C. Buchta, J. Camp, J. Jovanović, E. Puchhammer-Stöckl, R. Strassl, MATHIAS M. MüLLER, A. Griesmacher, S. Aberle, I. Görzer
{"title":"Results of a SARS-CoV-2 virus genome detection external quality assessment round focusing on sensitivity of assays and pooling of samples","authors":"C. Buchta, J. Camp, J. Jovanović, E. Puchhammer-Stöckl, R. Strassl, MATHIAS M. MüLLER, A. Griesmacher, S. Aberle, I. Görzer","doi":"10.1515/cclm-2022-0263","DOIUrl":null,"url":null,"abstract":"Abstract Objectives Results of earlier external quality assessment (EQA) rounds suggested remarkable differences in the sensitivity of SARS-CoV PCR assays. Although the test systems are intended to detect SARS-CoV-2 in individual samples, screening is often applied to sample pools to increase efficiency and decrease costs. However, it is unknown to what extent these tests actually meet the manufacturer’s specifications for sensitivity and how they perform when testing sample pools. Methods The sensitivity of assays in routine use was evaluated with a panel of positive samples in a round of a SARS-CoV-2 virus genome detection EQA scheme. The panel consisted of samples at or near the lower limit of detection (“weakly positive”). Laboratories that routinely test sample pools were asked to also analyze the pooled EQA samples according to their usual pool size and dilution method. Results All participants could detect a highly positive patient-derived sample (>106 copies/mL). Most (96%) of the test systems could detect at least 1,000 copies/mL, meeting the minimum acceptable benchmark, and many (94%) detected the vRNA in a sample with lower concentration (500 copies/mL). The false negative ratio increased to 16 and 26% for samples with 100 and 50 copies/mL, respectively. Conclusions The performance of most assays met or exceeded their specification on sensitivity. If assays are to be used to analyze sample pools, the sensitivity of the assay and the number of pooled samples must be balanced.","PeriodicalId":10388,"journal":{"name":"Clinical Chemistry and Laboratory Medicine (CCLM)","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2022-05-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"4","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical Chemistry and Laboratory Medicine (CCLM)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1515/cclm-2022-0263","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 4
Abstract
Abstract Objectives Results of earlier external quality assessment (EQA) rounds suggested remarkable differences in the sensitivity of SARS-CoV PCR assays. Although the test systems are intended to detect SARS-CoV-2 in individual samples, screening is often applied to sample pools to increase efficiency and decrease costs. However, it is unknown to what extent these tests actually meet the manufacturer’s specifications for sensitivity and how they perform when testing sample pools. Methods The sensitivity of assays in routine use was evaluated with a panel of positive samples in a round of a SARS-CoV-2 virus genome detection EQA scheme. The panel consisted of samples at or near the lower limit of detection (“weakly positive”). Laboratories that routinely test sample pools were asked to also analyze the pooled EQA samples according to their usual pool size and dilution method. Results All participants could detect a highly positive patient-derived sample (>106 copies/mL). Most (96%) of the test systems could detect at least 1,000 copies/mL, meeting the minimum acceptable benchmark, and many (94%) detected the vRNA in a sample with lower concentration (500 copies/mL). The false negative ratio increased to 16 and 26% for samples with 100 and 50 copies/mL, respectively. Conclusions The performance of most assays met or exceeded their specification on sensitivity. If assays are to be used to analyze sample pools, the sensitivity of the assay and the number of pooled samples must be balanced.