Dose assessment in high dose rate brachytherapy with cobalt-60 source for cervical cancer treatment: a phantom study

IF 0.7 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
B. K. Bour, S. Inkoom, S. Tagoe, J. Amuasi, E. Sasu, F. Hasford
{"title":"Dose assessment in high dose rate brachytherapy with cobalt-60 source for cervical cancer treatment: a phantom study","authors":"B. K. Bour, S. Inkoom, S. Tagoe, J. Amuasi, E. Sasu, F. Hasford","doi":"10.2478/pjmpe-2020-0029","DOIUrl":null,"url":null,"abstract":"Abstract Transition from low dose rate brachytherapy to high dose rate brachytherapy at our department necessitated the performance of dose verification test, which served as an end-to-end quality assurance procedure to verify and validate dose delivery in intracavitary brachytherapy of the cervix and the vaginal walls based on the Manchester system. An in-house water phantom was designed and constructed from Perspex sheets to represent the cervix region of a standard adult patient. The phantom was used to verify the whole dose delivery chain such as calibration of the cobalt-60 source in use, applicator, and source localization method, the output of treatment planning with dedicated treatment planning system, and actual dose delivery process. Since the above factors would influence the final dose delivered, doses were measured with calibrated gafchromic EBT3 films at various points within the in-house phantom for a number of clinical implants that were used to treat a patient based on departmental protocol. The measured doses were compared to those of the treatment planning system. The discrepancies between measured doses and their corresponding calculated doses obtained with the treatment planning system ranged from -29.67 to 40.34% (mean of ±13.27%). These compared similarly to other studies.","PeriodicalId":53955,"journal":{"name":"Polish Journal of Medical Physics and Engineering","volume":"37 1","pages":"243 - 250"},"PeriodicalIF":0.7000,"publicationDate":"2020-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Polish Journal of Medical Physics and Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2478/pjmpe-2020-0029","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 1

Abstract

Abstract Transition from low dose rate brachytherapy to high dose rate brachytherapy at our department necessitated the performance of dose verification test, which served as an end-to-end quality assurance procedure to verify and validate dose delivery in intracavitary brachytherapy of the cervix and the vaginal walls based on the Manchester system. An in-house water phantom was designed and constructed from Perspex sheets to represent the cervix region of a standard adult patient. The phantom was used to verify the whole dose delivery chain such as calibration of the cobalt-60 source in use, applicator, and source localization method, the output of treatment planning with dedicated treatment planning system, and actual dose delivery process. Since the above factors would influence the final dose delivered, doses were measured with calibrated gafchromic EBT3 films at various points within the in-house phantom for a number of clinical implants that were used to treat a patient based on departmental protocol. The measured doses were compared to those of the treatment planning system. The discrepancies between measured doses and their corresponding calculated doses obtained with the treatment planning system ranged from -29.67 to 40.34% (mean of ±13.27%). These compared similarly to other studies.
高剂量率钴-60源近距离治疗宫颈癌的剂量评估:一项幻象研究
我科低剂量率近距离放疗向高剂量率近距离放疗过渡,需要进行剂量验证试验,作为端到端的质量保证程序,验证和验证基于Manchester系统的宫颈及阴道壁腔内近距离放疗的剂量传递。用有机玻璃片设计和制造了一个内部水模,以代表标准成年患者的宫颈区域。幻影用于验证整个剂量传递链,如使用中的钴-60源的校准,施药器和源定位方法,专用治疗计划系统的治疗计划输出以及实际剂量传递过程。由于上述因素会影响最终剂量,因此根据部门方案,使用校准的变色EBT3薄膜在内部幻体的不同点测量剂量,用于治疗患者的许多临床植入物。将测量的剂量与治疗计划系统的剂量进行比较。测量剂量与相应的治疗计划系统计算剂量的差异范围为- 29.67% ~ 40.34%(平均值为±13.27%)。这些研究与其他研究相似。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Polish Journal of Medical Physics and Engineering
Polish Journal of Medical Physics and Engineering RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
1.30
自引率
0.00%
发文量
19
期刊介绍: Polish Journal of Medical Physics and Engineering (PJMPE) (Online ISSN: 1898-0309; Print ISSN: 1425-4689) is an official publication of the Polish Society of Medical Physics. It is a peer-reviewed, open access scientific journal with no publication fees. The issues are published quarterly online. The Journal publishes original contribution in medical physics and biomedical engineering.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信