{"title":"He 集積法による中性子照射量測定のための He 絶対量測定装置","authors":"深堀 智生, 中村 武彦, 森北 喜久馬, 飛松 浩樹, 井尻 秀信, 神田 幸則","doi":"10.15017/17593","DOIUrl":null,"url":null,"abstract":"In the helium accumulation method for neutron dosimetry, it is applied that the number of helium atoms produced through (n,xα) reaction is proportional to neutron fluence in fast neutron irradiation fields. Besides, activities measuring in the activation method saturate to the value depending on their half-lives. A helium atom measurement, system for detecting of helium atoms accumulated in neutron-irradiated samples has been constructed and tested. The samples are vaporized in a furnace to release helium contained in them. The number of helium atoms is measured with a quadrupole mass spectrometer calibrated by a standard helium gas supply which is consist of a known-volume vessel, a pressure gauge and a thermometer. It has been proposed in this work that helium-ion implanted aluminum disks are applied to calibrate the helium atom measurement system. This method has succeeded to obtain results agreed with the standard helium gas supply system.","PeriodicalId":11722,"journal":{"name":"Engineering sciences reports, Kyushu University","volume":"33 1","pages":"31-37"},"PeriodicalIF":0.0000,"publicationDate":"1984-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Engineering sciences reports, Kyushu University","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.15017/17593","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

In the helium accumulation method for neutron dosimetry, it is applied that the number of helium atoms produced through (n,xα) reaction is proportional to neutron fluence in fast neutron irradiation fields. Besides, activities measuring in the activation method saturate to the value depending on their half-lives. A helium atom measurement, system for detecting of helium atoms accumulated in neutron-irradiated samples has been constructed and tested. The samples are vaporized in a furnace to release helium contained in them. The number of helium atoms is measured with a quadrupole mass spectrometer calibrated by a standard helium gas supply which is consist of a known-volume vessel, a pressure gauge and a thermometer. It has been proposed in this work that helium-ion implanted aluminum disks are applied to calibrate the helium atom measurement system. This method has succeeded to obtain results agreed with the standard helium gas supply system.
He集成法测量中子辐照量的He绝对量测量装置
在中子剂量学的氦积累法中,采用(n,xα)反应产生的氦原子数与快中子辐照场中的中子通量成正比。此外,活化法测量的活度根据半衰期的不同而趋于饱和。建立了一套氦原子测量系统,用于检测中子辐照样品中积累的氦原子。样品在熔炉中汽化,释放出其中所含的氦。氦原子的数量是用四极杆质谱仪测量的,由一个已知体积的容器、一个压力表和一个温度计组成的标准氦气供应校准。本文提出用氦离子注入铝盘对氦原子测量系统进行标定。该方法成功地获得了与标准氦气供应系统一致的结果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信