{"title":"Performance evaluation of Siemens E.Cam® Spect System at Korle-Bu Teaching Hospital (Ghana","authors":"E. Sosu, F. Hasford, J. Amuasi","doi":"10.4314/JAST.V16I1-2.64789","DOIUrl":null,"url":null,"abstract":"Performance evaluation tests on Siemens e.camR Signature Series Single Photon Emission Computed Tomogra-phy (SPECT) System at Korle-Bu Teaching Hospital (Ghana) were conducted for compliance with Manufactur-erfs specifications and usersf requirements. The tests were performed according to National Electrical Manufac-turers Association protocols, and different measuring instruments and phantoms containing 99m-Tc were used. Extrinsic uniformity, System energy resolution, System spatial resolution (without scattering), Detector shield leakage and System planar sensitivity were evaluated. The System energy resolution measured was 9.38 % as against manufacturersf specification of . 9.9 %, Planar sensitivity value measured was 232.92 c/min/mCi while manufacturersf specification was 202 c/min/mCi, and Detector shield leakage values measured were 0.001 %, 0.022 % and 0.025 % against an accepted value of < 1 %. The performance evaluation confirmed that the Syst-em met requirements for clinical nuclear medicine imaging, and the values could be used as database for future quality control measurements.","PeriodicalId":9207,"journal":{"name":"British Journal of Applied Science and Technology","volume":"39 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2011-03-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"British Journal of Applied Science and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4314/JAST.V16I1-2.64789","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Performance evaluation tests on Siemens e.camR Signature Series Single Photon Emission Computed Tomogra-phy (SPECT) System at Korle-Bu Teaching Hospital (Ghana) were conducted for compliance with Manufactur-erfs specifications and usersf requirements. The tests were performed according to National Electrical Manufac-turers Association protocols, and different measuring instruments and phantoms containing 99m-Tc were used. Extrinsic uniformity, System energy resolution, System spatial resolution (without scattering), Detector shield leakage and System planar sensitivity were evaluated. The System energy resolution measured was 9.38 % as against manufacturersf specification of . 9.9 %, Planar sensitivity value measured was 232.92 c/min/mCi while manufacturersf specification was 202 c/min/mCi, and Detector shield leakage values measured were 0.001 %, 0.022 % and 0.025 % against an accepted value of < 1 %. The performance evaluation confirmed that the Syst-em met requirements for clinical nuclear medicine imaging, and the values could be used as database for future quality control measurements.