Performance evaluation of thrombus molecular markers thrombomodulin, thrombin-antithrombin complex, plasmin-α2-plasmin inhibitor complex, and tissue plasminogen activator-inhibitor complex by a chemiluminescence analyzer.

IF 0.3 4区 医学 Q3 MEDICINE, GENERAL & INTERNAL
Hippokratia Pub Date : 2022-04-01
T Lirong, X Minmin, Y Manqiu, L Yanping
{"title":"Performance evaluation of thrombus molecular markers thrombomodulin, thrombin-antithrombin complex, plasmin-α2-plasmin inhibitor complex, and tissue plasminogen activator-inhibitor complex by a chemiluminescence analyzer.","authors":"T Lirong,&nbsp;X Minmin,&nbsp;Y Manqiu,&nbsp;L Yanping","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To comprehensively evaluate the primary performance regarding the task of detecting thrombomodulin (TM), thrombin-antithrombin complex (TAT), plasmin-α2-plasmin inhibitor complex (PIC), and tissue plasminogen activator/plasminogen activator inhibitor-1 complex (t-PAIC) by Sysmex HisCL5000 high sensitivity chemiluminescence analyzer.</p><p><strong>Methods: </strong>The performance of the chemiluminescence analyzer was evaluated according to the Clinical and Laboratory Standards Institute (CLSI) documents for in-batch precision, daytime precision, carryover rate, linearity, and reference range.</p><p><strong>Results: </strong>The intra-batch and inter-day variation coefficients of the test items were all less than 5 %, and the contamination rate of each index was less than 10 %. The linear verification analysis showed that the correlation coefficients of TM, TAT, PIC, and t-PAIC were 0.9968, 0.9988, 0.9981, and 0.9930, respectively. The project recommended reference range was applicable to our laboratory.</p><p><strong>Conclusion: </strong>The high-sensitivity chemiluminescence analyzer has good performance in the detection of TM, TAT, PIC, and t-PAIC and is suitable for the detection of clinical specimens. HIPPOKRATIA 2022, 26 (2):78-82.</p>","PeriodicalId":50405,"journal":{"name":"Hippokratia","volume":"26 2","pages":"78-82"},"PeriodicalIF":0.3000,"publicationDate":"2022-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10177852/pdf/hippokratia-26-78.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Hippokratia","FirstCategoryId":"3","ListUrlMain":"","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: To comprehensively evaluate the primary performance regarding the task of detecting thrombomodulin (TM), thrombin-antithrombin complex (TAT), plasmin-α2-plasmin inhibitor complex (PIC), and tissue plasminogen activator/plasminogen activator inhibitor-1 complex (t-PAIC) by Sysmex HisCL5000 high sensitivity chemiluminescence analyzer.

Methods: The performance of the chemiluminescence analyzer was evaluated according to the Clinical and Laboratory Standards Institute (CLSI) documents for in-batch precision, daytime precision, carryover rate, linearity, and reference range.

Results: The intra-batch and inter-day variation coefficients of the test items were all less than 5 %, and the contamination rate of each index was less than 10 %. The linear verification analysis showed that the correlation coefficients of TM, TAT, PIC, and t-PAIC were 0.9968, 0.9988, 0.9981, and 0.9930, respectively. The project recommended reference range was applicable to our laboratory.

Conclusion: The high-sensitivity chemiluminescence analyzer has good performance in the detection of TM, TAT, PIC, and t-PAIC and is suitable for the detection of clinical specimens. HIPPOKRATIA 2022, 26 (2):78-82.

Abstract Image

通过化学发光分析仪对血栓分子标记物血栓调节蛋白、凝血酶-抗凝血酶复合物、纤溶酶-α2-纤溶酶抑制剂复合物和组织纤溶酶原激活物抑制剂复合物的性能评估。
目的:综合评价Sysmex HisCL5000高灵敏度化学发光分析仪检测血栓调节蛋白(TM)、凝血酶-抗凝血酶复合物(TAT)、纤溶酶-α2-纤溶酶抑制剂复合物(PIC)和组织纤溶酶原激活剂/纤溶酶原激活物抑制物-1复合物(t-PAIC)的主要性能。方法:根据临床和实验室标准研究所(CLSI)的文件评估化学发光分析仪的性能,包括批内精密度、日间精密度、携带率、线性度和参考范围。结果:检测项目批内和日间变异系数均小于5%,各项指标污染率均小于10%。线性验证分析表明,TM、TAT、PIC和t-PAIC的相关系数分别为0.9968、0.9988、0.9981和0.9930。项目推荐的参考范围适用于我们的实验室。结论:高灵敏度化学发光分析仪对TM、TAT、PIC和t-PAIC具有良好的检测性能,适用于临床标本的检测。HIPPOKRATIA 2022,26(2):78-82。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Hippokratia
Hippokratia MEDICINE, GENERAL & INTERNAL-
CiteScore
1.10
自引率
0.00%
发文量
0
审稿时长
6-12 weeks
期刊介绍: Hippokratia journal is a quarterly issued, open access, peer reviewed, general medical journal, published in Thessaloniki, Greece. It is a forum for all medical specialties. The journal is published continuously since 1997, its official language is English and all submitted manuscripts undergo peer review by two independent reviewers, assigned by the Editor (double blinded review process). Hippokratia journal is managed by its Editorial Board and has an International Advisory Committee and over 500 expert Reviewers covering all medical specialties and additionally Technical Reviewers, Statisticians, Image processing Experts and a journal Secretary. The Society “Friends of Hippokratia Journal” has the financial management of both the printed and electronic edition of the journal.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信