Comparison of Geometric Accuracy of Low-Dose and Standard-Dose Dental CBCT Imaging Protocols in CAD/CAM-Guided Dental Implant Surgery

IF 1.7 4区 医学 Q3 DENTISTRY, ORAL SURGERY & MEDICINE
Silvan Unger, Marion Penzenstadler, Adib Al-Haj Husain, Daniel Wiedemeier, Bernd Stadlinger, Silvio Valdec
{"title":"Comparison of Geometric Accuracy of Low-Dose and Standard-Dose Dental CBCT Imaging Protocols in CAD/CAM-Guided Dental Implant Surgery","authors":"Silvan Unger,&nbsp;Marion Penzenstadler,&nbsp;Adib Al-Haj Husain,&nbsp;Daniel Wiedemeier,&nbsp;Bernd Stadlinger,&nbsp;Silvio Valdec","doi":"10.11607/jomi.9851","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This preclinical comparison study assessed the diagnostic accuracy of low-dose CBCT protocols compared with standard-dose protocols in digital implant treatment planning and template-guided implant surgery.</p><p><strong>Materials and methods: </strong>Thirty mandibles of pig cadavers underwent both CBCT protocols on an Orthophos SL Unit (Dentsply-Sirona). Surface scans of the regions of interest were performed to create a digital diagnostic wax-up followed by 120 subsequent implant plannings (one implant per quadrant). Simple randomization (1:1) was assessed to assign each quadrant into one of the imaging protocols. Sixty implant surgical guides were manufactured using CAD/CAM technology, followed by the fully guided placement of 60 implants following the surgical protocol in randomized order. Geometric accuracy between the planned and definitive implant position was determined regarding apical distances between the central axes and angle deviation. Descriptive statistics and linear regressions were used for the statistical analysis of the data.</p><p><strong>Results: </strong>Regarding implant apex deviation using low-dose CBCT, the following differences were observed: apical deviation of 0.75 ± 0.63 mm and angular deviation of 2.5 ± 2.12 degrees, while the standard-dose CBCT showed the following results: apical deviation of 0.92 ± 0.55 mm and angular deviation of 3.06 ± 2.12 degrees. The regression analyses could not show evidence for a significant difference between the two CBCT protocols, neither with regard to the apical distance nor in view of the angular deviation.</p><p><strong>Conclusion: </strong>Low-dose CBCT imaging protocols providing accurate 3D anatomical information with an improved benefit-risk ratio according to the as low as diagnostically acceptable (ALADA) principle could become a promising option as a primary diagnostic modality as well as for radiologic follow-up.</p>","PeriodicalId":50298,"journal":{"name":"International Journal of Oral & Maxillofacial Implants","volume":"38 2","pages":"287-294"},"PeriodicalIF":1.7000,"publicationDate":"2023-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Oral & Maxillofacial Implants","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.11607/jomi.9851","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: This preclinical comparison study assessed the diagnostic accuracy of low-dose CBCT protocols compared with standard-dose protocols in digital implant treatment planning and template-guided implant surgery.

Materials and methods: Thirty mandibles of pig cadavers underwent both CBCT protocols on an Orthophos SL Unit (Dentsply-Sirona). Surface scans of the regions of interest were performed to create a digital diagnostic wax-up followed by 120 subsequent implant plannings (one implant per quadrant). Simple randomization (1:1) was assessed to assign each quadrant into one of the imaging protocols. Sixty implant surgical guides were manufactured using CAD/CAM technology, followed by the fully guided placement of 60 implants following the surgical protocol in randomized order. Geometric accuracy between the planned and definitive implant position was determined regarding apical distances between the central axes and angle deviation. Descriptive statistics and linear regressions were used for the statistical analysis of the data.

Results: Regarding implant apex deviation using low-dose CBCT, the following differences were observed: apical deviation of 0.75 ± 0.63 mm and angular deviation of 2.5 ± 2.12 degrees, while the standard-dose CBCT showed the following results: apical deviation of 0.92 ± 0.55 mm and angular deviation of 3.06 ± 2.12 degrees. The regression analyses could not show evidence for a significant difference between the two CBCT protocols, neither with regard to the apical distance nor in view of the angular deviation.

Conclusion: Low-dose CBCT imaging protocols providing accurate 3D anatomical information with an improved benefit-risk ratio according to the as low as diagnostically acceptable (ALADA) principle could become a promising option as a primary diagnostic modality as well as for radiologic follow-up.

CAD/CAM引导下种植牙手术中低剂量和标准剂量牙科CBCT成像方案几何精度的比较
目的:本临床前比较研究评估了低剂量CBCT方案与标准剂量方案在数字植入治疗计划和模板引导植入手术中的诊断准确性。材料和方法:30具猪尸体的下颌骨在Orthophos SL装置(Dentsply Sirona)上接受了两种CBCT方案。对感兴趣区域进行表面扫描,以创建数字诊断蜡,然后进行120次后续植入物规划(每个象限一个植入物)。评估简单随机化(1:1),将每个象限分配到一个成像方案中。使用CAD/CAM技术制造了60个植入物手术引导器,然后按照手术方案以随机顺序完全引导放置60个植入体。根据中心轴之间的心尖距离和角度偏差,确定计划和最终植入位置之间的几何精度。使用描述性统计和线性回归对数据进行统计分析。结果:对于使用低剂量CBCT的种植体顶端偏移,观察到以下差异:顶端偏移0.75±0.63 mm和角度偏移2.5±2.12度,而标准剂量CBCT显示以下结果:顶端偏移0.92±0.55 mm和角度偏差3.06±2.12度。回归分析无法显示两种CBCT方案之间存在显著差异的证据,无论是在根尖距离还是角度偏差方面。结论:根据低至诊断可接受(ALADA)原则,低剂量CBCT成像方案提供准确的三维解剖信息,并提高获益风险比,可能成为一种有前途的主要诊断模式和放射学随访选择。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.30
自引率
5.00%
发文量
115
审稿时长
6 months
期刊介绍: Edited by Steven E. Eckert, DDS, MS ISSN (Print): 0882-2786 ISSN (Online): 1942-4434 This highly regarded, often-cited journal integrates clinical and scientific data to improve methods and results of oral and maxillofacial implant therapy. It presents pioneering research, technology, clinical applications, reviews of the literature, seminal studies, emerging technology, position papers, and consensus studies, as well as the many clinical and therapeutic innovations that ensue as a result of these efforts. The editorial board is composed of recognized opinion leaders in their respective areas of expertise and reflects the international reach of the journal. Under their leadership, JOMI maintains its strong scientific integrity while expanding its influence within the field of implant dentistry. JOMI’s popular regular feature "Thematic Abstract Review" presents a review of abstracts of recently published articles on a specific topical area of interest each issue.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信