{"title":"Comparison of an ELISA-based screening test with liquid chromatography for the determination of aflatoxins in corn.","authors":"R. W. Beaver, M. A. James, T. Lin","doi":"10.1093/JAOAC/74.5.827","DOIUrl":null,"url":null,"abstract":"An enzyme-linked immunosorbent assay (ELISA) screening test (CITE PROBE) was compared to liquid chromatography (LC) for the determination of aflatoxins in naturally contaminated corn samples. The CITE PROBE, with a positive/negative cutoff of 5 ng/g aflatoxin B1, was correct (based on LC results) on 47 of 51 samples. Two of the incorrect responses by the CITE PROBE were false positives on samples containing 4.4 ng/g and 4.1 ng/g aflatoxins by LC. Another incorrect response was a false negative on a sample containing 5.5 ng/g aflatoxins by LC. The fourth incorrect response was a false positive on a sample containing 1.9 ng/g aflatoxins by LC. On the basis of these results, the CITE PROBE was determined to be a reliable screening method for the detection of greater than or equal to 5 ng/g aflatoxins in corn.","PeriodicalId":14752,"journal":{"name":"Journal - Association of Official Analytical Chemists","volume":"1 1","pages":"827-9"},"PeriodicalIF":0.0000,"publicationDate":"1991-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"7","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal - Association of Official Analytical Chemists","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1093/JAOAC/74.5.827","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 7
Abstract
An enzyme-linked immunosorbent assay (ELISA) screening test (CITE PROBE) was compared to liquid chromatography (LC) for the determination of aflatoxins in naturally contaminated corn samples. The CITE PROBE, with a positive/negative cutoff of 5 ng/g aflatoxin B1, was correct (based on LC results) on 47 of 51 samples. Two of the incorrect responses by the CITE PROBE were false positives on samples containing 4.4 ng/g and 4.1 ng/g aflatoxins by LC. Another incorrect response was a false negative on a sample containing 5.5 ng/g aflatoxins by LC. The fourth incorrect response was a false positive on a sample containing 1.9 ng/g aflatoxins by LC. On the basis of these results, the CITE PROBE was determined to be a reliable screening method for the detection of greater than or equal to 5 ng/g aflatoxins in corn.