{"title":"术中CT评估DBS记录微电极插入时的偏差。","authors":"Nicolas Massager,&nbsp;Anthony Nguyen,&nbsp;Henri-Benjamin Pouleau,&nbsp;Sophie Dethy,&nbsp;Daniele Morelli","doi":"10.1159/000528202","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Intraoperative microelectrodes recording with the Ben Gun microdrive system are often used during DBS surgery. An accurate location of these microelectrodes will directly influence the interest of this recording. We have studied the imprecision of implantation of these microelectrodes.</p><p><strong>Methods: </strong>We have analyzed the stereotactic position of 135 microelectrodes implanted with the Ben Gun microdrive during DBS surgery of 16 patients with advanced Parkinson's disease. An intracranial CT was obtained and integrated to a stereotactic planification system. We recorded the stereotactic coordinates of the 5 microelectrodes inserted simultaneously in a cross-shape. The coordinates of each microelectrode were compared with coordinates of the other 4 electrodes inserted simultaneously with the Ben Gun and visible on the same iCT image. Thus, this procedure avoids errors from image fusion and from brain shift. We calculate (1) the three-dimensional Euclidian deviation of microelectrodes, (2) the deviation in X- and Y-axes on reconstructed probe's eye view MR images, and (3) the deviation from the 2-mm theoretical distance between the central electrode and 4 satellite microelectrodes.</p><p><strong>Results: </strong>The median deviation was 0.64 mm in 3-D and 0.58 mm in 2-D probe's eye view. Satellite electrodes were located from the central electrode theoretically at 2.0 mm and practically within the range 1.9-2.1 mm, 1.5-2.5 mm, 1.0-3.0 mm, and 0.5-3.5 mm for, respectively, 9.3%, 53.7%, 88.0%, and 98.1%, thus highlighting the significant deviation from the theoretical distance. Position imprecisions were similar for the 4 satellite microelectrodes. The imprecision was similar in X-axis and Y-axes and statistically less in Z-axis. For bilateral implantation, the second implantation of the same patient was not associated with a greater risk of deviation of the microelectrodes than for the first side implanted.</p><p><strong>Conclusion: </strong>A significant percentage of microelectrodes for MER can deviate substantially from their theoretical target during DBS procedures. An iCT can be used to estimate the potential deviation of microelectrodes and improve the interpretation of MER during the procedure.</p>","PeriodicalId":1,"journal":{"name":"Accounts of Chemical Research","volume":null,"pages":null},"PeriodicalIF":16.4000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":"{\"title\":\"Deviation of DBS Recording Microelectrodes during Insertion Assessed by Intraoperative CT.\",\"authors\":\"Nicolas Massager,&nbsp;Anthony Nguyen,&nbsp;Henri-Benjamin Pouleau,&nbsp;Sophie Dethy,&nbsp;Daniele Morelli\",\"doi\":\"10.1159/000528202\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>Intraoperative microelectrodes recording with the Ben Gun microdrive system are often used during DBS surgery. An accurate location of these microelectrodes will directly influence the interest of this recording. We have studied the imprecision of implantation of these microelectrodes.</p><p><strong>Methods: </strong>We have analyzed the stereotactic position of 135 microelectrodes implanted with the Ben Gun microdrive during DBS surgery of 16 patients with advanced Parkinson's disease. An intracranial CT was obtained and integrated to a stereotactic planification system. We recorded the stereotactic coordinates of the 5 microelectrodes inserted simultaneously in a cross-shape. The coordinates of each microelectrode were compared with coordinates of the other 4 electrodes inserted simultaneously with the Ben Gun and visible on the same iCT image. Thus, this procedure avoids errors from image fusion and from brain shift. We calculate (1) the three-dimensional Euclidian deviation of microelectrodes, (2) the deviation in X- and Y-axes on reconstructed probe's eye view MR images, and (3) the deviation from the 2-mm theoretical distance between the central electrode and 4 satellite microelectrodes.</p><p><strong>Results: </strong>The median deviation was 0.64 mm in 3-D and 0.58 mm in 2-D probe's eye view. Satellite electrodes were located from the central electrode theoretically at 2.0 mm and practically within the range 1.9-2.1 mm, 1.5-2.5 mm, 1.0-3.0 mm, and 0.5-3.5 mm for, respectively, 9.3%, 53.7%, 88.0%, and 98.1%, thus highlighting the significant deviation from the theoretical distance. Position imprecisions were similar for the 4 satellite microelectrodes. The imprecision was similar in X-axis and Y-axes and statistically less in Z-axis. For bilateral implantation, the second implantation of the same patient was not associated with a greater risk of deviation of the microelectrodes than for the first side implanted.</p><p><strong>Conclusion: </strong>A significant percentage of microelectrodes for MER can deviate substantially from their theoretical target during DBS procedures. An iCT can be used to estimate the potential deviation of microelectrodes and improve the interpretation of MER during the procedure.</p>\",\"PeriodicalId\":1,\"journal\":{\"name\":\"Accounts of Chemical Research\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":16.4000,\"publicationDate\":\"2023-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Accounts of Chemical Research\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1159/000528202\",\"RegionNum\":1,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"CHEMISTRY, MULTIDISCIPLINARY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Accounts of Chemical Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1159/000528202","RegionNum":1,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
引用次数: 1

