Toward acquisition protocol standardization for estimating liver fat content using ultrasound attenuation coefficient imaging.

IF 2.4 3区 医学 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Ultrasonography Pub Date : 2023-07-01 DOI:10.14366/usg.23014
Giovanna Ferraioli, Ambra Raimondi, Annalisa De Silvestri, Carlo Filice, Richard G Barr
{"title":"Toward acquisition protocol standardization for estimating liver fat content using ultrasound attenuation coefficient imaging.","authors":"Giovanna Ferraioli,&nbsp;Ambra Raimondi,&nbsp;Annalisa De Silvestri,&nbsp;Carlo Filice,&nbsp;Richard G Barr","doi":"10.14366/usg.23014","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study's primary aim was to assess factors affecting ultrasound attenuation coefficient (AC) measurement repeatability using the Canon ultrasound (US) system. The secondary aim was to evaluate whether similar results were obtained with other vendors' AC algorithms.</p><p><strong>Methods: </strong>This prospective study was performed at two centers from February to November 2022. AC was obtained using two US systems (Aplio i800 of Canon Medical Systems and Arietta 850 of Fujifilm). An algorithm combining AC and the backscatter coefficient was also used (Sequoia US System, Siemens Healthineers). To evaluate inter-observer concordance, AC was obtained by two expert operators using different transducer positions with regions of interest (ROIs) varying in terms of depth and size. Intra-observer concordance was evaluated on measurements performed intercostally, subcostally, and in the left liver lobe. Lin's concordance correlation coefficient was used.</p><p><strong>Results: </strong>Thirty-four participants (mean age, 49.4±15.1 years; 18 females) were studied. AC values progressively decreased with depth. The measurements in intercostal spaces on bestquality US images using a 3-cm ROI with its upper edge 2 cm below the liver capsule during breath-hold showed the highest intra-observer and inter-observer concordance (0.92 [95% confidence interval, 0.88 to 0.95] and 0.89 [0.82 to 0.96], respectively). Measurements in the left lobe showed the lowest intra-observer and inter-observer concordance (0.67 [0.43 to 0.90] and 0.58 [0.12 to 1.00], respectively). Intercostal space measurements also had the highest repeatability for the other two ultrasound systems.</p><p><strong>Conclusion: </strong>AC values obtained in intercostal spaces on best-quality images using a 3-cm ROI placed with its top 2 cm below the liver capsule were highly repeatable.</p>","PeriodicalId":54227,"journal":{"name":"Ultrasonography","volume":null,"pages":null},"PeriodicalIF":2.4000,"publicationDate":"2023-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/d0/33/usg-23014.PMC10331055.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Ultrasonography","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.14366/usg.23014","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: This study's primary aim was to assess factors affecting ultrasound attenuation coefficient (AC) measurement repeatability using the Canon ultrasound (US) system. The secondary aim was to evaluate whether similar results were obtained with other vendors' AC algorithms.

Methods: This prospective study was performed at two centers from February to November 2022. AC was obtained using two US systems (Aplio i800 of Canon Medical Systems and Arietta 850 of Fujifilm). An algorithm combining AC and the backscatter coefficient was also used (Sequoia US System, Siemens Healthineers). To evaluate inter-observer concordance, AC was obtained by two expert operators using different transducer positions with regions of interest (ROIs) varying in terms of depth and size. Intra-observer concordance was evaluated on measurements performed intercostally, subcostally, and in the left liver lobe. Lin's concordance correlation coefficient was used.

Results: Thirty-four participants (mean age, 49.4±15.1 years; 18 females) were studied. AC values progressively decreased with depth. The measurements in intercostal spaces on bestquality US images using a 3-cm ROI with its upper edge 2 cm below the liver capsule during breath-hold showed the highest intra-observer and inter-observer concordance (0.92 [95% confidence interval, 0.88 to 0.95] and 0.89 [0.82 to 0.96], respectively). Measurements in the left lobe showed the lowest intra-observer and inter-observer concordance (0.67 [0.43 to 0.90] and 0.58 [0.12 to 1.00], respectively). Intercostal space measurements also had the highest repeatability for the other two ultrasound systems.

Conclusion: AC values obtained in intercostal spaces on best-quality images using a 3-cm ROI placed with its top 2 cm below the liver capsule were highly repeatable.

超声衰减系数成像估计肝脏脂肪含量的采集方案标准化研究。
目的:本研究的主要目的是评估影响佳能超声(US)系统超声衰减系数(AC)测量重复性的因素。第二个目的是评估是否与其他供应商的AC算法获得类似的结果。方法:本前瞻性研究于2022年2月至11月在两个中心进行。AC使用两种美国系统(佳能医疗系统的应用程序i800和富士胶片的Arietta 850)获得。还使用了一种结合交流和后向散射系数的算法(Sequoia US System, Siemens Healthineers)。为了评估观察者之间的一致性,两个专家算子使用不同的换能器位置获得AC,感兴趣区域(roi)在深度和大小方面不同。通过肋间、肋下和左肝叶的测量来评估观察者内部的一致性。采用林氏一致性相关系数。结果:34例受试者(平均年龄49.4±15.1岁;18只雌性)。交流电值随深度逐渐降低。在屏气期间,使用上边缘位于肝包膜下方2 cm的3cm ROI在最佳质量的US图像上测量肋间隙时,观察者内部和观察者之间的一致性最高(分别为0.92[95%置信区间,0.88至0.95]和0.89[0.82至0.96])。左脑叶的测量结果显示观察者内部和观察者之间的一致性最低(分别为0.67[0.43至0.90]和0.58[0.12至1.00])。肋间隙测量在其他两种超声系统中也具有最高的重复性。结论:使用顶部位于肝包膜下方2cm的3cm ROI在肋间隙获得的AC值具有高度可重复性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Ultrasonography
Ultrasonography Medicine-Radiology, Nuclear Medicine and Imaging
CiteScore
5.10
自引率
6.50%
发文量
78
审稿时长
15 weeks
期刊介绍: Ultrasonography, the official English-language journal of the Korean Society of Ultrasound in Medicine (KSUM), is an international peer-reviewed academic journal dedicated to practice, research, technology, and education dealing with medical ultrasound. It is renamed from the Journal of Korean Society of Ultrasound in Medicine in January 2014, and published four times per year: January 1, April 1, July 1, and October 1. Original articles, technical notes, topical reviews, perspectives, pictorial essays, and timely editorial materials are published in Ultrasonography covering state-of-the-art content. Ultrasonography aims to provide updated information on new diagnostic concepts and technical developments, including experimental animal studies using new equipment in addition to well-designed reviews of contemporary issues in patient care. Along with running KSUM Open, the annual international congress of KSUM, Ultrasonography also serves as a medium for cooperation among physicians and specialists from around the world who are focusing on various ultrasound technology and disease problems and relevant basic science.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信