注射异硫丹蓝染料后脉搏血氧计读数下降

Ji Kun Kim, Soon Yong Park, Myung Kook Lim, C. W. Lee, H. Kim, K. Kwun, Soo Jung Lee
{"title":"注射异硫丹蓝染料后脉搏血氧计读数下降","authors":"Ji Kun Kim, Soon Yong Park, Myung Kook Lim, C. W. Lee, H. Kim, K. Kwun, Soo Jung Lee","doi":"10.4048/JKBCS.2003.6.1.24","DOIUrl":null,"url":null,"abstract":"Purpose: A sentinel lymph node mapping with blue dye has been well accepted as a common procedure in breast cancer surgery. However, it is well known that blue dye absorbed into the circulation may interfere pulse oximetery reading. The aim of this study was to evaluate the change of pulse oximetery reading by isosulfan blue dye injection during sentinel lymph node mapping. Methods: Thirteen breast cancer patients with normal preoperative cardiopulmonary functions were studied. Four ml of isosulfan blue dye was injected subdermally when the patient became stable after induction of general anesthesia. The pulse oximetery was monitored continuously. Multiple arterial blood gas analyses (ABGA) were performed before dye injection and 10, 30, 40 minutes after dye injection. The results of oxygen saturturation by oximetery (SpO2) and the results of arterial oxygen tension (SaO2) and arterial oxygen saturation (SaO2) by ABGA were compared. Results: The value of both SaO2 and PaO2 measured by ABGA has not been altered by isosulfan dye injection. However SpO2 decreased by isosulfan dye injection. SpO2 decrease started 8.2±1.5 (2∼20) minutes after dye injection and returned to preinjection level by 85.7±5.6 (60∼126) minutes after injection. The lowest vaule of SpO2 was 95.6 ±1.2% (93∼97). Mean duration of SpO2 decrease was 77.5 ±6.2 (40∼117) minutes. The duration of SpO2 decrease was longer in the aged patients, but it was not statistically significant (p=0.3). There was no siginificant difference in duration of SpO2 decrease according to injection site, operation method, and body mass index (BMI). Conclusion: .Isosulfan dye injection using for sentinel lymph node mapping causes no change in true ABGA results but causes a mild reversible decrease in SpO2, It is important to look for other causes when SpO2 decrease is significant and persistent. (Journal of Korean Breast Cancer Society 2003;6:24-28)","PeriodicalId":414717,"journal":{"name":"Journal of Korean Breast Cancer Society","volume":"247 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2003-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Decrease in Pulse Oximeter Readings Following Injection of Isosulfan Blue Dye\",\"authors\":\"Ji Kun Kim, Soon Yong Park, Myung Kook Lim, C. W. Lee, H. Kim, K. Kwun, Soo Jung Lee\",\"doi\":\"10.4048/JKBCS.2003.6.1.24\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Purpose: A sentinel lymph node mapping with blue dye has been well accepted as a common procedure in breast cancer surgery. However, it is well known that blue dye absorbed into the circulation may interfere pulse oximetery reading. The aim of this study was to evaluate the change of pulse oximetery reading by isosulfan blue dye injection during sentinel lymph node mapping. Methods: Thirteen breast cancer patients with normal preoperative cardiopulmonary functions were studied. Four ml of isosulfan blue dye was injected subdermally when the patient became stable after induction of general anesthesia. The pulse oximetery was monitored continuously. Multiple arterial blood gas analyses (ABGA) were performed before dye injection and 10, 30, 40 minutes after dye injection. The results of oxygen saturturation by oximetery (SpO2) and the results of arterial oxygen tension (SaO2) and arterial oxygen saturation (SaO2) by ABGA were compared. Results: The value of both SaO2 and PaO2 measured by ABGA has not been altered by isosulfan dye injection. However SpO2 decreased by isosulfan dye injection. SpO2 decrease started 8.2±1.5 (2∼20) minutes after dye injection and returned to preinjection level by 85.7±5.6 (60∼126) minutes