{"title":"高静压下锰电阻计的性能评价","authors":"T. Yuan, Yan-Chong Lu, Bao-Hui Yuan","doi":"10.1109/IMTC.1989.36923","DOIUrl":null,"url":null,"abstract":"The performance of manganin resistance gauges under high static pressure is evaluated. The accuracy of several interpolation- (or extrapolation-) based mathematical-models is compared. It is shown that if the gauges used at pressures below 1 GPa are calibrated at a given constant temperature and used at that temperature, the complete mathematical-model is unnecessary. However, a complete model for the variation of resistance with both pressure and temperature is required in the pressure range from 1 GPa up to 20 GPa, because the temperature changes due to compression or expansion of the pressure-transmitting medium can be quite pronounced.<<ETX>>","PeriodicalId":298343,"journal":{"name":"6th IEEE Conference Record., Instrumentation and Measurement Technology Conference","volume":"1 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1989-04-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Performance comments on manganin resistance gauges under high static pressures\",\"authors\":\"T. Yuan, Yan-Chong Lu, Bao-Hui Yuan\",\"doi\":\"10.1109/IMTC.1989.36923\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The performance of manganin resistance gauges under high static pressure is evaluated. The accuracy of several interpolation- (or extrapolation-) based mathematical-models is compared. It is shown that if the gauges used at pressures below 1 GPa are calibrated at a given constant temperature and used at that temperature, the complete mathematical-model is unnecessary. However, a complete model for the variation of resistance with both pressure and temperature is required in the pressure range from 1 GPa up to 20 GPa, because the temperature changes due to compression or expansion of the pressure-transmitting medium can be quite pronounced.<<ETX>>\",\"PeriodicalId\":298343,\"journal\":{\"name\":\"6th IEEE Conference Record., Instrumentation and Measurement Technology Conference\",\"volume\":\"1 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1989-04-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"6th IEEE Conference Record., Instrumentation and Measurement Technology Conference\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/IMTC.1989.36923\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"6th IEEE Conference Record., Instrumentation and Measurement Technology Conference","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/IMTC.1989.36923","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Performance comments on manganin resistance gauges under high static pressures
The performance of manganin resistance gauges under high static pressure is evaluated. The accuracy of several interpolation- (or extrapolation-) based mathematical-models is compared. It is shown that if the gauges used at pressures below 1 GPa are calibrated at a given constant temperature and used at that temperature, the complete mathematical-model is unnecessary. However, a complete model for the variation of resistance with both pressure and temperature is required in the pressure range from 1 GPa up to 20 GPa, because the temperature changes due to compression or expansion of the pressure-transmitting medium can be quite pronounced.<>