{"title":"Reliability of the Pressure Indicating Film for Occlusal Force Analysis: A Pilot Study","authors":"R. Idris, NI Tasri, SF Yusof, TW Lim, Y. Shoji","doi":"10.24191/cos.v6i0.17495","DOIUrl":null,"url":null,"abstract":"Objective: The aim of this study was to evaluate the reliability of the pressure indicating film in measuring pressure exerted on it with and without Polyethylene (PE) sleeve as infection control purposes, and to analyze the pressure produced with its software for occlusal force study. Materials and Methods: The optimization of the pressure indicating film for occlusal force analysis commenced with the design and calibration of this sheet. The film was designed into horseshoe shape to suit the shape of maxillary and mandibular arches. The calibration was initiated with 5 different types of pressure which were 15 MPa, 25 MPa, 30 MPa, 35 MPa and 45 MPa exerted on two groups of the film: (i) with PE sleeve and (ii) without PE sleeve. Three readings were recorded for each group and mean value was documented. Then, the films were calibrated by its software for pressure analysis. Results: There was no significance difference found between the film with and without PE sleeve during the calibration stage (P>0.05). In all groups of pressure, there was no significant difference documented between pressure exerted and read out value. Conclusion: The results suggested that the film can be used for occlusal force analysis and improvement of the film with addition of PE sleeve for hygienic purpose is suitable to form the basis of clinical occlusal forces study.","PeriodicalId":10525,"journal":{"name":"Compendium of Oral Science","volume":"45 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2019-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Compendium of Oral Science","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.24191/cos.v6i0.17495","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Objective: The aim of this study was to evaluate the reliability of the pressure indicating film in measuring pressure exerted on it with and without Polyethylene (PE) sleeve as infection control purposes, and to analyze the pressure produced with its software for occlusal force study. Materials and Methods: The optimization of the pressure indicating film for occlusal force analysis commenced with the design and calibration of this sheet. The film was designed into horseshoe shape to suit the shape of maxillary and mandibular arches. The calibration was initiated with 5 different types of pressure which were 15 MPa, 25 MPa, 30 MPa, 35 MPa and 45 MPa exerted on two groups of the film: (i) with PE sleeve and (ii) without PE sleeve. Three readings were recorded for each group and mean value was documented. Then, the films were calibrated by its software for pressure analysis. Results: There was no significance difference found between the film with and without PE sleeve during the calibration stage (P>0.05). In all groups of pressure, there was no significant difference documented between pressure exerted and read out value. Conclusion: The results suggested that the film can be used for occlusal force analysis and improvement of the film with addition of PE sleeve for hygienic purpose is suitable to form the basis of clinical occlusal forces study.
压力指示膜用于咬合力分析的可靠性:一项初步研究
目的:评价压力指示膜在有聚乙烯(PE)套和无聚乙烯(PE)套的情况下对其施加压力作为感染控制目的的可靠性,并分析其软件产生的压力用于咬合力研究。材料和方法:优化的压力指示膜的咬合力分析开始于该片的设计和校准。薄膜被设计成马蹄形,以适应上颌和下颌弓的形状。在两组薄膜上分别施加15 MPa、25 MPa、30 MPa、35 MPa和45 MPa 5种不同的压力(1)有PE套筒和2)不带PE套筒。每组记录三次读数并记录平均值。然后,通过其压力分析软件对薄膜进行校准。结果:带PE套筒与不带PE套筒的膜在校正阶段无显著性差异(P>0.05)。在所有压力组中,施加压力和读出值之间没有显著差异。结论:该膜可用于牙合力分析,添加PE卫生套筒对膜进行改进,可作为临床牙合力研究的基础。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信