Richard W. Quine, George A. Rinard, Yilin Shi, Laura Buchanan, Joshua R. Biller, Sandra S. Eaton, Gareth R. Eaton
{"title":"UHF EPR spectrometer operating at frequencies between 400 MHz and 1 GHz","authors":"Richard W. Quine, George A. Rinard, Yilin Shi, Laura Buchanan, Joshua R. Biller, Sandra S. Eaton, Gareth R. Eaton","doi":"10.1002/cmr.b.21328","DOIUrl":null,"url":null,"abstract":"<p>A spectrometer was designed and constructed to facilitate measurements of <i>T</i><sub>1</sub>, <i>T</i><sub>2</sub>, spin echo signal-to-noise, and resonator quality factor, <i>Q</i>, between about 400 and 1000 MHz. Pulse patterns are generated at 250 MHz and mixed with the output from a second source to perform excitation and detection. A cross-loop resonator was constructed in which the same sample could be measured in the same resonator over the full range of frequencies. An air-core, four-coil, water-cooled electromagnet with a large experimental volume was built.</p>","PeriodicalId":50623,"journal":{"name":"Concepts in Magnetic Resonance Part B-Magnetic Resonance Engineering","volume":"46B 3","pages":"123-133"},"PeriodicalIF":0.9000,"publicationDate":"2016-10-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1002/cmr.b.21328","citationCount":"7","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Concepts in Magnetic Resonance Part B-Magnetic Resonance Engineering","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/cmr.b.21328","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"CHEMISTRY, PHYSICAL","Score":null,"Total":0}
引用次数: 7
Abstract
A spectrometer was designed and constructed to facilitate measurements of T1, T2, spin echo signal-to-noise, and resonator quality factor, Q, between about 400 and 1000 MHz. Pulse patterns are generated at 250 MHz and mixed with the output from a second source to perform excitation and detection. A cross-loop resonator was constructed in which the same sample could be measured in the same resonator over the full range of frequencies. An air-core, four-coil, water-cooled electromagnet with a large experimental volume was built.
期刊介绍:
Concepts in Magnetic Resonance Part B brings together engineers and physicists involved in the design and development of hardware and software employed in magnetic resonance techniques. The journal welcomes contributions predominantly from the fields of magnetic resonance imaging (MRI), nuclear magnetic resonance (NMR), and electron paramagnetic resonance (EPR), but also encourages submissions relating to less common magnetic resonance imaging and analytical methods.
Contributors come from both academia and industry, to report the latest advancements in the development of instrumentation and computer programming to underpin medical, non-medical, and analytical magnetic resonance techniques.