Yasser Awaad, Tamer Rizk, Iram Siddiqui, Norbert Roosen, Kelly McIntosh, G Michael Waines
{"title":"Complications of intrathecal baclofen pump: prevention and cure.","authors":"Yasser Awaad, Tamer Rizk, Iram Siddiqui, Norbert Roosen, Kelly McIntosh, G Michael Waines","doi":"10.5402/2012/575168","DOIUrl":null,"url":null,"abstract":"<p><p>Increasingly, spasticity is managed with surgically implanted Intrathecal Baclofen pumps. Intrathecal Baclofen pump revision surgery unrelated to programmable pump end-of-life is not uncommon, requiring special attention during pre-, intra-, and postoperative management. We aimed to identify and describe complications of Intrathecal Baclofen pump as well as to report avoidance and management of complications. Methods and Materials. Through 2002-2006, at the department of neurosurgery, Henry Ford and Oakwood Health Systems, Intrathecal Baclofen pumps were implanted in 44 patients: 24 children versus 20 adults; 30 \"primary-implant-patients\"; 14 \"revision-only patients\". We evaluated reasons for revision surgeries and diagnostic workup requirements. Results. Eight primary-implant-patients required 14 revisions and 7 of revision-only patients needed 13 procedures. Seven patients with slowly increasing baclofen-resistant spasticity had either (i) unsuspected pump-catheter connector defects, (ii) an X-ray-documented pump-catheter connector defect, (iii) X-ray-demonstrated fractured catheter with intrathecal fragment. Implant infections occurred in 4 cases. Scintigraphy revealed occult CSF leakage N=1 and intrinsic pump failure N=1. Conclusion. Intrathecal Baclofen pumps, although very gratifying, have a high, technique-related complication incidence during implant life. Meticulous technique, high clinical suspicion, appropriate workup, and timely surgical management can reduce surgical complications of Intrathecal Baclofen pump implantation.</p>","PeriodicalId":14626,"journal":{"name":"ISRN Neurology","volume":"2012 ","pages":"575168"},"PeriodicalIF":0.0000,"publicationDate":"2012-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.5402/2012/575168","citationCount":"68","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ISRN Neurology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5402/2012/575168","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 68
Abstract
Increasingly, spasticity is managed with surgically implanted Intrathecal Baclofen pumps. Intrathecal Baclofen pump revision surgery unrelated to programmable pump end-of-life is not uncommon, requiring special attention during pre-, intra-, and postoperative management. We aimed to identify and describe complications of Intrathecal Baclofen pump as well as to report avoidance and management of complications. Methods and Materials. Through 2002-2006, at the department of neurosurgery, Henry Ford and Oakwood Health Systems, Intrathecal Baclofen pumps were implanted in 44 patients: 24 children versus 20 adults; 30 "primary-implant-patients"; 14 "revision-only patients". We evaluated reasons for revision surgeries and diagnostic workup requirements. Results. Eight primary-implant-patients required 14 revisions and 7 of revision-only patients needed 13 procedures. Seven patients with slowly increasing baclofen-resistant spasticity had either (i) unsuspected pump-catheter connector defects, (ii) an X-ray-documented pump-catheter connector defect, (iii) X-ray-demonstrated fractured catheter with intrathecal fragment. Implant infections occurred in 4 cases. Scintigraphy revealed occult CSF leakage N=1 and intrinsic pump failure N=1. Conclusion. Intrathecal Baclofen pumps, although very gratifying, have a high, technique-related complication incidence during implant life. Meticulous technique, high clinical suspicion, appropriate workup, and timely surgical management can reduce surgical complications of Intrathecal Baclofen pump implantation.