Feasibility of output quality assurance considering gantry angle using “Stealth Chamber”

IF 0.7 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
H. Miura, Fumika Takeichi, Tsubasa Enosaki, K. Yamada, S. Ozawa, Y. Nagata
{"title":"Feasibility of output quality assurance considering gantry angle using “Stealth Chamber”","authors":"H. Miura, Fumika Takeichi, Tsubasa Enosaki, K. Yamada, S. Ozawa, Y. Nagata","doi":"10.2478/pjmpe-2021-0035","DOIUrl":null,"url":null,"abstract":"Abstract Purpose: This study was conducted to demonstrate the feasibility of X-ray output constancy quality assurance (QA) of a linear accelerator for various gantry angles using the Stealth Chamber. Methods: The X-ray output constancy of a Varian TrueBeam STx was evaluated under various gantry angles and a 10 × 10 cm2 field size using a Stealth Chamber. Specifically, 10X and 10X-flattening-filter-free beams with dose rates of 600 and 2400 monitor units (MU)/min, respectively, were used. The Stealth Chamber was attached to the gantry head, and irradiation was performed every 45° for gantry angles of 0-315°. To evaluate the variations in the output constancy with respect to the gantry angle, the acquired values were normalized to the value corresponding to a 0° gantry angle. The obtained results were utilized to determine the correction factors for all gantry angles. To verify the correction factors, additional measurements were performed for five days. Results: The maximum variation in the output constancy measurement relative to the output constancy at a 0° gantry angle was found to be approximately 4.0% for both energy beams at a gantry angle of 180°. Furthermore, the measured values were dependent on the gantry angle. Upon applying the correction factor, the variation in the output constancy with respect to the gantry angle was less than 0.5%. Conclusions: Output constancy QA using the Stealth Chamber for various gantry angles was found to be feasible with the application of a correction factor.","PeriodicalId":53955,"journal":{"name":"Polish Journal of Medical Physics and Engineering","volume":"103 1","pages":"299 - 302"},"PeriodicalIF":0.7000,"publicationDate":"2021-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Polish Journal of Medical Physics and Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2478/pjmpe-2021-0035","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 1

Abstract

Abstract Purpose: This study was conducted to demonstrate the feasibility of X-ray output constancy quality assurance (QA) of a linear accelerator for various gantry angles using the Stealth Chamber. Methods: The X-ray output constancy of a Varian TrueBeam STx was evaluated under various gantry angles and a 10 × 10 cm2 field size using a Stealth Chamber. Specifically, 10X and 10X-flattening-filter-free beams with dose rates of 600 and 2400 monitor units (MU)/min, respectively, were used. The Stealth Chamber was attached to the gantry head, and irradiation was performed every 45° for gantry angles of 0-315°. To evaluate the variations in the output constancy with respect to the gantry angle, the acquired values were normalized to the value corresponding to a 0° gantry angle. The obtained results were utilized to determine the correction factors for all gantry angles. To verify the correction factors, additional measurements were performed for five days. Results: The maximum variation in the output constancy measurement relative to the output constancy at a 0° gantry angle was found to be approximately 4.0% for both energy beams at a gantry angle of 180°. Furthermore, the measured values were dependent on the gantry angle. Upon applying the correction factor, the variation in the output constancy with respect to the gantry angle was less than 0.5%. Conclusions: Output constancy QA using the Stealth Chamber for various gantry angles was found to be feasible with the application of a correction factor.
考虑龙门角度的“隐身腔”输出质量保证的可行性
摘要目的:研究利用隐身室对直线加速器进行不同门架角度x射线输出恒定质量保证(QA)的可行性。方法:在不同的门架角度和10 × 10 cm2的视场尺寸下,使用隐身室评估瓦里安TrueBeam STx的x射线输出常数。具体来说,使用剂量率分别为600和2400监测单位(MU)/min的10X和10X无平坦滤波光束。隐身室附在龙门头上,在0-315°的龙门角范围内,每45°照射一次。为了评估输出常数相对于龙门角的变化,将获得的值归一化为对应于0°龙门角的值。利用所得结果确定了所有龙门角度的修正系数。为了验证校正因子,进行了5天的额外测量。结果:在门架角度为180°时,两束能量束的输出常数测量值相对于0°门架角度时的输出常数的最大变化约为4.0%。此外,测量值依赖于龙门角度。在应用校正因子后,输出常数相对于龙门角的变化小于0.5%。结论:应用校正因子,采用隐身腔对不同龙门角度的输出恒定性质量保证是可行的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Polish Journal of Medical Physics and Engineering
Polish Journal of Medical Physics and Engineering RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
1.30
自引率
0.00%
发文量
19
期刊介绍: Polish Journal of Medical Physics and Engineering (PJMPE) (Online ISSN: 1898-0309; Print ISSN: 1425-4689) is an official publication of the Polish Society of Medical Physics. It is a peer-reviewed, open access scientific journal with no publication fees. The issues are published quarterly online. The Journal publishes original contribution in medical physics and biomedical engineering.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信