A comparative study between Pentacam and optical biometry measurements of anterior-chamber depth and keratometric values in normal and myopic eyes

Hend El-Sayed, Nashwa Lamie, Fatma A. Atwa, Nermien El-Dien Mohammed
{"title":"A comparative study between Pentacam and optical biometry measurements of anterior-chamber depth and keratometric values in normal and myopic eyes","authors":"Hend El-Sayed, Nashwa Lamie, Fatma A. Atwa, Nermien El-Dien Mohammed","doi":"10.4103/sjamf.sjamf_178_21","DOIUrl":null,"url":null,"abstract":"Purpose Accurate measurement of the anterior-segment parameters is important for diagnosis of various pathologies such as cataract and glaucoma. Anterior-chamber depth (ACD) and corneal radii (R1, R2) are important parameters for biometry and refractive surgery. Considering ACD into the calculation of the intraocular lens power is an effective method to reduce postoperative errors. Aim To compare the anterior-segment parameters in the form of ACD, K-readings, and astigmatism measurements by using Pentacam and optical biometer in both normal and myopic eyes. Patients and methods A comparative observational case–control study was conducted at the outpatient clinic of Al-Zahraa University Hospital and Al-Rwad Eye Hospital from April to August 2020. Pentacam measurements was taken at Al-Rwad Eye Hospital and optical biometry was done at Al-Zahraa University Hospital. Our study was performed on 60 phakic eyes classified into two groups: group I (30 normal eyes) and group II (30 myopic eyes). The ACD, keratometry, and corneal astigmatism were measured using corneal topography WaveLight Allegro Oculyzer Pentacam and then optical biometry using Nidek AL Scan Optical Biometer (Nidek Co. Ltd). Results Our study showed that there was no significant difference between the optical biometer and Pentacam in measuring ACD, corneal power, and astigmatism measurements for both normal and myopic groups. Conclusion The WaveLight Allegro Oculyzer Pentacam and the Nidek AL Scan Optical Biometer are excellent noncontact devices and the results obtained using the two devices were well correlated and comparable, so they could be used interchangeably in most clinical settings.","PeriodicalId":22975,"journal":{"name":"The Scientific Journal of Al-Azhar Medical Faculty, Girls","volume":"75 1","pages":"824 - 830"},"PeriodicalIF":0.0000,"publicationDate":"2021-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The Scientific Journal of Al-Azhar Medical Faculty, Girls","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/sjamf.sjamf_178_21","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Purpose Accurate measurement of the anterior-segment parameters is important for diagnosis of various pathologies such as cataract and glaucoma. Anterior-chamber depth (ACD) and corneal radii (R1, R2) are important parameters for biometry and refractive surgery. Considering ACD into the calculation of the intraocular lens power is an effective method to reduce postoperative errors. Aim To compare the anterior-segment parameters in the form of ACD, K-readings, and astigmatism measurements by using Pentacam and optical biometer in both normal and myopic eyes. Patients and methods A comparative observational case–control study was conducted at the outpatient clinic of Al-Zahraa University Hospital and Al-Rwad Eye Hospital from April to August 2020. Pentacam measurements was taken at Al-Rwad Eye Hospital and optical biometry was done at Al-Zahraa University Hospital. Our study was performed on 60 phakic eyes classified into two groups: group I (30 normal eyes) and group II (30 myopic eyes). The ACD, keratometry, and corneal astigmatism were measured using corneal topography WaveLight Allegro Oculyzer Pentacam and then optical biometry using Nidek AL Scan Optical Biometer (Nidek Co. Ltd). Results Our study showed that there was no significant difference between the optical biometer and Pentacam in measuring ACD, corneal power, and astigmatism measurements for both normal and myopic groups. Conclusion The WaveLight Allegro Oculyzer Pentacam and the Nidek AL Scan Optical Biometer are excellent noncontact devices and the results obtained using the two devices were well correlated and comparable, so they could be used interchangeably in most clinical settings.
Pentacam与光学生物测量法测量正常和近视眼前房深度和角膜测量值的比较研究
目的准确测量前段参数对白内障、青光眼等多种病变的诊断具有重要意义。前房深度(ACD)和角膜半径(R1, R2)是生物测量和屈光手术的重要参数。将ACD纳入人工晶状体度数的计算中是减少术后误差的有效方法。目的比较Pentacam和光学生物计在正常眼和近视眼的前节参数ACD、k值和散光测量值。患者和方法于2020年4 - 8月在Al-Zahraa大学医院和Al-Rwad眼科医院门诊进行比较观察性病例对照研究。在Al-Rwad眼科医院进行了Pentacam测量,在Al-Zahraa大学医院进行了光学生物测定。研究对象为60只晶状眼,分为两组:ⅰ组(30只正常眼)和ⅱ组(30只近视眼)。采用角膜地形图waveight Allegro Oculyzer Pentacam测量ACD、角膜屈光度和角膜散光,然后使用Nidek AL Scan optical Biometer (Nidek Co. Ltd)进行光学生物测定。结果光学生物计和Pentacam在正常组和近视组的ACD、角膜度数和散光测量方面无显著差异。结论:waveight Allegro Oculyzer Pentacam与Nidek AL扫描光学生物计是一种优秀的非接触式仪器,两种仪器的检测结果具有良好的相关性和可比性,可在大多数临床环境中互换使用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信