用带有适配器的国际标准定义的鼓室温度计用黑体源前额辐射温度计的实验室精度评价。

Q3 Engineering
Li Wang, Choon Heng Goh
{"title":"用带有适配器的国际标准定义的鼓室温度计用黑体源前额辐射温度计的实验室精度评价。","authors":"Li Wang,&nbsp;Choon Heng Goh","doi":"10.1080/03091902.2022.2096135","DOIUrl":null,"url":null,"abstract":"<p><p>In view of the gaps between the current clinical thermometer standards and the commercially available products, in particular the introduction of forehead radiation thermometers into the market, ISO initiated a review on standard for clinical thermometers. The original clinical thermometer standard includes requirements for evaluating tympanic (ear) thermometers using identified reference blackbody sources. However, one of the challenges in reviewing the standard is to understand if the reference blackbody sources identified for ear thermometers are still usable for evaluating the laboratory accuracy of forehead thermometers. In order to answer this challenge, four brand new forehead thermometers from two different manufacturers are evaluated at NMC. Two reference blackbody sources, one JIS standard source with very small opening and one NMC source with larger opening, are used for the evaluation. In order to use JIS standard source, two types of adaptors, one supplied by one of the manufactures, one 3D printed, are evaluated. The measurement results together with their measurement uncertainties are presented and analysed. Suitable adaptors are identified and recommended.</p>","PeriodicalId":39637,"journal":{"name":"Journal of Medical Engineering and Technology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Evaluation of laboratory accuracy of forehead radiation thermometers using blackbody sources for tympanic thermometers defined by international standards with adaptors.\",\"authors\":\"Li Wang,&nbsp;Choon Heng Goh\",\"doi\":\"10.1080/03091902.2022.2096135\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>In view of the gaps between the current clinical thermometer standards and the commercially available products, in particular the introduction of forehead radiation thermometers into the market, ISO initiated a review on standard for clinical thermometers. The original clinical thermometer standard includes requirements for evaluating tympanic (ear) thermometers using identified reference blackbody sources. However, one of the challenges in reviewing the standard is to understand if the reference blackbody sources identified for ear thermometers are still usable for evaluating the laboratory accuracy of forehead thermometers. In order to answer this challenge, four brand new forehead thermometers from two different manufacturers are evaluated at NMC. Two reference blackbody sources, one JIS standard source with very small opening and one NMC source with larger opening, are used for the evaluation. In order to use JIS standard source, two types of adaptors, one supplied by one of the manufactures, one 3D printed, are evaluated. The measurement results together with their measurement uncertainties are presented and analysed. Suitable adaptors are identified and recommended.</p>\",\"PeriodicalId\":39637,\"journal\":{\"name\":\"Journal of Medical Engineering and Technology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Engineering and Technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1080/03091902.2022.2096135\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Engineering\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Engineering and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1080/03091902.2022.2096135","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Engineering","Score":null,"Total":0}
引用次数: 0

摘要

鉴于现时的临床体温计标准与市售产品之间存在差距,特别是额头辐射体温计进入市场后,ISO开始检讨临床体温计的标准。最初的临床温度计标准包括使用确定的参考黑体源评估鼓室(耳)温度计的要求。然而,审查该标准的挑战之一是了解耳部温度计的参考黑体源是否仍然可用于评估前额温度计的实验室精度。为了应对这一挑战,NMC对来自两家不同制造商的四款全新额头温度计进行了评估。两个参考黑体源,一个JIS标准源具有非常小的开口和一个NMC源具有较大的开口,用于评估。为了使用JIS标准源,对两种类型的适配器进行了评估,一种是由一家制造商提供的,一种是3D打印的。给出了测量结果及其测量不确定度,并进行了分析。确定并推荐合适的适配器。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Evaluation of laboratory accuracy of forehead radiation thermometers using blackbody sources for tympanic thermometers defined by international standards with adaptors.

In view of the gaps between the current clinical thermometer standards and the commercially available products, in particular the introduction of forehead radiation thermometers into the market, ISO initiated a review on standard for clinical thermometers. The original clinical thermometer standard includes requirements for evaluating tympanic (ear) thermometers using identified reference blackbody sources. However, one of the challenges in reviewing the standard is to understand if the reference blackbody sources identified for ear thermometers are still usable for evaluating the laboratory accuracy of forehead thermometers. In order to answer this challenge, four brand new forehead thermometers from two different manufacturers are evaluated at NMC. Two reference blackbody sources, one JIS standard source with very small opening and one NMC source with larger opening, are used for the evaluation. In order to use JIS standard source, two types of adaptors, one supplied by one of the manufactures, one 3D printed, are evaluated. The measurement results together with their measurement uncertainties are presented and analysed. Suitable adaptors are identified and recommended.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Medical Engineering and Technology
Journal of Medical Engineering and Technology Engineering-Biomedical Engineering
CiteScore
4.60
自引率
0.00%
发文量
77
期刊介绍: The Journal of Medical Engineering & Technology is an international, independent, multidisciplinary, bimonthly journal promoting an understanding of the physiological processes underlying disease processes and the appropriate application of technology. Features include authoritative review papers, the reporting of original research, and evaluation reports on new and existing techniques and devices. Each issue of the journal contains a comprehensive information service which provides news relevant to the world of medical technology, details of new products, book reviews, and selected contents of related journals.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信