Evaluation of Visual Outcomes and 3-Month Refractive Stability of a New Hydrophobic Acrylic Intraocular Lens.

IF 1.8 Q3 OPHTHALMOLOGY
Clayton Blehm, Brad Hall
{"title":"Evaluation of Visual Outcomes and 3-Month Refractive Stability of a New Hydrophobic Acrylic Intraocular Lens.","authors":"Clayton Blehm,&nbsp;Brad Hall","doi":"10.2147/OPTH.S415400","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To determine the refractive stability of a new hydrophobic acrylic intraocular lens (IOL) when implanted bilaterally.</p><p><strong>Methods: </strong>This was a prospective, evaluator masked, single surgeon study of 58 eyes of 29 patients. Patients were bilaterally implanted with the Clareon monofocal IOL (CNA0T0, Alcon Vision LLC). Refractive stability was evaluated between 1 and 3 months postoperatively. At 3 months postoperatively, data were also collected for binocular uncorrected and distance corrected visual acuities at distance (4 m) and intermediate (80 cm and 66 cm) and binocular defocus curve.</p><p><strong>Results: </strong>Postoperative refraction was statistically equivalent between 1 and 3 months postoperatively (p < 0.001). Mean postoperative uncorrected distance visual acuity was -0.01 ± 0.10 logMAR, and mean corrected distance visual acuity was -0.04 ± 0.06 logMAR. Mean postoperative uncorrected intermediate visual acuity was 0.16 ± 0.13 logMAR and 0.24 ± 0.14 logMAR at 80 cm and 66 cm, respectively. With distance correction in place, mean visual acuity at 80cm and 60cm was 0.16 ± 0.13 logMAR and 0.23 ± 0.14 logMAR, respectively.</p><p><strong>Conclusion: </strong>The Clareon monofocal IOL can provide stable refraction, excellent distance vision, and functional intermediate vision postoperatively.</p>","PeriodicalId":10442,"journal":{"name":"Clinical ophthalmology","volume":"17 ","pages":"1859-1864"},"PeriodicalIF":1.8000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/ba/d8/opth-17-1859.PMC10328104.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical ophthalmology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2147/OPTH.S415400","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To determine the refractive stability of a new hydrophobic acrylic intraocular lens (IOL) when implanted bilaterally.

Methods: This was a prospective, evaluator masked, single surgeon study of 58 eyes of 29 patients. Patients were bilaterally implanted with the Clareon monofocal IOL (CNA0T0, Alcon Vision LLC). Refractive stability was evaluated between 1 and 3 months postoperatively. At 3 months postoperatively, data were also collected for binocular uncorrected and distance corrected visual acuities at distance (4 m) and intermediate (80 cm and 66 cm) and binocular defocus curve.

Results: Postoperative refraction was statistically equivalent between 1 and 3 months postoperatively (p < 0.001). Mean postoperative uncorrected distance visual acuity was -0.01 ± 0.10 logMAR, and mean corrected distance visual acuity was -0.04 ± 0.06 logMAR. Mean postoperative uncorrected intermediate visual acuity was 0.16 ± 0.13 logMAR and 0.24 ± 0.14 logMAR at 80 cm and 66 cm, respectively. With distance correction in place, mean visual acuity at 80cm and 60cm was 0.16 ± 0.13 logMAR and 0.23 ± 0.14 logMAR, respectively.

Conclusion: The Clareon monofocal IOL can provide stable refraction, excellent distance vision, and functional intermediate vision postoperatively.

Abstract Image

Abstract Image

新型疏水丙烯酸人工晶状体的视力及3个月屈光稳定性评价。
目的:观察双侧植入新型疏水丙烯酸人工晶状体的屈光稳定性。方法:这是一项前瞻性、评估者蒙面、单外科医生研究,涉及29例患者58只眼。患者双侧植入Clareon单焦点人工晶状体(CNA0T0, Alcon Vision LLC)。术后1 - 3个月评估屈光稳定性。术后3个月,收集双眼未矫正视力和距离矫正视力(4 m)、中间(80 cm和66 cm)和双眼离焦曲线的数据。结果:术后1 ~ 3个月屈光度差异有统计学意义(p < 0.001)。术后未矫正距离平均视力为-0.01±0.10 logMAR,矫正距离平均视力为-0.04±0.06 logMAR。在80 cm和66 cm处,术后未矫正的平均中间视力分别为0.16±0.13 logMAR和0.24±0.14 logMAR。距离校正到位后,80cm和60cm处的平均视力分别为0.16±0.13 logMAR和0.23±0.14 logMAR。结论:Clareon单焦点人工晶状体术后屈光稳定,远视力良好,中间视力功能良好。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Clinical ophthalmology
Clinical ophthalmology OPHTHALMOLOGY-
CiteScore
3.50
自引率
9.10%
发文量
499
审稿时长
16 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信