基于传输衰减校正的新型HRRT高通量大鼠酒店的量化精度:模拟研究

S. Keller, Elina N. L'Estrade, B. Dall, M. Palner, M. Herth
{"title":"基于传输衰减校正的新型HRRT高通量大鼠酒店的量化精度:模拟研究","authors":"S. Keller, Elina N. L'Estrade, B. Dall, M. Palner, M. Herth","doi":"10.1109/NSSMIC.2016.8069467","DOIUrl":null,"url":null,"abstract":"Aim: To parallelize scanning and save time and cut costs of preclinical studies we have designed a new hotel holding 4 rats in the HRRT, which has a spatial resolution close to that of preclinical PET scanners. In this work we test the quantitative accuracy on phantoms in the hotel using different attenuation corrections methods on the HRRT. Material and Methods: The rat hotel has 4 compartments made of acrylic plastic with an 8 mm base plate and a 3 mm half-cylinder lid. Four 50 ml syringes filled with [18F]-FDG in water were used as phantoms and scanned in the rat hotel for 20 min. on the HRRT and a high statistics speed 10 transmission scan was acquired. Three μ-map processing/reconstruction methods — MAP-TR with either human head (HH) or water phantom (WP) prior and TXTV — were used and μ-maps and PET images reconstructed with each of the 3 μ-maps evaluated. Results: The μ-maps all underestimated the LAC of the acrylic plastic material as compared to CT, and the base plate thickness was underestimated. Activity concentrations were thus also underestimated: −4.6% using HH −8.7% using TXTV and −13.8% with WP. No noteworthy local variations were found. Conclusion: We found a global underestimation of PET activity, which was within a ±5% acceptance range using MAP-TR with the human head prior and a long transmission scan (speed 10). Fine tuning HH or TXTV parameters might give further improvements.","PeriodicalId":184587,"journal":{"name":"2016 IEEE Nuclear Science Symposium, Medical Imaging Conference and Room-Temperature Semiconductor Detector Workshop (NSS/MIC/RTSD)","volume":"117 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2016-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"10","resultStr":"{\"title\":\"Quantification accuracy of a new HRRT high throughput rat hotel using transmission-based attenuation correction: A phantom study\",\"authors\":\"S. Keller, Elina N. L'Estrade, B. Dall, M. Palner, M. Herth\",\"doi\":\"10.1109/NSSMIC.2016.8069467\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Aim: To parallelize scanning and save time and cut costs of preclinical studies we have designed a new hotel holding 4 rats in the HRRT, which has a spatial resolution close to that of preclinical PET scanners. In this work we test the quantitative accuracy on phantoms in the hotel using different attenuation corrections methods on the HRRT. Material and Methods: The rat hotel has 4 compartments made of acrylic plastic with an 8 mm base plate and a 3 mm half-cylinder lid. Four 50 ml syringes filled with [18F]-FDG in water were used as phantoms and scanned in the rat hotel for 20 min. on the HRRT and a high statistics speed 10 transmission scan was acquired. Three μ-map processing/reconstruction methods — MAP-TR with either human head (HH) or water phantom (WP) prior and TXTV — were used and μ-maps and PET images reconstructed with each of the 3 μ-maps evaluated. Results: The μ-maps all underestimated the LAC of the acrylic plastic material as compared to CT, and the base plate thickness was underestimated. Activity concentrations were thus also underestimated: −4.6% using HH −8.7% using TXTV and −13.8% with WP. No noteworthy local variations were found. Conclusion: We found a global underestimation of PET activity, which was within a ±5% acceptance range using MAP-TR with the human head prior and a long transmission scan (speed 10). Fine tuning HH or TXTV parameters might give further improvements.\",\"PeriodicalId\":184587,\"journal\":{\"name\":\"2016 IEEE Nuclear Science Symposium, Medical Imaging Conference and Room-Temperature Semiconductor Detector Workshop (NSS/MIC/RTSD)\",\"volume\":\"117 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2016-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"10\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2016 IEEE Nuclear Science Symposium, Medical Imaging Conference and Room-Temperature Semiconductor Detector Workshop (NSS/MIC/RTSD)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/NSSMIC.2016.8069467\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2016 IEEE Nuclear Science Symposium, Medical Imaging Conference and Room-Temperature Semiconductor Detector Workshop (NSS/MIC/RTSD)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/NSSMIC.2016.8069467","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 10

摘要

目的:为了并行扫描,节省临床前研究的时间和成本,我们在HRRT中设计了一个新的容纳4只大鼠的酒店,其空间分辨率接近临床前PET扫描仪。在这项工作中,我们使用不同的衰减校正方法在HRRT上测试了酒店中幽灵的定量精度。材料和方法:老鼠旅馆有4个隔间,由丙烯酸塑料制成,有一个8毫米的底板和一个3毫米的半圆柱形盖子。取4支50 ml注射器,充入[18F]-FDG水,作为模型,在大鼠宾馆HRRT上扫描20分钟,获得高统计速度10透射扫描。采用人头(HH)或水影(WP)先验的MAP-TR和TXTV三种μ图处理/重建方法,对每一种μ图重建的μ图和PET图像进行评价。结果:与CT相比,μ图均低估了丙烯酸塑料材料的LAC,基底厚度均低估。因此,活性浓度也被低估:HH组为- 4.6%,TXTV组为- 8.7%,WP组为- 13.8%。没有发现明显的局部变异。结论:我们发现PET活性的整体低估在±5%的接受范围内,使用MAP-TR与人类头部事先和长传输扫描(速度10)。微调HH或TXTV参数可能会带来进一步的改进。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Quantification accuracy of a new HRRT high throughput rat hotel using transmission-based attenuation correction: A phantom study
Aim: To parallelize scanning and save time and cut costs of preclinical studies we have designed a new hotel holding 4 rats in the HRRT, which has a spatial resolution close to that of preclinical PET scanners. In this work we test the quantitative accuracy on phantoms in the hotel using different attenuation corrections methods on the HRRT. Material and Methods: The rat hotel has 4 compartments made of acrylic plastic with an 8 mm base plate and a 3 mm half-cylinder lid. Four 50 ml syringes filled with [18F]-FDG in water were used as phantoms and scanned in the rat hotel for 20 min. on the HRRT and a high statistics speed 10 transmission scan was acquired. Three μ-map processing/reconstruction methods — MAP-TR with either human head (HH) or water phantom (WP) prior and TXTV — were used and μ-maps and PET images reconstructed with each of the 3 μ-maps evaluated. Results: The μ-maps all underestimated the LAC of the acrylic plastic material as compared to CT, and the base plate thickness was underestimated. Activity concentrations were thus also underestimated: −4.6% using HH −8.7% using TXTV and −13.8% with WP. No noteworthy local variations were found. Conclusion: We found a global underestimation of PET activity, which was within a ±5% acceptance range using MAP-TR with the human head prior and a long transmission scan (speed 10). Fine tuning HH or TXTV parameters might give further improvements.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信