after injection. The lowest vaule of SpO2 was 95.6 ±1.2% (93∼97). Mean duration of SpO2 decrease was 77.5 ±6.2 (40∼117) minutes. The duration of SpO2 decrease was longer in the aged patients, but it was not statistically significant (p=0.3). There was no siginificant difference in duration of SpO2 decrease according to injection site, operation method, and body mass index (BMI). Conclusion: .Isosulfan dye injection using for sentinel lymph node mapping causes no change in true ABGA results but causes a mild reversible decrease in SpO2, It is important to look for other causes when SpO2 decrease is significant and persistent. (Journal of Korean Breast Cancer Society 2003;6:24-28)\",\"PeriodicalId\":414717,\"journal\":{\"name\":\"Journal of Korean Breast Cancer Society\",\"volume\":\"247 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2003-03-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Korean Breast Cancer Society\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4048/JKBCS.2003.6.1.24\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Korean Breast Cancer Society","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4048/JKBCS.2003.6.1.24","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:前哨淋巴结定位与蓝色染料已被广泛接受为乳腺癌手术的常见程序。然而,众所周知,被血液循环吸收的蓝色染料可能会干扰脉搏血氧计的读数。本研究的目的是评估异硫丹蓝染料注射在前哨淋巴结定位期间脉搏血氧读数的变化。方法:对13例术前心肺功能正常的乳腺癌患者进行分析。全麻诱导后病情稳定后皮下注射异硫丹蓝染料4ml。连续监测脉搏血氧。染色前及染色后10、30、40分钟行动脉血气分析(ABGA)。比较血氧计测定血氧饱和度(SpO2)和ABGA测定动脉血氧张力(SaO2)和动脉血氧饱和度(SaO2)的结果。结果:ABGA法测定的SaO2和PaO2值未受异硫丹染料注射的影响。注射异硫丹染料后,SpO2降低。SpO2在染料注射后8.2±1.5(2 ~ 20)分钟开始下降,在注射后85.7±5.6(60 ~ 126)分钟恢复到注射前水平。SpO2的最低值为95.6±1.2%(93 ~ 97)。SpO2降低的平均持续时间为77.5±6.2(40 ~ 117)分钟。老年患者SpO2下降持续时间更长,但差异无统计学意义(p=0.3)。注射部位、手术方式、体重指数(BMI)不同,SpO2降低持续时间无显著差异。结论:异硫丹染料注射用于前哨淋巴结测图对ABGA的真实结果没有影响,但引起SpO2的轻度可逆下降,当SpO2下降明显且持续时,寻找其他原因是很重要的。(韩国乳腺癌学会杂志2003;6:24-28)
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Decrease in Pulse Oximeter Readings Following Injection of Isosulfan Blue Dye
Purpose: A sentinel lymph node mapping with blue dye has been well accepted as a common procedure in breast cancer surgery. However, it is well known that blue dye absorbed into the circulation may interfere pulse oximetery reading. The aim of this study was to evaluate the change of pulse oximetery reading by isosulfan blue dye injection during sentinel lymph node mapping. Methods: Thirteen breast cancer patients with normal preoperative cardiopulmonary functions were studied. Four ml of isosulfan blue dye was injected subdermally when the patient became stable after induction of general anesthesia. The pulse oximetery was monitored continuously. Multiple arterial blood gas analyses (ABGA) were performed before dye injection and 10, 30, 40 minutes after dye injection. The results of oxygen saturturation by oximetery (SpO2) and the results of arterial oxygen tension (SaO2) and arterial oxygen saturation (SaO2) by ABGA were compared. Results: The value of both SaO2 and PaO2 measured by ABGA has not been altered by isosulfan dye injection. However SpO2 decreased by isosulfan dye injection. SpO2 decrease started 8.2±1.5 (2∼20) minutes after dye injection and returned to preinjection level by 85.7±5.6 (60∼126) minutes after injection. The lowest vaule of SpO2 was 95.6 ±1.2% (93∼97). Mean duration of SpO2 decrease was 77.5 ±6.2 (40∼117) minutes. The duration of SpO2 decrease was longer in the aged patients, but it was not statistically significant (p=0.3). There was no siginificant difference in duration of SpO2 decrease according to injection site, operation method, and body mass index (BMI). Conclusion: .Isosulfan dye injection using for sentinel lymph node mapping causes no change in true ABGA results but causes a mild reversible decrease in SpO2, It is important to look for other causes when SpO2 decrease is significant and persistent. (Journal of Korean Breast Cancer Society 2003;6:24-28)
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信