Early results for iCMR in atrial flutter

J. Tomala, S. Ulbrich, U. Richter, Y. Huo, T. Gaspar
{"title":"Early results for iCMR in atrial flutter","authors":"J. Tomala, S. Ulbrich, U. Richter, Y. Huo, T. Gaspar","doi":"10.1093/EHJCI/JEAA356.256","DOIUrl":null,"url":null,"abstract":"\n \n \n Type of funding sources: None.\n \n \n \n  As a standard of care, ablation of typical atrial flutter involves performing cardiac catheterization under x-ray fluoroscopy. The unique ability of magnetic resonance (MR) to provide real-time functional imaging in multiple views without ionizing radiation exposure has the potential to be a powerful tool for diagnostic and interventional procedures. Real-time MR imaging-guided radiofrequency (RF) ablation has been used as a part of clinical trials.\n \n \n \n  To implement the MR imaging-guided RF ablation of typical atrial flutter in the clinical routine.\n \n \n \n  From January to July 2020, 15 consecutive patients with typical atrial flutter have been referred for ablation. Patient preparation, conscious sedation and groin puncture took place inside the 1.5 Tesla MR scanner as a part of an MR-only workflow. The catheter advancement as well as the RF ablation procedure have been performed under direct visualisation of catheters with integrated micro-coils using an interactive sequence with active tip tracking and automatic slice positioning in the plane where the catheter tip has been detected. During catheter advancement a single frontal plane has been used to visualize the passage of the catheter through femoral veins and inferior vena cava. With both diagnostic and ablation catheters in the right atrium, an axial stack of balanced steady-state free precession acquisitions has been acquired and used to reconstruct planes in the short- and long cardiac axis. These have been used to provide a \"left- and right anterior oblique\" view familiar to an interventional electrophysiologist. The subsequent catheter placement and ablation have been performed under direct visualisation in these two planes (see figure).\n \n \n \n  The ablation was successful in 14 patients, one patient had to undergo a conventional procedure on the following day. The mean time to reach right atrium and coronary sinus was 4 [3,5] and 7 [6,10] minutes, respectively. The mean total ablation duration and procedure time was 18 [12,26] and 43 [33,58] minutes, respectively. There were no adverse events.\n \n \n \n  The implementation of the MR imaging-guided RF ablation of typical atrial flutter in the clinical routine is feasible and leads to similar success rates and procedure times as the conventional fluoroscopy-based ablation.\n Abstract Figure.\n","PeriodicalId":11963,"journal":{"name":"European Journal of Echocardiography","volume":"10 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2021-02-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Journal of Echocardiography","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1093/EHJCI/JEAA356.256","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Type of funding sources: None.  As a standard of care, ablation of typical atrial flutter involves performing cardiac catheterization under x-ray fluoroscopy. The unique ability of magnetic resonance (MR) to provide real-time functional imaging in multiple views without ionizing radiation exposure has the potential to be a powerful tool for diagnostic and interventional procedures. Real-time MR imaging-guided radiofrequency (RF) ablation has been used as a part of clinical trials.  To implement the MR imaging-guided RF ablation of typical atrial flutter in the clinical routine.  From January to July 2020, 15 consecutive patients with typical atrial flutter have been referred for ablation. Patient preparation, conscious sedation and groin puncture took place inside the 1.5 Tesla MR scanner as a part of an MR-only workflow. The catheter advancement as well as the RF ablation procedure have been performed under direct visualisation of catheters with integrated micro-coils using an interactive sequence with active tip tracking and automatic slice positioning in the plane where the catheter tip has been detected. During catheter advancement a single frontal plane has been used to visualize the passage of the catheter through femoral veins and inferior vena cava. With both diagnostic and ablation catheters in the right atrium, an axial stack of balanced steady-state free precession acquisitions has been acquired and used to reconstruct planes in the short- and long cardiac axis. These have been used to provide a "left- and right anterior oblique" view familiar to an interventional electrophysiologist. The subsequent catheter placement and ablation have been performed under direct visualisation in these two planes (see figure).  The ablation was successful in 14 patients, one patient had to undergo a conventional procedure on the following day. The mean time to reach right atrium and coronary sinus was 4 [3,5] and 7 [6,10] minutes, respectively. The mean total ablation duration and procedure time was 18 [12,26] and 43 [33,58] minutes, respectively. There were no adverse events.  The implementation of the MR imaging-guided RF ablation of typical atrial flutter in the clinical routine is feasible and leads to similar success rates and procedure times as the conventional fluoroscopy-based ablation. Abstract Figure.
心房扑动的iCMR早期结果
资金来源类型:无。作为一种标准的治疗,典型心房扑动的消融包括在x线透视下进行心导管插入术。磁共振(MR)在没有电离辐射暴露的情况下提供多视图实时功能成像的独特能力,有可能成为诊断和介入程序的强大工具。实时磁共振成像引导射频(RF)消融已被用作临床试验的一部分。目的:在临床常规应用磁共振成像引导射频消融治疗典型心房扑动。从2020年1月至7月,连续15例典型心房扑动患者接受了消融治疗。患者准备,清醒镇静和腹股沟穿刺在1.5特斯拉磁共振扫描仪内进行,作为磁共振工作流程的一部分。导管推进和射频消融过程是在集成微线圈的导管直接可视化下进行的,使用具有主动尖端跟踪和在检测到导管尖端的平面上自动切片定位的交互式序列。在导管推进过程中,使用单个额平面来观察导管通过股静脉和下腔静脉的情况。在右心房使用诊断导管和消融导管,获得了平衡稳态自由进动采集的轴向堆栈,并用于重建心脏短轴和长轴的平面。这些已被用于提供介入电生理学家熟悉的“左右前斜位”视图。随后的导管放置和消融在这两个平面的直接目视下进行(见图)。消融在14例患者中成功,1例患者必须在第二天接受常规手术。到达右心房和冠状窦的平均时间分别为4[3,5]和7[6,10]分钟。平均总消融时间和手术时间分别为18[12,26]和43[33,58]分钟。没有不良事件发生。磁共振成像引导射频消融治疗典型心房扑动在临床常规中是可行的,其成功率和手术时间与传统的基于透视的消融相似。抽象的图。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
European Journal of Echocardiography
European Journal of Echocardiography 医学-心血管系统
自引率
0.00%
发文量
0
审稿时长
>12 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书