The bootstrap method to improve statistical analysis of dosimetric data for radiotherapy outcomes

A. Chaikh, J. Bresciani, J. Balosso
{"title":"The bootstrap method to improve statistical analysis of dosimetric data for radiotherapy outcomes","authors":"A. Chaikh, J. Bresciani, J. Balosso","doi":"10.14319/IJCTO.51.2","DOIUrl":null,"url":null,"abstract":"Purpose: The purpose of this study is to validate a new technique in radiotherapy, the medical physicist needs to evaluate the dosimetric benefit and the risk of toxicity before integrating it in the clinical use. Methods: We validate a sound decision tool based on bootstrap method to help the radio oncologist and the medical physicist to usefully analyze the dosimetric data obtained from small-sized samples, with few patients. Statistical investigation principles are presented in the framework of a clinical example based on 36 patients with 6 different cancer sites treated with radiotherapy. For each patient, two treatment plans were generated. In plan 1, the dose was calculated using Modified Batho's (MB) density correction method integrated with pencil beam convolution (PBC) as type (a) algorithm. In plan 2, the dose was calculated using Anisotropic Analytical Algorithm (AAA) as type (b) algorithm. The delivered doses in monitor units (MUs) were compared using the two plans. Then, the bootstrap method was applied to the original data set to assess the dose differences and evaluate the impact of sample size on the 95% confidence interval (95%.CI). Shapiro-Wilks and Wilcoxon signed-rank tests were used to assess the normality of the data and determine the p-value. In addition, Spearman’s rank test was used to calculate the correlation coefficient between the doses calculated with both algorithms. Results: A significant difference was observed between AAA and MB for all tested radiation sites. Spearman’s test indicated a good correlation between the doses calculated with both methods. The bootstrap simulation with 1000 random samplings can be used for small populations with n = 10 and provides a true estimation. Conclusion: one must be cautious when implementing this method for radiotherapy: the data should be representative of the real variations of the cases and the cases should be as homogeneous as possible to avoid bias of over/under estimation of the results.","PeriodicalId":13931,"journal":{"name":"International Journal of Cancer Therapy and Oncology","volume":"11 1","pages":"512"},"PeriodicalIF":0.0000,"publicationDate":"2017-08-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Cancer Therapy and Oncology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.14319/IJCTO.51.2","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Purpose: The purpose of this study is to validate a new technique in radiotherapy, the medical physicist needs to evaluate the dosimetric benefit and the risk of toxicity before integrating it in the clinical use. Methods: We validate a sound decision tool based on bootstrap method to help the radio oncologist and the medical physicist to usefully analyze the dosimetric data obtained from small-sized samples, with few patients. Statistical investigation principles are presented in the framework of a clinical example based on 36 patients with 6 different cancer sites treated with radiotherapy. For each patient, two treatment plans were generated. In plan 1, the dose was calculated using Modified Batho's (MB) density correction method integrated with pencil beam convolution (PBC) as type (a) algorithm. In plan 2, the dose was calculated using Anisotropic Analytical Algorithm (AAA) as type (b) algorithm. The delivered doses in monitor units (MUs) were compared using the two plans. Then, the bootstrap method was applied to the original data set to assess the dose differences and evaluate the impact of sample size on the 95% confidence interval (95%.CI). Shapiro-Wilks and Wilcoxon signed-rank tests were used to assess the normality of the data and determine the p-value. In addition, Spearman’s rank test was used to calculate the correlation coefficient between the doses calculated with both algorithms. Results: A significant difference was observed between AAA and MB for all tested radiation sites. Spearman’s test indicated a good correlation between the doses calculated with both methods. The bootstrap simulation with 1000 random samplings can be used for small populations with n = 10 and provides a true estimation. Conclusion: one must be cautious when implementing this method for radiotherapy: the data should be representative of the real variations of the cases and the cases should be as homogeneous as possible to avoid bias of over/under estimation of the results.
自举法改进放射治疗结果剂量学数据的统计分析
目的:本研究的目的是验证一种新的放射治疗技术,医学物理学家在将其纳入临床应用之前需要评估剂量学效益和毒性风险。方法:我们验证了一种基于自举法的决策工具,以帮助放射肿瘤学家和医学物理学家有效地分析从少量患者的小样本中获得的剂量学数据。统计调查的原则,提出了一个临床实例的框架,基于36名患者的6个不同的癌症部位接受放疗。对于每个病人,产生了两个治疗方案。方案1采用改良Batho (MB)密度校正法结合铅笔束卷积(PBC)作为类型(a)算法计算剂量。在方案2中,剂量计算采用各向异性分析算法(AAA)作为类型(b)算法。比较两种方案给药剂量的监测单位(MUs)。然后,对原始数据集应用bootstrap方法评估剂量差异,并评估样本量对95%置信区间(95%. ci)的影响。使用Shapiro-Wilks和Wilcoxon符号秩检验来评估数据的正态性并确定p值。此外,采用Spearman秩检验计算两种算法计算的剂量之间的相关系数。结果:在所有测试的辐射部位,AAA和MB之间观察到显著差异。斯皮尔曼的试验表明,两种方法计算出的剂量之间存在良好的相关性。具有1000个随机样本的自举模拟可以用于n = 10的小总体,并提供真实的估计。结论:应用该方法进行放疗时必须谨慎:数据应代表病例的真实变化,病例应尽可能均匀,以避免对结果估计过高/过低的偏倚。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书