摘要

术中微电极记录与Ben Gun微驱动系统在DBS手术中经常被使用。这些微电极的准确位置将直接影响记录的兴趣。我们研究了这些微电极植入的不精确性。方法:对16例晚期帕金森病患者DBS手术中植入Ben Gun微驱动器的135个微电极的立体定向位置进行分析。获得颅内CT并整合到立体定向平面化系统。我们记录了同时插入的5个微电极呈十字形的立体定向坐标。将每个微电极的坐标与与Ben Gun同时插入的其他4个电极的坐标进行比较,并在同一iCT图像上可见。因此,该方法避免了图像融合和脑偏移带来的错误。我们计算了(1)微电极的三维欧几里得偏差,(2)重建探头眼视MR图像的X轴和y轴偏差,以及(3)中心电极与4个卫星微电极之间2mm理论距离的偏差。结果:三维位差为0.64 mm,二维位差为0.58 mm。卫星电极距离中心电极的理论距离为2.0 mm,实际距离分别为1.9 ~ 2.1 mm、1.5 ~ 2.5 mm、1.0 ~ 3.0 mm和0.5 ~ 3.5 mm,分别为9.3%、53.7%、88.0%和98.1%,与理论距离偏差较大。4个卫星微电极的位置误差相似。不精度在x轴和y轴上相似,在z轴上统计上较小。对于双侧植入,同一患者的第二次植入与第一次植入相比,微电极偏离的风险并不大。结论:在DBS过程中,相当大比例的MER微电极可能会严重偏离其理论目标。iCT可用于估计微电极的潜在偏差,并在过程中改善对MER的解释。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Deviation of DBS Recording Microelectrodes during Insertion Assessed by Intraoperative CT.

Introduction: Intraoperative microelectrodes recording with the Ben Gun microdrive system are often used during DBS surgery. An accurate location of these microelectrodes will directly influence the interest of this recording. We have studied the imprecision of implantation of these microelectrodes.

Methods: We have analyzed the stereotactic position of 135 microelectrodes implanted with the Ben Gun microdrive during DBS surgery of 16 patients with advanced Parkinson's disease. An intracranial CT was obtained and integrated to a stereotactic planification system. We recorded the stereotactic coordinates of the 5 microelectrodes inserted simultaneously in a cross-shape. The coordinates of each microelectrode were compared with coordinates of the other 4 electrodes inserted simultaneously with the Ben Gun and visible on the same iCT image. Thus, this procedure avoids errors from image fusion and from brain shift. We calculate (1) the three-dimensional Euclidian deviation of microelectrodes, (2) the deviation in X- and Y-axes on reconstructed probe's eye view MR images, and (3) the deviation from the 2-mm theoretical distance between the central electrode and 4 satellite microelectrodes.

Results: The median deviation was 0.64 mm in 3-D and 0.58 mm in 2-D probe's eye view. Satellite electrodes were located from the central electrode theoretically at 2.0 mm and practically within the range 1.9-2.1 mm, 1.5-2.5 mm, 1.0-3.0 mm, and 0.5-3.5 mm for, respectively, 9.3%, 53.7%, 88.0%, and 98.1%, thus highlighting the significant deviation from the theoretical distance. Position imprecisions were similar for the 4 satellite microelectrodes. The imprecision was similar in X-axis and Y-axes and statistically less in Z-axis. For bilateral implantation, the second implantation of the same patient was not associated with a greater risk of deviation of the microelectrodes than for the first side implanted.

Conclusion: A significant percentage of microelectrodes for MER can deviate substantially from their theoretical target during DBS procedures. An iCT can be used to estimate the potential deviation of microelectrodes and improve the interpretation of MER during the procedure.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Accounts of Chemical Research
Accounts of Chemical Research 化学-化学综合
CiteScore
31.40
自引率
1.10%
发文量
312
审稿时长
2 months
期刊介绍: Accounts of Chemical Research presents short, concise and critical articles offering easy-to-read overviews of basic research and applications in all areas of chemistry and biochemistry. These short reviews focus on research from the author’s own laboratory and are designed to teach the reader about a research project. In addition, Accounts of Chemical Research publishes commentaries that give an informed opinion on a current research problem. Special Issues online are devoted to a single topic of unusual activity and significance. Accounts of Chemical Research replaces the traditional article abstract with an article "Conspectus." These entries synopsize the research affording the reader a closer look at the content and significance of an article. Through this provision of a more detailed description of the article contents, the Conspectus enhances the article's discoverability by search engines and the exposure for the research.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信