external accessories such as cameras and observer tubes.</p>\",\"PeriodicalId\":23219,\"journal\":{\"name\":\"Transactions. Section on Ophthalmology. American Academy of Ophthalmology and Otolaryngology\",\"volume\":\"83 5\",\"pages\":\"869-71\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1977-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Transactions. Section on Ophthalmology. American Academy of Ophthalmology and Otolaryngology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Transactions. Section on Ophthalmology. American Academy of Ophthalmology and Otolaryngology","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

在眼科显微手术中,显微镜的无菌问题通常是通过使用可消毒的手柄保护器,或在整个显微镜上使用某种形式的可消毒的塑料窗帘来解决的。这两种系统都有缺点。一个简单而有效的替代方案,已设计蔡司操作显微镜,并可以修改为其他类型的仪器提出。不需要修改显微镜。该仪器由两部分组成:一个金属支架安装在操作显微镜的附属鞋上,另一个是坚硬的塑料灭菌部分,它可以用三个灭菌螺钉从下面固定在显微镜上。整个组件可以在15秒内安装。它提供手术领域的完全无菌保护;它为操作和移动显微镜提供无菌表面;它不会干扰辅助照明系统;它还可以开放使用外部配件,如摄像头和观察管。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Microscope sterility system.

The problem of microscope sterility during ophthalmic microsurgery has generally been managed by using sterilizable handle protectors, or by using some form of plastic sterilizable drape for the entire microscope. Both of these systems have disadvantages. A simple and effective alternative which has been designed for Zeiss operating microscopes and can be modified for other types of instruments is presented. No modification of the microscope is required. The instrument consists of two pieces: a metal bracket which mounts in the accessory shoe of the operating microscope, and a rigid plastic sterilizable portion which is readily attached on the microscope from below with three sterilizable screws. The entire assembly can be attached within 15 seconds. It affords complete sterility protection of the operative field; it provides a sterile surface for handling and moving the microscope; it does not interfere with accessory lighting systerms; and it allows open access to external accessories such as cameras and observer tubes.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信