Chen Sun, Zhenghua Huang, Yingxue Ma, Ye Gu, Qi Li
{"title":"[Cochlear electrode array misplacement into the superior semicircular canal: a case report and literature review].","authors":"Chen Sun, Zhenghua Huang, Yingxue Ma, Ye Gu, Qi Li","doi":"10.13201/j.issn.2096-7993.2023.04.015","DOIUrl":null,"url":null,"abstract":"<p><p><b></b> Electrode array misplacement is a rare complication of cochlear implant. This article reports an 11-year-old boy who was mistakenly implanted the cochlear electrode array into the superior semicircular canal during the initial cochlear implant. After the diagnosis was confirmed, he underwent a second cochlear implant and the electrode array were successfully implanted into the cochlea. This article conducted a systematic review of the literature on electrode array misplacement, and the causes of electrode array misplacement were analyzed from different implantation position.</p>","PeriodicalId":18104,"journal":{"name":"Lin chuang er bi yan hou tou jing wai ke za zhi = Journal of clinical otorhinolaryngology, head, and neck surgery","volume":"37 4","pages":"310-312"},"PeriodicalIF":0.0000,"publicationDate":"2023-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10406583/pdf/lcebyhtjwkzz-37-4-310.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Lin chuang er bi yan hou tou jing wai ke za zhi = Journal of clinical otorhinolaryngology, head, and neck surgery","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.13201/j.issn.2096-7993.2023.04.015","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Electrode array misplacement is a rare complication of cochlear implant. This article reports an 11-year-old boy who was mistakenly implanted the cochlear electrode array into the superior semicircular canal during the initial cochlear implant. After the diagnosis was confirmed, he underwent a second cochlear implant and the electrode array were successfully implanted into the cochlea. This article conducted a systematic review of the literature on electrode array misplacement, and the causes of electrode array misplacement were analyzed from different implantation position.