Optimization and performance validation for the fully automatic blood type analysis system

Rongrong Pang, Libo Zhang, Yong Wang, Wenping Han, Min Huang, Qiang Fu
{"title":"Optimization and performance validation for the fully automatic \nblood type analysis system","authors":"Rongrong Pang, Libo Zhang, Yong Wang, Wenping Han, Min Huang, Qiang Fu","doi":"10.46701/apjbg.2019012019112","DOIUrl":null,"url":null,"abstract":"To evaluate and validate the application of fully automatic blood type analysis system parameters under actual lab conditions. All key system parameters were optimized and validated accordingly. The optimized parameters were centrifugal speed at 550 rpm; centrifugal time 20 min; resuspension speed 1,200 rpm; resuspension time: 45 s; incubation temperature 25°C ; incubation time 400 s; incubation rate: 0 rpm. The sampled red blood cell concentration was 3%. The ratio of plasma to red blood cells reagent was 60 μL:30 μL; the ratio of antibody (reagent) to sampled diluted red blood cell was 30 μL:30 μL. After applying our key parameters for optimization and validation, the automatic blood type detection system's performance was found to meet the relevant requirements, effectively improving the accuracy and reliability of the detection system.","PeriodicalId":87226,"journal":{"name":"Blood and genomics","volume":"30 1 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2019-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood and genomics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.46701/apjbg.2019012019112","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

To evaluate and validate the application of fully automatic blood type analysis system parameters under actual lab conditions. All key system parameters were optimized and validated accordingly. The optimized parameters were centrifugal speed at 550 rpm; centrifugal time 20 min; resuspension speed 1,200 rpm; resuspension time: 45 s; incubation temperature 25°C ; incubation time 400 s; incubation rate: 0 rpm. The sampled red blood cell concentration was 3%. The ratio of plasma to red blood cells reagent was 60 μL:30 μL; the ratio of antibody (reagent) to sampled diluted red blood cell was 30 μL:30 μL. After applying our key parameters for optimization and validation, the automatic blood type detection system's performance was found to meet the relevant requirements, effectively improving the accuracy and reliability of the detection system.
全自动血型分析系统的优化与性能验证
评估和验证全自动血型分析系统参数在实际实验室条件下的应用。对所有关键系统参数进行了优化和验证。优化参数为离心转速550 rpm;离心时间20 min;再悬浮速度1200转/分;复苏时间:45 s;培养温度25℃;孵育时间400 s;孵育速率:0 rpm。采集的红细胞浓度为3%。血浆与红细胞试剂的比例为60 μL:30 μL;抗体(试剂)与被稀释红细胞的比例为30 μL:30 μL。应用我们的关键参数进行优化验证后,发现自动血型检测系统的性能满足相关要求,有效提高了检测系统的准确性和可靠性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书