探讨使用自动体外除颤器测量胸腔阻抗定量循环的潜力

M. Risdal, H. Losert, J. Nysaether, T. Eftestoel, F. Sterz, S. O. Aase
{"title":"探讨使用自动体外除颤器测量胸腔阻抗定量循环的潜力","authors":"M. Risdal, H. Losert, J. Nysaether, T. Eftestoel, F. Sterz, S. O. Aase","doi":"10.1109/CIC.2005.1588228","DOIUrl":null,"url":null,"abstract":"It has been suggested to acquire circulatory information from patients undergoing resuscitation from cardiac arrest by analyzing their thoracic electrical impedance using modified automated external defibrillators (AEDs). To investigate the potential of this idea, we studied the correlation between two impedance-derived parameters related to circulation, the negative peak of the impedance fluctuation (Zpeak) and its first time derivative (dZpeak), and arterial blood pressure measurements in 26 patients undergoing resuscitation and 32 hemodynamically stable patients. The highest correlation coefficient, rho=0.4338 was found between the systolic blood pressure and the magnitude of the negative peak of the first time derivative of the impedance. The poor correlation indicates that the impedance-derived parameters are not suitable for quantification of circulation, but can be used to indicate circulation","PeriodicalId":239491,"journal":{"name":"Computers in Cardiology, 2005","volume":"10 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1900-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"4","resultStr":"{\"title\":\"Examining the potential of using thorax impedance measured by automated external defibrillators for quantification of circulation\",\"authors\":\"M. Risdal, H. Losert, J. Nysaether, T. Eftestoel, F. Sterz, S. O. Aase\",\"doi\":\"10.1109/CIC.2005.1588228\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"It has been suggested to acquire circulatory information from patients undergoing resuscitation from cardiac arrest by analyzing their thoracic electrical impedance using modified automated external defibrillators (AEDs). To investigate the potential of this idea, we studied the correlation between two impedance-derived parameters related to circulation, the negative peak of the impedance fluctuation (Zpeak) and its first time derivative (dZpeak), and arterial blood pressure measurements in 26 patients undergoing resuscitation and 32 hemodynamically stable patients. The highest correlation coefficient, rho=0.4338 was found between the systolic blood pressure and the magnitude of the negative peak of the first time derivative of the impedance. The poor correlation indicates that the impedance-derived parameters are not suitable for quantification of circulation, but can be used to indicate circulation\",\"PeriodicalId\":239491,\"journal\":{\"name\":\"Computers in Cardiology, 2005\",\"volume\":\"10 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1900-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"4\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Computers in Cardiology, 2005\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/CIC.2005.1588228\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Computers in Cardiology, 2005","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/CIC.2005.1588228","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 4

摘要

有人建议使用改良的自动体外除颤器(aed)分析患者的胸电阻抗,以获取心脏骤停复苏患者的循环信息。为了研究这一想法的潜力,我们研究了与循环相关的两个阻抗衍生参数,阻抗波动的负峰值(Zpeak)及其第一次导数(dZpeak)与26例复苏患者和32例血流动力学稳定患者的动脉血压测量之间的相关性。收缩压与阻抗一阶导数负峰幅度的相关系数最高,rho=0.4338。相关性较差表明阻抗导出的参数不适合用于循环的量化,但可以用来指示循环
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Examining the potential of using thorax impedance measured by automated external defibrillators for quantification of circulation
It has been suggested to acquire circulatory information from patients undergoing resuscitation from cardiac arrest by analyzing their thoracic electrical impedance using modified automated external defibrillators (AEDs). To investigate the potential of this idea, we studied the correlation between two impedance-derived parameters related to circulation, the negative peak of the impedance fluctuation (Zpeak) and its first time derivative (dZpeak), and arterial blood pressure measurements in 26 patients undergoing resuscitation and 32 hemodynamically stable patients. The highest correlation coefficient, rho=0.4338 was found between the systolic blood pressure and the magnitude of the negative peak of the first time derivative of the impedance. The poor correlation indicates that the impedance-derived parameters are not suitable for quantification of circulation, but can be used to indicate circulation
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信