Design And Fabricate A Brace Joint With A New Intermittent Correction Varum Mechanism for Patients With Medial Compartment Knee Osteoarthritis and to Investigate Its Effect on A Patient's Gait Parameters (Case Report)
Roshanak Baghaei Roodsari, A. Esteki, G. Aminian, Esmaeil Ebrahimi Takamjani, M. Mousavi, S. Hoseinzadeh
{"title":"Design And Fabricate A Brace Joint With A New Intermittent Correction Varum Mechanism for Patients With Medial Compartment Knee Osteoarthritis and to Investigate Its Effect on A Patient's Gait Parameters (Case Report)","authors":"Roshanak Baghaei Roodsari, A. Esteki, G. Aminian, Esmaeil Ebrahimi Takamjani, M. Mousavi, S. Hoseinzadeh","doi":"10.32598/rj.22.4.3379.1","DOIUrl":null,"url":null,"abstract":"Osteoarthritis Knee is a common, progressive, and chronic disease that affects approximately 13% of the population over the age of 60. Osteoarthritis causes progressive disability and changes in gait parameters, decreased function and quality of life of patients by causing malalignment of the knee in the frontal plane following destruction of articular cartilage. Braces as a mechanical intervention are a successful treatment for osteoarthritis knee and its associated malalignment. Using a three-point pressure mechanism, braces create a natural alignment in the knee by straps attached to the thigh and shank shells, which reduce the medial knee load and correct the valgus alignment. This distributes the force properly while walking. However, applying a constant force on three pressure points of the leg, thigh and affected knee joint reduces the Adduction angle of the knee in all phases of gait cycle, while the correction of the adduction angle of the knee in patients is essential only in phases between 30% to 60% of the stance phase, when the knee deviation reaches its peak. The aim of this study is to design and fabricate a brace joint with a new intermittent correction mechanism for patients with osteoarthritis medial compartment knee and to investigate its effect on a patient's gait parameters. Technique The new joint design of the knee brace was based on the conversion of the knee extension movement at the end of the swing on the sagittal plate to the abduction movement on the frontal plate by stretching the converter piece. Participant gait was assessed in two ways with and without braces. Kinematic variables including knee joint angles in the sagittal and frontal planes and spatio-temporal variables of patient gait were measured and recorded. Results The overall results of the new brace joint function on the knee showed that the angle of flexion of the knee in the swing phase increased from 44.72 ° to 46.19 ° when using brace. Also, knee adduction angle in the stance phase decreased from 4.25 ° to 2.3 ° and patient's walking speed while wearing the brace increased from 0.88 m / s to 0.93 m / s. On the other hand, step length increased from 1.125 m to 1.451 m when using brace and the percentage of stance phase decreased from 63.53 to 62.68 when using brace. Result In this one-sample study, a knee brace with a new joint was able to help correct the direction of knee joint in the frontal and sagittal plane and affect the patient's gait parameters. It seems that if the same results are obtained in a large-scale clinical study, this brace could be a suitable replacement for invasive methods and inappropriate orthoses","PeriodicalId":46374,"journal":{"name":"JOURNAL OF REHABILITATION","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2022-03-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"JOURNAL OF REHABILITATION","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.32598/rj.22.4.3379.1","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Osteoarthritis Knee is a common, progressive, and chronic disease that affects approximately 13% of the population over the age of 60. Osteoarthritis causes progressive disability and changes in gait parameters, decreased function and quality of life of patients by causing malalignment of the knee in the frontal plane following destruction of articular cartilage. Braces as a mechanical intervention are a successful treatment for osteoarthritis knee and its associated malalignment. Using a three-point pressure mechanism, braces create a natural alignment in the knee by straps attached to the thigh and shank shells, which reduce the medial knee load and correct the valgus alignment. This distributes the force properly while walking. However, applying a constant force on three pressure points of the leg, thigh and affected knee joint reduces the Adduction angle of the knee in all phases of gait cycle, while the correction of the adduction angle of the knee in patients is essential only in phases between 30% to 60% of the stance phase, when the knee deviation reaches its peak. The aim of this study is to design and fabricate a brace joint with a new intermittent correction mechanism for patients with osteoarthritis medial compartment knee and to investigate its effect on a patient's gait parameters. Technique The new joint design of the knee brace was based on the conversion of the knee extension movement at the end of the swing on the sagittal plate to the abduction movement on the frontal plate by stretching the converter piece. Participant gait was assessed in two ways with and without braces. Kinematic variables including knee joint angles in the sagittal and frontal planes and spatio-temporal variables of patient gait were measured and recorded. Results The overall results of the new brace joint function on the knee showed that the angle of flexion of the knee in the swing phase increased from 44.72 ° to 46.19 ° when using brace. Also, knee adduction angle in the stance phase decreased from 4.25 ° to 2.3 ° and patient's walking speed while wearing the brace increased from 0.88 m / s to 0.93 m / s. On the other hand, step length increased from 1.125 m to 1.451 m when using brace and the percentage of stance phase decreased from 63.53 to 62.68 when using brace. Result In this one-sample study, a knee brace with a new joint was able to help correct the direction of knee joint in the frontal and sagittal plane and affect the patient's gait parameters. It seems that if the same results are obtained in a large-scale clinical study, this brace could be a suitable replacement for invasive methods and inappropriate orthoses
膝关节骨性关节炎是一种常见的进行性慢性疾病,大约13%的60岁以上人群患有此病。骨关节炎可导致进行性残疾,步态参数改变,患者的功能和生活质量下降,其原因是关节软骨破坏后,导致膝关节额平面错位。牙套作为一种机械干预手段是治疗膝关节骨关节炎及其相关的关节错位的成功方法。使用三点压力机制,支架通过连接在大腿和小腿壳上的带子在膝盖上形成自然对齐,从而减少膝盖内侧的负荷并纠正外翻对齐。这样可以在行走时适当地分配力。然而,在腿部、大腿和受累膝关节的三个压力点上施加恒定的力,在步态周期的各个阶段,膝关节内收角都会减小,而患者的膝关节内收角只有在站立阶段的30%到60%之间的阶段才需要纠正,此时膝关节偏差达到峰值。本研究的目的是为骨关节炎内侧室膝关节患者设计和制造一种具有新型间歇矫正机制的支架关节,并研究其对患者步态参数的影响。新膝关节支架的关节设计是基于通过拉伸转换片将矢状板摆动结束时的膝关节伸展运动转化为额板上的外展运动。参与者的步态以两种方式评估,有和没有牙套。测量并记录膝关节矢状面和关节面角度等运动学变量以及患者步态的时空变量。结果新型支架关节功能对膝关节的整体效果显示,使用支架时,摆动阶段膝关节屈曲角度由44.72°增加到46.19°。站立期膝关节内收角从4.25°减小到2.3°,患者佩戴支架时的步行速度从0.88 m / s增加到0.93 m / s,步长从使用支架时的1.125 m增加到1.451 m,站立期百分比从使用支架时的63.53降低到62.68。结果在单样本研究中,新关节膝关节支具能够帮助纠正膝关节在额矢状面方向,并影响患者的步态参数。如果在大规模的临床研究中获得相同的结果,这种支具可能是一种合适的替代侵入性方法和不合适的矫形器