Evaluation of the rotary kinematics between actual and set speeds of X-Smart Plus, VDW.Silver and iRoot motors.

Q2 Medicine
Renata de Castro Monteiro-Netto, Dieimes Braambati, Rodrigo Arruda-Vasconcelos, Adriana de-Jesus Soares, Marcos Frozoni
{"title":"Evaluation of the rotary kinematics between actual and set speeds of X-Smart Plus, VDW.Silver and iRoot motors.","authors":"Renata de Castro Monteiro-Netto,&nbsp;Dieimes Braambati,&nbsp;Rodrigo Arruda-Vasconcelos,&nbsp;Adriana de-Jesus Soares,&nbsp;Marcos Frozoni","doi":"10.1590/0103-6440202304953","DOIUrl":null,"url":null,"abstract":"<p><p>The present study evaluated the actual rotational speed of three different endodontic motors compared to the values provided by the manufacturers. A total of three endodontic motors (X-Smart Plus, VDW.Silver, and iRoot) were tested at 400 rpm and 800 rpm and 2 N/cm2 torque. The kinematics of the devices was recorded by using a custom angle-measuring disc with a 50-mm diameter attached to the handpiece provided by the manufacturer, whereas their movement was captured by a high-speed camera at 2,400 frames per second, 800 x 800 pixel-resolution and distance of 0.3 m from the target object. Statistical analysis was performed at a significance level of 5%. At 400 rpm, the iRoot motor had a value of 17.94 rpm above that indicated by the manufacturer, which was significantly different from those of X-Smart Plus (5.20 rpm below that indicated by the manufacturer) and VDW.Silver (0.62 rpm above that indicated by the manufacturer) motors (P < 0.05). At 800 rpm, the iRoot motor had a value of 51.34 rpm below that indicated by the manufacturer, whereas the X-Smart Plus motor had a value of 13.00 rpm below that indicated by the manufacturer (P > 0.05). The VDW.Silver motor statistically differed from the iRoot and X-Smart Plus ones, showing a value of 1.68 rpm above that indicated by the manufacturer. In conclusion, the X-Smart Plus, VDW.Silver, and iRoot motors showed lower variations in the rotational speed values compared to those reported by their manufacturers. The endodontic motors presented different behaviors between them, with the VDW.Silver motor presents the most accurate values and the iRoot presents the most divergent values.</p>","PeriodicalId":9211,"journal":{"name":"Brazilian dental journal","volume":"34 2","pages":"21-26"},"PeriodicalIF":0.0000,"publicationDate":"2023-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10208292/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Brazilian dental journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1590/0103-6440202304953","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

The present study evaluated the actual rotational speed of three different endodontic motors compared to the values provided by the manufacturers. A total of three endodontic motors (X-Smart Plus, VDW.Silver, and iRoot) were tested at 400 rpm and 800 rpm and 2 N/cm2 torque. The kinematics of the devices was recorded by using a custom angle-measuring disc with a 50-mm diameter attached to the handpiece provided by the manufacturer, whereas their movement was captured by a high-speed camera at 2,400 frames per second, 800 x 800 pixel-resolution and distance of 0.3 m from the target object. Statistical analysis was performed at a significance level of 5%. At 400 rpm, the iRoot motor had a value of 17.94 rpm above that indicated by the manufacturer, which was significantly different from those of X-Smart Plus (5.20 rpm below that indicated by the manufacturer) and VDW.Silver (0.62 rpm above that indicated by the manufacturer) motors (P < 0.05). At 800 rpm, the iRoot motor had a value of 51.34 rpm below that indicated by the manufacturer, whereas the X-Smart Plus motor had a value of 13.00 rpm below that indicated by the manufacturer (P > 0.05). The VDW.Silver motor statistically differed from the iRoot and X-Smart Plus ones, showing a value of 1.68 rpm above that indicated by the manufacturer. In conclusion, the X-Smart Plus, VDW.Silver, and iRoot motors showed lower variations in the rotational speed values compared to those reported by their manufacturers. The endodontic motors presented different behaviors between them, with the VDW.Silver motor presents the most accurate values and the iRoot presents the most divergent values.

Abstract Image

Abstract Image

X-Smart Plus, VDW的实际速度和设定速度之间的旋转运动学评估。银色和iRoot电机。
本研究评估了三种不同牙髓马达的实际转速,并与制造商提供的数值进行了比较。共有3台牙髓马达(X-Smart Plus、VDW、VDW、VDW)。Silver和iRoot分别在400 rpm和800 rpm和2 N/cm2的扭矩下进行了测试。设备的运动是通过使用制造商提供的附着在手机上的直径为50毫米的定制角度测量盘来记录的,而它们的运动是由高速摄像机以每秒2400帧,800 x 800像素的分辨率和距离目标物体0.3 m的距离来捕捉的。统计学分析为显著性水平5%。在400转时,iRoot电机的转速比制造商指示的转速高17.94转,与X-Smart Plus的转速(比制造商指示的转速低5.20转)和VDW有显著差异。银色(0.62 rpm以上制造商指示)电机(P < 0.05)。在800 rpm时,iRoot电机的转速比制造商指示的转速低51.34 rpm,而X-Smart Plus电机的转速比制造商指示的转速低13.00 rpm (P > 0.05)。就是secu * tanu减去VDW的。Silver电机的转速与iRoot和X-Smart Plus的转速在统计上有所不同,比制造商提供的转速高1.68 rpm。总之,X-Smart Plus, VDW。与制造商报告的转速值相比,Silver和iRoot电机的转速值变化较小。牙髓马达表现出不同的行为,与VDW一致。银电机呈现最准确的值,iRoot呈现最发散的值。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Brazilian dental journal
Brazilian dental journal Dentistry-Dentistry (all)
CiteScore
2.20
自引率
0.00%
发文量
69
审稿时长
12 weeks
期刊介绍: Brazilian Dental Journal, publishes Full-Length Papers, Short Communications and Case Reports, dealing with dentistry or related disciplines and edited six times a year.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信