{"title":"Effect of Different Retightening Procedures on Screw Stability in External-Hex and Internal Trichannel Connections: An In Vitro Study.","authors":"Jian Sun, Nobuhiro Yoda, Yoshiki Matsudate, Guang Hong, Tetsuo Kawata, Keiichi Sasaki","doi":"10.11607/ijp.6863","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the effect of different abutment screw retightening procedures on the screw stability between the abutment and implant.</p><p><strong>Materials and methods: </strong>External hexagon connection (EHC) and internal trichannel connection (ITC) implants were employed in this study. Each abutment screw was used for the following specific applications (n = 5 for each group): group 0 = abutment was tightened and retightened with a 10-minute interval; group 1 = abutment was tightened and retightened with a 10-minute interval, then loaded for 500,000 cycles; group 2 = abutment was loaded for 83,000 cycles during tightening and retightening, then loaded for 500,000 cycles; group 3 = abutment was loaded for 167,000 cycles during tightening and retightening, then loaded for 500,000 cycles; and group 4 = abutment was loaded for 250,000 cycles during tightening and retightening, then loaded for 500,000 cycles. Both tightening and retightening torques were set at 35 Ncm, and dynamic oblique loading between 0 N and 200 N was applied on the abutment in all groups. The settling values of the abutments after retightening and the abutment removal torque values (RTVs) after testing were measured.</p><p><strong>Results: </strong>No significant differences in settling values were found between EHC and ITC. There were significant differences in the RTVs of EHC implants, but ITC implants showed no difference in RTV with different prosthetic retightening applications.</p><p><strong>Conclusion: </strong>The retightening application affected the joint stability of EHC implants but did not affect the settlement and joint stability of ITC implants. Int J Prosthodont 2023;36:181-188. doi: 10.11607/ijp.6863.</p>","PeriodicalId":50292,"journal":{"name":"International Journal of Prosthodontics","volume":"36 2","pages":"181-188"},"PeriodicalIF":2.1000,"publicationDate":"2023-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Prosthodontics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.11607/ijp.6863","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To evaluate the effect of different abutment screw retightening procedures on the screw stability between the abutment and implant.
Materials and methods: External hexagon connection (EHC) and internal trichannel connection (ITC) implants were employed in this study. Each abutment screw was used for the following specific applications (n = 5 for each group): group 0 = abutment was tightened and retightened with a 10-minute interval; group 1 = abutment was tightened and retightened with a 10-minute interval, then loaded for 500,000 cycles; group 2 = abutment was loaded for 83,000 cycles during tightening and retightening, then loaded for 500,000 cycles; group 3 = abutment was loaded for 167,000 cycles during tightening and retightening, then loaded for 500,000 cycles; and group 4 = abutment was loaded for 250,000 cycles during tightening and retightening, then loaded for 500,000 cycles. Both tightening and retightening torques were set at 35 Ncm, and dynamic oblique loading between 0 N and 200 N was applied on the abutment in all groups. The settling values of the abutments after retightening and the abutment removal torque values (RTVs) after testing were measured.
Results: No significant differences in settling values were found between EHC and ITC. There were significant differences in the RTVs of EHC implants, but ITC implants showed no difference in RTV with different prosthetic retightening applications.
Conclusion: The retightening application affected the joint stability of EHC implants but did not affect the settlement and joint stability of ITC implants. Int J Prosthodont 2023;36:181-188. doi: 10.11607/ijp.6863.
期刊介绍:
Official Journal of the European Association for Osseointegration (EAO), the International College of Prosthodontists (ICP), the German Society of Prosthodontics and Dental Materials Science (DGPro), and the Italian Academy of Prosthetic Dentistry (AIOP)
Prosthodontics demands a clinical research emphasis on patient- and dentist-mediated concerns in the management of oral rehabilitation needs. It is about making and implementing the best clinical decisions to enhance patients'' quality of life via applied biologic architecture - a role that far exceeds that of traditional prosthetic dentistry, with its emphasis on materials and techniques. The International Journal of Prosthodontics is dedicated to exploring and developing this conceptual shift in the role of today''s prosthodontist, clinician, and educator alike. The editorial board is composed of a distinguished team of leading international scholars.