{"title":"TESTING THE COBALT ION EMISSION ON THE STAND\nFOR DIFFERENT ANGLES OF POSITIONING THE COMPONENTS\nOF THE METAL–METAL HIP JOINT ENDOPROSTHESIS","authors":"T. Wiśniewski, M. Libera","doi":"10.5604/01.3001.0015.6899","DOIUrl":null,"url":null,"abstract":"The paper deals with the subject related to the assessment of the influence of the axis angle of the metal\ncomponents of the hip joint on the emission of cobalt ions. The tribological tests were carried out with the use\nof a simulator for the examination of hip joint endoprostheses, the structure of which enables the fixation of\nendoprosthesis components in accordance with the anatomical structure of the human hip joint. During the\ntests, the simulator performs flexion and extension movements as well as loads occurring in the human hip\njoint while walking. Loss-wear tests were carried out for nine variants of the “head–cup” system settings.\nThese settings were determined on the basis of CT images obtained from patients after arthroplasty. After\nthe tribological tests were completed, samples of the lubricating fluid with the wear products suspended in it\nwere collected in order to determine the concentration of cobalt ions, which was carried out using the atomic\nabsorption spectrometry method. As a result, the influence of the head antetorsion angle (α) and the acetabular\nanteversion angle (β) on the concentration of cobalt ions was analysed.\n\n","PeriodicalId":35004,"journal":{"name":"Tribologia: Finnish Journal of Tribology","volume":"27 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2021-12-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Tribologia: Finnish Journal of Tribology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5604/01.3001.0015.6899","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Engineering","Score":null,"Total":0}
引用次数: 0
Abstract
The paper deals with the subject related to the assessment of the influence of the axis angle of the metal
components of the hip joint on the emission of cobalt ions. The tribological tests were carried out with the use
of a simulator for the examination of hip joint endoprostheses, the structure of which enables the fixation of
endoprosthesis components in accordance with the anatomical structure of the human hip joint. During the
tests, the simulator performs flexion and extension movements as well as loads occurring in the human hip
joint while walking. Loss-wear tests were carried out for nine variants of the “head–cup” system settings.
These settings were determined on the basis of CT images obtained from patients after arthroplasty. After
the tribological tests were completed, samples of the lubricating fluid with the wear products suspended in it
were collected in order to determine the concentration of cobalt ions, which was carried out using the atomic
absorption spectrometry method. As a result, the influence of the head antetorsion angle (α) and the acetabular
anteversion angle (β) on the concentration of cobalt ions was analysed.