{"title":"在医院设施中使用有源植入医疗设备(aimd)和极低频磁场的患者","authors":"L. Korpinen, R. Pääkkönen","doi":"10.23919/EMF-MED.2018.8526040","DOIUrl":null,"url":null,"abstract":"Europeans are aging, and they are increasingly using active implanted medical devices. We have conducted measurements of magnetic flux densities in hospitals in order to find possible risks to sensitive devices and thus to patient safety. The highest values 62 µT were found at electric facilities, but in treatment rooms, the values were less than 3 µT. Overall, the electromagnetic disturbance should be evaluated when planning hospitals.","PeriodicalId":134768,"journal":{"name":"2018 EMF-Med 1st World Conference on Biomedical Applications of Electromagnetic Fields (EMF-Med)","volume":"17 7","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2018-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Patients with active implanted medical devices (AIMDs) and extremely low-frequency magnetic fields at hospital facilities\",\"authors\":\"L. Korpinen, R. Pääkkönen\",\"doi\":\"10.23919/EMF-MED.2018.8526040\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Europeans are aging, and they are increasingly using active implanted medical devices. We have conducted measurements of magnetic flux densities in hospitals in order to find possible risks to sensitive devices and thus to patient safety. The highest values 62 µT were found at electric facilities, but in treatment rooms, the values were less than 3 µT. Overall, the electromagnetic disturbance should be evaluated when planning hospitals.\",\"PeriodicalId\":134768,\"journal\":{\"name\":\"2018 EMF-Med 1st World Conference on Biomedical Applications of Electromagnetic Fields (EMF-Med)\",\"volume\":\"17 7\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2018-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2018 EMF-Med 1st World Conference on Biomedical Applications of Electromagnetic Fields (EMF-Med)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.23919/EMF-MED.2018.8526040\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2018 EMF-Med 1st World Conference on Biomedical Applications of Electromagnetic Fields (EMF-Med)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.23919/EMF-MED.2018.8526040","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Patients with active implanted medical devices (AIMDs) and extremely low-frequency magnetic fields at hospital facilities
Europeans are aging, and they are increasingly using active implanted medical devices. We have conducted measurements of magnetic flux densities in hospitals in order to find possible risks to sensitive devices and thus to patient safety. The highest values 62 µT were found at electric facilities, but in treatment rooms, the values were less than 3 µT. Overall, the electromagnetic disturbance should be evaluated when planning hospitals.