Could the radiographic image quality be affected by the excessive use of the photostimulable phosphor plate?

Q2 Medicine
Débora Costa Ruiz, Amanda Farias Gomes, Rocharles Cavalcante Fontenele, Francisco Haiter Neto, Deborah Queiroz Freitas, Francisco Carlos Groppo
{"title":"Could the radiographic image quality be affected by the excessive use of the photostimulable phosphor plate?","authors":"Débora Costa Ruiz,&nbsp;Amanda Farias Gomes,&nbsp;Rocharles Cavalcante Fontenele,&nbsp;Francisco Haiter Neto,&nbsp;Deborah Queiroz Freitas,&nbsp;Francisco Carlos Groppo","doi":"10.1590/0103-6440202305225","DOIUrl":null,"url":null,"abstract":"<p><p>The study aimed to assess the influence of excessive use of a photostimulable phosphor plate (PSP) on the density, image noise, and contrast of radiographs. For that, radiographs of an acrylic block were acquired with a PSP of the Express intraoral system to assess the density and image noise. Initially, five images were obtained and exported (first group). After 400 exposures to X-rays and scannings of PSP, other five images were obtained and exported (second group). The same procedure was done after 800 (third group), 1200 (fourth group), 1600 (fifth group), and 2000 acquisitions (sixth group), resulting in 30 images to be evaluated. The mean and standard deviation of the gray values were calculated for the images using the ImageJ software. For contrast analysis, radiographs of an aluminum step-wedge were acquired with a new PSP following the same acquisition intervals. The percentage of contrast variation was calculated. Another two unused PSP receptors were employed to evaluate the method's reproducibility. The comparison of the results among the acquisition groups was performed with one-way Analysis of Variance (α=0.05). Intraclass Correlation Coefficient (ICC) assessed the reproducibility of the receptors. Image noise did not differ among the groups (p>0.05). There was a slight increase in density after 400 acquisitions and a slight difference in contrast for all acquisition groups without a pattern of increase or decrease (p<0.05). ICC showed excellent reliability for the methods. Therefore, excessive PSP use slightly affected the radiograph's density and contrast.</p>","PeriodicalId":9211,"journal":{"name":"Brazilian dental journal","volume":"34 1","pages":"39-44"},"PeriodicalIF":0.0000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10027106/pdf/","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Brazilian dental journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1590/0103-6440202305225","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 1

Abstract

The study aimed to assess the influence of excessive use of a photostimulable phosphor plate (PSP) on the density, image noise, and contrast of radiographs. For that, radiographs of an acrylic block were acquired with a PSP of the Express intraoral system to assess the density and image noise. Initially, five images were obtained and exported (first group). After 400 exposures to X-rays and scannings of PSP, other five images were obtained and exported (second group). The same procedure was done after 800 (third group), 1200 (fourth group), 1600 (fifth group), and 2000 acquisitions (sixth group), resulting in 30 images to be evaluated. The mean and standard deviation of the gray values were calculated for the images using the ImageJ software. For contrast analysis, radiographs of an aluminum step-wedge were acquired with a new PSP following the same acquisition intervals. The percentage of contrast variation was calculated. Another two unused PSP receptors were employed to evaluate the method's reproducibility. The comparison of the results among the acquisition groups was performed with one-way Analysis of Variance (α=0.05). Intraclass Correlation Coefficient (ICC) assessed the reproducibility of the receptors. Image noise did not differ among the groups (p>0.05). There was a slight increase in density after 400 acquisitions and a slight difference in contrast for all acquisition groups without a pattern of increase or decrease (p<0.05). ICC showed excellent reliability for the methods. Therefore, excessive PSP use slightly affected the radiograph's density and contrast.

Abstract Image

Abstract Image

过度使用可光刺激磷光片会否影响放射影像质素?
本研究旨在评估过度使用光刺激磷光片(PSP)对x线片密度、图像噪声和对比度的影响。为此,使用Express口腔内系统的PSP获取丙烯酸块的x线片以评估密度和图像噪声。最初,获得5张图像并输出(第一组)。经过400次x射线曝光和PSP扫描后,获得另外5张图像并输出(第二组)。对800张(第三组)、1200张(第四组)、1600张(第五组)和2000张(第六组)进行同样的处理,得到30张待评估的图像。利用ImageJ软件计算图像灰度值的均值和标准差。为了对比分析,在相同的采集间隔后,用新的PSP获得了铝阶梯楔形的x线片。计算对比度变化的百分比。另外两个未使用的PSP受体用于评价方法的再现性。各组间比较采用单因素方差分析(α=0.05)。用类内相关系数(Intraclass Correlation Coefficient, ICC)评价受体的可重复性。各组图像噪声差异无统计学意义(p>0.05)。在400次收购后,密度略有增加,而所有收购组的对比略有差异,没有增加或减少的模式(p
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Brazilian dental journal
Brazilian dental journal Dentistry-Dentistry (all)
CiteScore
2.20
自引率
0.00%
发文量
69
审稿时长
12 weeks
期刊介绍: Brazilian Dental Journal, publishes Full-Length Papers, Short Communications and Case Reports, dealing with dentistry or related disciplines and edited six times a year.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信