{"title":"三维全身PET单光子衰减校正的临床评价","authors":"C. Watson, A. Schaefer, W. K. Luk, C. Kirsch","doi":"10.1109/NSSMIC.1998.773867","DOIUrl":null,"url":null,"abstract":"Reports on the first clinical evaluation of the /sup 137/Cs single photon transmission measurement that has been implemented on the Siemens/CTI ECAT ART. Thorax phantom studies are performed to verify the accuracy of the transmission measurement. Pre-injection (with patient repositioning) and post-injection transmission scans and attenuation corrected emission images of 25 patients are compared with respect to estimated linear attenuation coefficients, specific activities and standardized uptake values in lesions. The authors find that both pre- and post-injection transmission provide accurate estimates of attenuation correction factors. Post-injection estimates for attenuation coefficients and specific activities in larger regional ROIs are nearly the same as pre-injection values, but exhibit 18% greater pixel standard deviation. Estimates of standardized uptake in smaller lesion ROIs are 13% higher with post-injection correction, but there is no difference in the ROI pixel variances.","PeriodicalId":129202,"journal":{"name":"1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255)","volume":"12 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1998-11-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"35","resultStr":"{\"title\":\"Clinical evaluation of single-photon attenuation correction for 3D whole-body PET\",\"authors\":\"C. Watson, A. Schaefer, W. K. Luk, C. Kirsch\",\"doi\":\"10.1109/NSSMIC.1998.773867\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Reports on the first clinical evaluation of the /sup 137/Cs single photon transmission measurement that has been implemented on the Siemens/CTI ECAT ART. Thorax phantom studies are performed to verify the accuracy of the transmission measurement. Pre-injection (with patient repositioning) and post-injection transmission scans and attenuation corrected emission images of 25 patients are compared with respect to estimated linear attenuation coefficients, specific activities and standardized uptake values in lesions. The authors find that both pre- and post-injection transmission provide accurate estimates of attenuation correction factors. Post-injection estimates for attenuation coefficients and specific activities in larger regional ROIs are nearly the same as pre-injection values, but exhibit 18% greater pixel standard deviation. Estimates of standardized uptake in smaller lesion ROIs are 13% higher with post-injection correction, but there is no difference in the ROI pixel variances.\",\"PeriodicalId\":129202,\"journal\":{\"name\":\"1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255)\",\"volume\":\"12 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1998-11-08\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"35\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/NSSMIC.1998.773867\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/NSSMIC.1998.773867","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 35

摘要

报道了在Siemens/CTI ECAT ART上实施的/sup 137/Cs单光子传输测量的首次临床评估。进行胸影研究以验证传输测量的准确性。比较25例患者注射前(患者重新定位)和注射后透射扫描和衰减校正发射图像的估计线性衰减系数、特定活动和标准化摄取值。作者发现,注入前和注入后的传输都提供了衰减校正因子的准确估计。在较大区域roi中,注入后的衰减系数和比活度估价值与注入前的值几乎相同,但显示出18%的像素标准差。注射后校正后,较小病灶ROI的标准化摄取估计值高出13%,但ROI像素方差没有差异。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Clinical evaluation of single-photon attenuation correction for 3D whole-body PET
Reports on the first clinical evaluation of the /sup 137/Cs single photon transmission measurement that has been implemented on the Siemens/CTI ECAT ART. Thorax phantom studies are performed to verify the accuracy of the transmission measurement. Pre-injection (with patient repositioning) and post-injection transmission scans and attenuation corrected emission images of 25 patients are compared with respect to estimated linear attenuation coefficients, specific activities and standardized uptake values in lesions. The authors find that both pre- and post-injection transmission provide accurate estimates of attenuation correction factors. Post-injection estimates for attenuation coefficients and specific activities in larger regional ROIs are nearly the same as pre-injection values, but exhibit 18% greater pixel standard deviation. Estimates of standardized uptake in smaller lesion ROIs are 13% higher with post-injection correction, but there is no difference in the ROI pixel variances.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信