{"title":"用于糖化血红蛋白检测的基于图像的定量荧光免疫分析阅读器:校准和可重复性研究","authors":"K. Chowdhury, J. Joseph, Nethra Reddy, Jayaraman Kiruthi Vasan, M. Sivaprakasam","doi":"10.1109/MeMeA.2016.7533702","DOIUrl":null,"url":null,"abstract":"Image based readers for quantitative fluorescence immunoassays offer advantages of reliability and robustness over traditional scanning readers and has potential applications in point-of-care diagnostics. We had previously demonstrated an image based quantitative method to scan fluorescence labeled lateral flow test strips. Here we present an improved design, and report the calibration method and repeatability of the instrument. A new wideband emission filter was used instead of the previous narrowband filter and enhanced signal levels were obtained. Standard HbA1c calibrator set was used to evaluate the area ratios for various concentration levels of the analyte and generate the calibration curve. A straight line fit to the calibration data showed R-squared value to be 0.99. Repeatability of the instrument in successive test runs was evaluated to be fairly good with CV less than 4%. Inter-day repeatability of the instrument was also studied and the results were moderate.","PeriodicalId":221120,"journal":{"name":"2016 IEEE International Symposium on Medical Measurements and Applications (MeMeA)","volume":"60 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2016-05-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"An image based quantitative fluorescence immunoassay reader for HbA1c testing: Calibration & repeatability study\",\"authors\":\"K. Chowdhury, J. Joseph, Nethra Reddy, Jayaraman Kiruthi Vasan, M. Sivaprakasam\",\"doi\":\"10.1109/MeMeA.2016.7533702\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Image based readers for quantitative fluorescence immunoassays offer advantages of reliability and robustness over traditional scanning readers and has potential applications in point-of-care diagnostics. We had previously demonstrated an image based quantitative method to scan fluorescence labeled lateral flow test strips. Here we present an improved design, and report the calibration method and repeatability of the instrument. A new wideband emission filter was used instead of the previous narrowband filter and enhanced signal levels were obtained. Standard HbA1c calibrator set was used to evaluate the area ratios for various concentration levels of the analyte and generate the calibration curve. A straight line fit to the calibration data showed R-squared value to be 0.99. Repeatability of the instrument in successive test runs was evaluated to be fairly good with CV less than 4%. Inter-day repeatability of the instrument was also studied and the results were moderate.\",\"PeriodicalId\":221120,\"journal\":{\"name\":\"2016 IEEE International Symposium on Medical Measurements and Applications (MeMeA)\",\"volume\":\"60 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2016-05-15\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"2016 IEEE International Symposium on Medical Measurements and Applications (MeMeA)\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1109/MeMeA.2016.7533702\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"2016 IEEE International Symposium on Medical Measurements and Applications (MeMeA)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/MeMeA.2016.7533702","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

摘要

基于图像的定量荧光免疫分析阅读器提供了可靠性和鲁棒性优于传统的扫描阅读器,并有潜在的应用在点护理诊断。我们之前已经演示了一种基于图像的定量方法来扫描荧光标记的横向流动测试条。本文提出了一种改进的设计,并报告了仪器的校准方法和可重复性。采用一种新的宽带发射滤波器代替原有的窄带滤波器,提高了信号电平。使用标准的糖化血红蛋白校准器对不同浓度水平的分析物进行面积比评估,并生成校准曲线。对标定数据进行直线拟合,r平方值为0.99。在连续的试验运行中,仪器的重复性被评价为相当好,CV小于4%。还研究了仪器的日间重复性,结果是中等的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
An image based quantitative fluorescence immunoassay reader for HbA1c testing: Calibration & repeatability study
Image based readers for quantitative fluorescence immunoassays offer advantages of reliability and robustness over traditional scanning readers and has potential applications in point-of-care diagnostics. We had previously demonstrated an image based quantitative method to scan fluorescence labeled lateral flow test strips. Here we present an improved design, and report the calibration method and repeatability of the instrument. A new wideband emission filter was used instead of the previous narrowband filter and enhanced signal levels were obtained. Standard HbA1c calibrator set was used to evaluate the area ratios for various concentration levels of the analyte and generate the calibration curve. A straight line fit to the calibration data showed R-squared value to be 0.99. Repeatability of the instrument in successive test runs was evaluated to be fairly good with CV less than 4%. Inter-day repeatability of the instrument was also studied and the results were moderate.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信