Alterations in Pre/Post Oxygenator Flows Due to Fibrin Deposition in the CardioHelp System-A Case Report.

Q2 Health Professions
Tyler Wahl, Angela Stokes, Caleb Varner, Burak Zeybek, Amit Bardia
{"title":"Alterations in Pre/Post Oxygenator Flows Due to Fibrin Deposition in the CardioHelp System-A Case Report.","authors":"Tyler Wahl,&nbsp;Angela Stokes,&nbsp;Caleb Varner,&nbsp;Burak Zeybek,&nbsp;Amit Bardia","doi":"10.1182/ject-239-241","DOIUrl":null,"url":null,"abstract":"<p><p>We present a 62-year-old patient with COVID-19 pneumonia on Veno-venous (VV) Extracorporeal Membrane Oxygenation (ECMO) with unique perturbations to pre and post oxygenator pressures due to fibrin deposition in despite being on a Heparin/Bivalirudin infusion and activated Partial Thromboplastin Time (aPTT) within therapeutic range of 60-80 seconds. On Day 8 of ECMO support, it was noticed that flows steadily decreased despite unchanged RPMs. Unlike typical blood flow to circuit pressure relationships, the circuit pressures did not correlate with the observed decreased flow. The Delta Pressure (Δ<i>P</i>) was not elevated. The patient's vitals were stable. On inspection post change-out, clots were noted in the oxygenator outlets. Oxygenator clots are usually associated with increased Δ<i>P</i>. In this scenario, clots in the oxygenator blocked 1 of the 4 outlets in the oxygenator causing the flow, pressures, and Δ<i>P</i> to drop consecutively. Due to reduced flow, the Δ<i>P</i> was not elevated despite extensive clots. The fibrin clot location in the CardioHelp ECMO circuit may lead to unexpected pressure and flow alterations. Sole reliance on Δ<i>P</i> as a marker for oxygenator clots may be misleading. Careful monitoring and timely diagnosis of coagulation status may lead to changes in anticoagulation goals and meaningfully impact patient outcomes.</p>","PeriodicalId":39644,"journal":{"name":"Journal of Extra-Corporeal Technology","volume":"54 3","pages":"239-241"},"PeriodicalIF":0.0000,"publicationDate":"2022-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC9891478/pdf/ject-239-241.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Extra-Corporeal Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1182/ject-239-241","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 0

Abstract

We present a 62-year-old patient with COVID-19 pneumonia on Veno-venous (VV) Extracorporeal Membrane Oxygenation (ECMO) with unique perturbations to pre and post oxygenator pressures due to fibrin deposition in despite being on a Heparin/Bivalirudin infusion and activated Partial Thromboplastin Time (aPTT) within therapeutic range of 60-80 seconds. On Day 8 of ECMO support, it was noticed that flows steadily decreased despite unchanged RPMs. Unlike typical blood flow to circuit pressure relationships, the circuit pressures did not correlate with the observed decreased flow. The Delta Pressure (ΔP) was not elevated. The patient's vitals were stable. On inspection post change-out, clots were noted in the oxygenator outlets. Oxygenator clots are usually associated with increased ΔP. In this scenario, clots in the oxygenator blocked 1 of the 4 outlets in the oxygenator causing the flow, pressures, and ΔP to drop consecutively. Due to reduced flow, the ΔP was not elevated despite extensive clots. The fibrin clot location in the CardioHelp ECMO circuit may lead to unexpected pressure and flow alterations. Sole reliance on ΔP as a marker for oxygenator clots may be misleading. Careful monitoring and timely diagnosis of coagulation status may lead to changes in anticoagulation goals and meaningfully impact patient outcomes.

心脏辅助系统中纤维蛋白沉积导致氧合器前后血流的改变——一例报告
我们报告了一名62岁的COVID-19肺炎患者,在静脉-静脉(VV)体外膜氧合(ECMO)治疗中,尽管使用肝素/比伐鲁定输注并在60-80秒的治疗范围内激活了部分凝血活素时间(aPTT),但由于纤维蛋白沉积而导致氧合器前后压力的独特扰动。在ECMO支持的第8天,尽管rpm不变,但注意到流量稳步下降。与典型的血流与回路压力的关系不同,回路压力与观察到的血流减少无关。δ压(ΔP)没有升高。病人的生命体征稳定。换岗检查时,氧合器出口发现血块。氧合器凝块通常与ΔP升高有关。在这种情况下,氧合器中的血块堵塞了氧合器4个出口中的1个,导致流量、压力和ΔP连续下降。由于血流减少,尽管存在大量血栓,但ΔP并未升高。纤维蛋白凝块在CardioHelp ECMO回路中的位置可能导致意想不到的压力和流量改变。单纯依赖ΔP作为氧合器血栓的标记可能会产生误导。仔细监测和及时诊断凝血状态可能导致抗凝目标的改变,并对患者的预后产生重大影响。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Extra-Corporeal Technology
Journal of Extra-Corporeal Technology Medicine-Medicine (all)
CiteScore
1.90
自引率
0.00%
发文量
12
期刊介绍: The Journal of Extracorporeal Technology is dedicated to the study and practice of Basic Science and Clinical issues related to extracorporeal circulation. Areas emphasized in the Journal include: •Cardiopulmonary Bypass •Cardiac Surgery •Cardiovascular Anesthesia •Hematology •Blood Management •Physiology •Fluid Dynamics •Laboratory Science •Coagulation and Hematology •Transfusion •Business Practices •Pediatric Perfusion •Total Quality Management • Evidence-Based Practices
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信