利用铋片的computed tomography fluoroscopy评估受术者辐射画质

Keisuke Takiguchi, Atsushi Urikura, Tsukasa Yoshida, Kenichi Hirosawa, Takahiro Ito, Yoshihiro Nakaya
{"title":"利用铋片的computed tomography fluoroscopy评估受术者辐射画质","authors":"Keisuke Takiguchi, Atsushi Urikura, Tsukasa Yoshida, Kenichi Hirosawa, Takahiro Ito, Yoshihiro Nakaya","doi":"10.6009/JJRT.2016_JSRT_72.11.1137","DOIUrl":null,"url":null,"abstract":"OBJECTIVE: The purpose of this study was to assess the dose reduction and the image quality using bismuth sheets during the computed tomography fluoroscopy (CTF). MATERIALS AND METHODS: The bismuth sheets of 1-mm thick were put on the upper mylar ring to reduce the frontal X-ray. The dose rates of an operator were measured using a torso phantom in the patient position during the CTF. The torso phantom was set on the gantry rotation center (center) and the lower position from the center (off-center). The image quality of the CTF image was assessed using an original phantom that mimics the normal liver parenchyma and the low attenuation lesions. The image contrast and contrast-to-noise ratio (CNR) were compared with and without the bismuth sheets. RESULTS: The bismuth sheets reduced the dose rate of the operator, regardless of whether the torso phantom was set at the center or the off-center. The reduction rate of exposure at the center and the off-center were 42.3% and 34.5%, respectively. There were no significant differences in the image contrast and the CNR, although the bismuth sheets increased the CT values of the liver parenchyma and the low attenuation lesions. CONCLUSION: The bismuth sheets were effective for the reduction of exposure to the operator without degrading the image quality of CTF images.","PeriodicalId":393304,"journal":{"name":"Nihon Hōshasen Gijutsu Gakkai zasshi","volume":"376 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1900-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"ビスマスシートを用いたcomputed tomography fluoroscopyにおける術者被ばくと画質の評価\",\"authors\":\"Keisuke Takiguchi, Atsushi Urikura, Tsukasa Yoshida, Kenichi Hirosawa, Takahiro Ito, Yoshihiro Nakaya\",\"doi\":\"10.6009/JJRT.2016_JSRT_72.11.1137\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"OBJECTIVE: The purpose of this study was to assess the dose reduction and the image quality using bismuth sheets during the computed tomography fluoroscopy (CTF). MATERIALS AND METHODS: The bismuth sheets of 1-mm thick were put on the upper mylar ring to reduce the frontal X-ray. The dose rates of an operator were measured using a torso phantom in the patient position during the CTF. The torso phantom was set on the gantry rotation center (center) and the lower position from the center (off-center). The image quality of the CTF image was assessed using an original phantom that mimics the normal liver parenchyma and the low attenuation lesions. The image contrast and contrast-to-noise ratio (CNR) were compared with and without the bismuth sheets. RESULTS: The bismuth sheets reduced the dose rate of the operator, regardless of whether the torso phantom was set at the center or the off-center. The reduction rate of exposure at the center and the off-center were 42.3% and 34.5%, respectively. There were no significant differences in the image contrast and the CNR, although the bismuth sheets increased the CT values of the liver parenchyma and the low attenuation lesions. CONCLUSION: The bismuth sheets were effective for the reduction of exposure to the operator without degrading the image quality of CTF images.\",\"PeriodicalId\":393304,\"journal\":{\"name\":\"Nihon Hōshasen Gijutsu Gakkai zasshi\",\"volume\":\"376 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1900-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Nihon Hōshasen Gijutsu Gakkai zasshi\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.6009/JJRT.2016_JSRT_72.11.1137\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hōshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/JJRT.2016_JSRT_72.11.1137","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:本研究的目的是评估在计算机断层透视(CTF)中使用铋片的剂量减少和图像质量。材料与方法:将1mm厚的铋片置于上密拉环上以减少正面x线。在CTF期间,操作员的剂量率是使用躯干幻像在患者位置测量的。躯干幻像设置在龙门旋转中心(center)和离中心较低的位置(off-center)。使用模拟正常肝实质和低衰减病变的原始模体评估CTF图像的图像质量。比较了添加和未添加铋片的图像对比度和噪比(CNR)。结果:无论躯干假体设置在中心还是偏离中心,铋片都降低了操作者的剂量率。中心处和非中心处的暴露降低率分别为42.3%和34.5%。虽然铋片增加了肝实质和低衰减病变的CT值,但图像对比度和CNR无显著差异。结论:铋片在不降低CTF图像质量的前提下,能有效减少操作者接触。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
ビスマスシートを用いたcomputed tomography fluoroscopyにおける術者被ばくと画質の評価
OBJECTIVE: The purpose of this study was to assess the dose reduction and the image quality using bismuth sheets during the computed tomography fluoroscopy (CTF). MATERIALS AND METHODS: The bismuth sheets of 1-mm thick were put on the upper mylar ring to reduce the frontal X-ray. The dose rates of an operator were measured using a torso phantom in the patient position during the CTF. The torso phantom was set on the gantry rotation center (center) and the lower position from the center (off-center). The image quality of the CTF image was assessed using an original phantom that mimics the normal liver parenchyma and the low attenuation lesions. The image contrast and contrast-to-noise ratio (CNR) were compared with and without the bismuth sheets. RESULTS: The bismuth sheets reduced the dose rate of the operator, regardless of whether the torso phantom was set at the center or the off-center. The reduction rate of exposure at the center and the off-center were 42.3% and 34.5%, respectively. There were no significant differences in the image contrast and the CNR, although the bismuth sheets increased the CT values of the liver parenchyma and the low attenuation lesions. CONCLUSION: The bismuth sheets were effective for the reduction of exposure to the operator without degrading the image quality of CTF images.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信