Development of the Method and Technique to Analyze Acetone Hydrogenation Products by Gas Chromatography

E. Yakovleva, Mural Nurbol, G. A. Bukhtiyarova
{"title":"Development of the Method and Technique to Analyze Acetone Hydrogenation Products by Gas Chromatography","authors":"E. Yakovleva, Mural Nurbol, G. A. Bukhtiyarova","doi":"10.18412/1816-0387-2023-1-29-36","DOIUrl":null,"url":null,"abstract":"Capillary chromatographic columns with different stationary phases were studied in order to compare their ability to estimate purity of isopropyl alcohol obtained by hydrogenation of acetone. The study was performed with the columns based on polyethylene glycol 20 M terminated with 2-nitroterephthalic acid (PEG20M/FFAP); poly(1-trimethylsilyl-1-propyne) (PTMSP032); and trifluoropropyl (25 %) methyl silicone elastomer (SKTFT 50X). A comparison of the measurement time, asymmetry factors (As) for components of the mixtures, and resolution (Rs) for the pairs of compounds acetone/isopropanol and isopropanol/internal standard gave grounds to choose a capillary column PEG20M/FFAP. A technique was developed for measuring the weight fractions of acetone and isopropanol using the method of internal standard in the gas phase. n-Butanol served as the internal standard. The detection limit was found to be 1.45 for acetone, 1.43 for isopropanol, and 1.28·10–12 g/s for n-butanol. Relative standard deviation (the recurrence factor) did not exceed 4.3 % at a confidence level P = 0.95.","PeriodicalId":17783,"journal":{"name":"Kataliz v promyshlennosti","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2023-01-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Kataliz v promyshlennosti","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.18412/1816-0387-2023-1-29-36","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Capillary chromatographic columns with different stationary phases were studied in order to compare their ability to estimate purity of isopropyl alcohol obtained by hydrogenation of acetone. The study was performed with the columns based on polyethylene glycol 20 M terminated with 2-nitroterephthalic acid (PEG20M/FFAP); poly(1-trimethylsilyl-1-propyne) (PTMSP032); and trifluoropropyl (25 %) methyl silicone elastomer (SKTFT 50X). A comparison of the measurement time, asymmetry factors (As) for components of the mixtures, and resolution (Rs) for the pairs of compounds acetone/isopropanol and isopropanol/internal standard gave grounds to choose a capillary column PEG20M/FFAP. A technique was developed for measuring the weight fractions of acetone and isopropanol using the method of internal standard in the gas phase. n-Butanol served as the internal standard. The detection limit was found to be 1.45 for acetone, 1.43 for isopropanol, and 1.28·10–12 g/s for n-butanol. Relative standard deviation (the recurrence factor) did not exceed 4.3 % at a confidence level P = 0.95.
丙酮加氢产物气相色谱分析方法与技术的发展
研究了不同固定相的毛细管色谱柱对丙酮加氢所得异丙醇纯度的测定能力。色谱柱为聚乙二醇20M,端接2-硝基对苯二甲酸(PEG20M/FFAP);保利(1-trimethylsilyl-1-propyne) (PTMSP032);和三氟丙基(25%)甲基有机硅弹性体(SKTFT 50X)。比较了测定时间、混合物组分的不对称系数(As)、丙酮/异丙醇和异丙醇/内标对化合物的分辨率(Rs),选择了PEG20M/FFAP毛细管柱。建立了用内标法测定丙酮和异丙醇气相质量分数的方法。正丁醇作为内标。丙酮的检出限为1.45,异丙醇的检出限为1.43,正丁醇的检出限为1.28·10 ~ 12 g/s。相对标准偏差(复发因子)在置信水平P = 0.95下未超过4.3%。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信