{"title":"三维全身PET单光子衰减校正的临床评价","authors":"C. Watson, A. Schaefer, W. K. Luk, C. Kirsch","doi":"10.1109/NSSMIC.1998.773867","DOIUrl":null,"url":null,"abstract":"Reports on the first clinical evaluation of the /sup 137/Cs single photon transmission measurement that has been implemented on the Siemens/CTI ECAT ART. Thorax phantom studies are performed to verify the accuracy of the transmission measurement. Pre-injection (with patient repositioning) and post-injection transmission scans and attenuation corrected emission images of 25 patients are compared with respect to estimated linear attenuation coefficients, specific activities and standardized uptake values in lesions. The authors find that both pre- and post-injection transmission provide accurate estimates of attenuation correction factors. Post-injection estimates for attenuation coefficients and specific activities in larger regional ROIs are nearly the same as pre-injection values, but exhibit 18% greater pixel standard deviation. Estimates of standardized uptake in smaller lesion ROIs are 13% higher with post-injection correction, but there is no difference in the ROI pixel variances.","PeriodicalId":129202,"journal":{"name":"1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255)","volume":"12 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1998-11-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"35","resultStr":"{\"title\":\"Clinical evaluation of single-photon attenuation correction for 3D whole-body PET\",\"authors\":\"C. Watson, A. Schaefer, W. K. Luk, C. Kirsch\",\"doi\":\"10.1109/NSSMIC.1998.773867\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Reports on the first clinical evaluation of the /sup 137/Cs single photon transmission measurement that has been implemented on the Siemens/CTI ECAT ART. Thorax phantom studies are performed to verify the accuracy of the transmission measurement. Pre-injection (with patient repositioning) and post-injection transmission scans and attenuation corrected emission images of 25 patients are compared with respect to estimated linear attenuation coefficients, specific activities and standardized uptake values in lesions. The authors find that both pre- and post-injection transmission provide accurate estimates of attenuation correction factors. Post-injection estimates for attenuation coefficients and specific activities in larger regional ROIs are nearly the same as pre-injection values, but exhibit 18% greater pixel standard deviation. Estimates of standardized uptake in smaller lesion ROIs are 13% higher with post-injection correction, but there is no difference in the ROI pixel variances.\",\"PeriodicalId\":129202,\"journal\":{\"name\":\"1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255)\",\"volume\":\"12 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1998-11-08\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"35\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/NSSMIC.1998.773867\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/NSSMIC.1998.773867","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Clinical evaluation of single-photon attenuation correction for 3D whole-body PET
Reports on the first clinical evaluation of the /sup 137/Cs single photon transmission measurement that has been implemented on the Siemens/CTI ECAT ART. Thorax phantom studies are performed to verify the accuracy of the transmission measurement. Pre-injection (with patient repositioning) and post-injection transmission scans and attenuation corrected emission images of 25 patients are compared with respect to estimated linear attenuation coefficients, specific activities and standardized uptake values in lesions. The authors find that both pre- and post-injection transmission provide accurate estimates of attenuation correction factors. Post-injection estimates for attenuation coefficients and specific activities in larger regional ROIs are nearly the same as pre-injection values, but exhibit 18% greater pixel standard deviation. Estimates of standardized uptake in smaller lesion ROIs are 13% higher with post-injection correction, but there is no difference in the ROI pixel variances.