{"title":"Evaluation of the immunoradiometric and electrochemiluminescence method for the measurement of serum insulin in children","authors":"Agnieszka Ochocińska, Rafał Śnitko, E. Czekuć-Kryśkiewicz, A. Kępka, M. Szalecki, R. Janas","doi":"10.1080/15321819.2015.1126601","DOIUrl":null,"url":null,"abstract":"Human insulin is a polypeptide hormone produced, stored, and secreted by the ß-cells in the pancreatic islets of Langerhans. Its secretion is stimulated by an increase of the glucose concentration in circulation. Non-radioactive assays are frequently used in many laboratories to measure hormone concentrations, as an alternative to the traditional “gold standard” radioimmuno- and immunoradiometric assays. The precise and reliable determination of the insulin concentration is an important concern in numerous diagnostic procedures. The aim of this study was to compare two commercially available assays (manual and automated) for measurement of serum insulin concentrations. Aliquots of the 86 randomly selected serum samples were measured by Elecsys Insulin Assay (cobas e411 immunoassay analyzer, Roche Diagnostics GmbH, Mannheim, Germany) and DIAsource INS-IRMA Kit (DIAsource ImmunoAssays S.A., Louvain-la-Neuve, Belgium). Compared assays exhibit good correlation (r = 0.996). Insulin concentrations were on average 4.2 μIU/mL lower (p < 0.05) with the cobas e411 immunoassay analyzer when compared to those measured with DIAsouce Immunoassay. Our findings suggest that electrochemiluminescence method on the cobas e411 analyzer and manual IRMA method offered by the DIAsource for the serum insulin determination could be considered interchangeable.","PeriodicalId":15987,"journal":{"name":"Journal of Immunoassay and Immunochemistry","volume":"17 1","pages":"243 - 250"},"PeriodicalIF":0.0000,"publicationDate":"2016-03-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"10","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Immunoassay and Immunochemistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1080/15321819.2015.1126601","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 10

Abstract

Human insulin is a polypeptide hormone produced, stored, and secreted by the ß-cells in the pancreatic islets of Langerhans. Its secretion is stimulated by an increase of the glucose concentration in circulation. Non-radioactive assays are frequently used in many laboratories to measure hormone concentrations, as an alternative to the traditional “gold standard” radioimmuno- and immunoradiometric assays. The precise and reliable determination of the insulin concentration is an important concern in numerous diagnostic procedures. The aim of this study was to compare two commercially available assays (manual and automated) for measurement of serum insulin concentrations. Aliquots of the 86 randomly selected serum samples were measured by Elecsys Insulin Assay (cobas e411 immunoassay analyzer, Roche Diagnostics GmbH, Mannheim, Germany) and DIAsource INS-IRMA Kit (DIAsource ImmunoAssays S.A., Louvain-la-Neuve, Belgium). Compared assays exhibit good correlation (r = 0.996). Insulin concentrations were on average 4.2 μIU/mL lower (p < 0.05) with the cobas e411 immunoassay analyzer when compared to those measured with DIAsouce Immunoassay. Our findings suggest that electrochemiluminescence method on the cobas e411 analyzer and manual IRMA method offered by the DIAsource for the serum insulin determination could be considered interchangeable.
免疫放射法和电化学发光法测定儿童血清胰岛素的评价
人胰岛素是一种多肽激素,由朗格汉斯胰岛的ß-细胞产生、储存和分泌。血液循环中葡萄糖浓度的增加刺激其分泌。许多实验室经常使用非放射性测定法来测量激素浓度,作为传统的“金标准”放射免疫和免疫放射测定法的替代方法。胰岛素浓度的精确和可靠的测定是许多诊断程序中的一个重要问题。本研究的目的是比较两种市售的测定血清胰岛素浓度的方法(手动和自动)。86份随机选择的血清样本的等份采用Elecsys胰岛素测定法(cobas e411免疫测定分析仪,罗氏诊断有限公司,曼海姆,德国)和DIAsource INS-IRMA试剂盒(DIAsource ImmunoAssays s.a., Louvain-la-Neuve,比利时)进行测定。对比分析结果显示相关性良好(r = 0.996)。使用cobas e411免疫分析仪测定胰岛素浓度比使用DIAsouce免疫分析仪测定胰岛素浓度平均降低4.2 μIU/mL (p < 0.05)。我们的研究结果表明,cobas e411分析仪上的电化学发光法与DIAsource提供的手动IRMA法测定血清胰岛素可以被认为是可互换的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信