{"title":"Environmental applications of thermal extraction cone penetrometry and ultrafast gas chromatography/mass spectrometry","authors":"Albert Robbat Jr.","doi":"10.1002/fact.1006","DOIUrl":null,"url":null,"abstract":"<p>This article describes the collection and analysis of organic contaminants at depth without bringing soil to the surface. A thermal extraction cone penetrometer (TECP) probe was used to extract soil-bound semivolatile organics, transferring them to the surface for analysis by ultrafast gas chromatography/mass spectrometry (GC/MS). Findings showed that polychlorinated biphenyls, polycyclic aromatic hydrocarbons, chlorinated pesticides, and explosives could be collected and analyzed in 5 min when the soil-water content was <20% and in 15 min when it was between 20 and 35%. When the TECP was directly connected to the GC/MS, trinitrotoluene (TNT) and four of its synthetic precursors were speciated in 20 sec. Moreover, 51 VOCs were detected by membrane-inlet mass spectrometry in ∼10 sec. Organics were “sniffed” from a vial and identified using the Ion Fingerprint Detection™ software. The algorithms provide the means to untangle complex mass spectra making real-time, electronic nose detection and identification by MS possible. © 2001 John Wiley & Sons, Inc. Field Analyt Chem Technol 5: 60–68, 2001</p>","PeriodicalId":100527,"journal":{"name":"Field Analytical Chemistry & Technology","volume":"5 1-2","pages":"60-68"},"PeriodicalIF":0.0000,"publicationDate":"2001-05-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1002/fact.1006","citationCount":"6","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Field Analytical Chemistry & Technology","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/fact.1006","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 6
Abstract
This article describes the collection and analysis of organic contaminants at depth without bringing soil to the surface. A thermal extraction cone penetrometer (TECP) probe was used to extract soil-bound semivolatile organics, transferring them to the surface for analysis by ultrafast gas chromatography/mass spectrometry (GC/MS). Findings showed that polychlorinated biphenyls, polycyclic aromatic hydrocarbons, chlorinated pesticides, and explosives could be collected and analyzed in 5 min when the soil-water content was <20% and in 15 min when it was between 20 and 35%. When the TECP was directly connected to the GC/MS, trinitrotoluene (TNT) and four of its synthetic precursors were speciated in 20 sec. Moreover, 51 VOCs were detected by membrane-inlet mass spectrometry in ∼10 sec. Organics were “sniffed” from a vial and identified using the Ion Fingerprint Detection™ software. The algorithms provide the means to untangle complex mass spectra making real-time, electronic nose detection and identification by MS possible. © 2001 John Wiley & Sons, Inc. Field Analyt Chem Technol 5: 60–68, 2001
热萃取锥透法和超快速气相色谱/质谱法的环境应用
本文介绍了在不将土壤带到地表的情况下,在深层收集和分析有机污染物。采用热萃取锥探针(TECP)提取土壤结合的半挥发性有机物,并将其转移到土壤表面,通过超快速气相色谱/质谱(GC/MS)进行分析。结果表明:土壤含水量为20%时,5 min即可采集到多氯联苯、多环芳烃、氯化农药和炸药,20% ~ 35%时,15 min即可采集到。当TECP直接连接到GC/MS时,三硝基甲苯(TNT)及其四种合成前体在20秒内被确定。此外,通过膜入口质谱法在10秒内检测到51种VOCs。从小瓶中“嗅”有机物,并使用离子指纹检测™软件进行鉴定。该算法提供了解决复杂质谱的方法,使实时电子鼻检测和质谱鉴定成为可能。©2001 John Wiley &儿子,Inc。化学工程学报(英文版)(5):60-68,2001
本文章由计算机程序翻译,如有差异,请以英文原文为准。